ASTM F1801-1997(2004) Standard Practice for Corrosion Fatigue Testing of Metallic Implant Materials《金属植入物腐蚀疲劳试验的标准操作规程》.pdf
《ASTM F1801-1997(2004) Standard Practice for Corrosion Fatigue Testing of Metallic Implant Materials《金属植入物腐蚀疲劳试验的标准操作规程》.pdf》由会员分享,可在线阅读,更多相关《ASTM F1801-1997(2004) Standard Practice for Corrosion Fatigue Testing of Metallic Implant Materials《金属植入物腐蚀疲劳试验的标准操作规程》.pdf(6页珍藏版)》请在麦多课文档分享上搜索。
1、Designation: F 1801 97 (Reapproved 2004)Standard Practice forCorrosion Fatigue Testing of Metallic Implant Materials1This standard is issued under the fixed designation F 1801; the number immediately following the designation indicates the year oforiginal adoption or, in the case of revision, the ye
2、ar of last revision. A number in parentheses indicates the year of last reapproval. Asuperscript epsilon (e) indicates an editorial change since the last revision or reapproval.1. Scope1.1 This practice covers the procedure for performing cor-rosion fatigue tests to obtain S-N fatigue curves or stat
3、isticallyderived fatigue strength values, or both, for metallic implantmaterials. This practice describes the testing of axially loadedfatigue specimens subjected to a constant amplitude, periodicforcing function in saline solution at 37C and in air at roomtemperature. The environmental test method
4、for implant mate-rials may be adapted to other modes of fatigue loading such asbending or torsion. While this practice is not intended to applyto fatigue tests on implantable components or devices, it doesprovide guidelines for fatigue tests with standard specimens inan environment related to physio
5、logical conditions.1.2 The values stated in SI units are to be regarded as thestandard.1.3 This standard does not purport to address all of thesafety concerns, if any, associated with its use. It is theresponsibility of the user of this standard to establish appro-priate safety and health practices
6、and determine the applica-bility of regulatory limitations prior to use.2. Referenced Documents2.1 ASTM Standards:2E 4 Practices for Force Verification of Testing MachinesE 466 Practice for Conducting Force Controlled ConstantAmplitude Axial Fatigue Tests of Metallic MaterialsE 467 Practice for Veri
7、fication of Constant Amplitude Dy-namic Loads on Displacements in an Axial Load FatigueTesting MachineE 468 Practice for Presentation of Constant Amplitude Fa-tigue Test Results for Metallic MaterialsE 739 Practice for Statistical Analysis of Linear or Linear-ized Stress-Life (S-N) or Strain-Life (e
8、-N) Fatigue DataE 1012 Practice for Verification of Specimen AlignmentUnder Tensile LoadingE 1150 Definitions of Terms Relating to Fatigue3F 86 Practice for Surface Preparation and Marking of Me-tallic Surgical ImplantsF 601 Practice for Fluorescent Penetrant Inspection of Me-tallic Surgical Implant
9、sG 15 Terminology Relating to Corrosion and CorrosionTesting2.2 ANSI Standard:ANSI B46.1 Surface Texture43. Terminology3.1 Definitions:3.1.1 The terminology used in conjunction with this practicecomplies to Terminology E 1150 and Terminology G 15.3.2 Definitions of Terms Specific to This Standard:3.
10、2.1 S-N curvesS-N curves (also known as Whler-curves) show the correlation between the applied stress (S) andthe counted number (N) of cycles to failure.4. Significance and Use4.1 Implants, particularly orthopedic devices, are usuallyexposed to dynamic forces. Thus, implant materials must havehigh f
11、atigue resistance in the physiological environment.4.1.1 This practice provides a procedure for fatigue testingin a simulated physiological environment. Axial tension-tension fatigue tests in an environmental test chamber arerecommended as a standard procedure. The axial fatigueloading shall comply
12、with Practice E 466 and Practice E 467.4.1.1.1 Bending and rotating bending beam fatigue tests ortorsion tests may be performed in a similar environmental cell.4.1.2 This practice is intended to assess the fatigue andcorrosion fatigue properties of materials that are employed orprojected to be emplo
13、yed for implants. This practice is suitablefor studying the effects of different material treatments andsurface conditions on the fatigue behavior of implant materials.The loading mode of the actual implants may be different fromthat of this practice. Determining the fatigue behavior ofimplants and
14、implant components may require separate teststhat consider the specific design and loading mode.4.1.3 As a substitute for body fluid, 0.9 % saline solution isrecommended as a standard environment. One of the various1This practice is under the jurisdiction of ASTM Committee F04 on Medical andSurgical
15、 Materials and Devices and is the direct responsibility of SubcommitteeF04.15 on Material Test Methods.Current edition approved Aug. 1, 2004. Published August 2004. Originallyapproved in 1997. Last previous edition approved in 1997 as F 1801 97.2For referenced ASTM standards, visit the ASTM website,
16、 www.astm.org, orcontact ASTM Customer Service at serviceastm.org. For Annual Book of ASTMStandards volume information, refer to the standards Document Summary page onthe ASTM website.3Withdrawn.4Available from American National Standards Institute (ANSI), 25 W. 43rd St.,4th Floor, New York, NY 1003
17、6.1Copyright ASTM International, 100 Barr Harbor Drive, PO Box C700, West Conshohocken, PA 19428-2959, United States.Ringers solutions or another substitute for body fluid may alsobe suitable for particular tests. However, these various solu-tions may not give equal fatigue endurance results. Thechl
18、oride ions are the most critical constituent in these solutionsin initiating corrosion fatigue.4.1.4 Because implants are manufactured from highly cor-rosion resistant materials, no visible corrosion may be detect-able by optical or electron-optical (SEM) means. Only adecrease of fatigue strength in
19、 the high cyclic life range may benoticeable. Therefore, S-N curves covering a broad fatigueloading range should be generated in 0.9 % saline solution(Ringers solutions) and air. Comparison of fatigue curvesgenerated in air and saline solution may be the only way toassess the effect of the saline en
20、vironment.4.1.5 Where the fatigue behavior of a material system isalready established, it may suffice to test modifications of thematerial properties or surface condition in only a selected stressrange.4.1.6 The recommended loading frequency of one hertzcorresponds to the frequency of weight-bearing
21、 during walk-ing. For screening tests, higher test frequencies may be used;but it must be realized that higher frequencies may affect theresults.4.1.7 Summary of Standard ConditionsFor inter-laboratory comparisons the following conditions are consid-ered as the standard test. Axial tension-tension t
22、ests withcylindrical specimens in 37C 0.9 % saline solution and airunder a loading frequency of 1 Hz.5. Testing Equipment5.1 The mechanics of the testing machine should be ana-lyzed to ensure that the machine is capable of maintaining thedesired form and magnitude of loading for the duration of thet
23、est (see Practice E 4).5.2 Axial Fatigue Testing:5.2.1 Tension-tension fatigue tests may be performed on oneof the following types of axial fatigue testing machines:5.2.1.1 Mechanical,5.2.1.2 Electromechanical or magnetically driven, and5.2.1.3 Hydraulic or electrohydraulic.5.2.2 The machine shall h
24、ave a load-monitoring system,such as a transducer mounted in series with the specimen. Thetest loads shall be monitored continuously in the early stage ofthe test and periodically thereafter, to ensure that the desiredload is maintained. The magnitude of the varying loads,measured dynamically as des
- 1.请仔细阅读文档,确保文档完整性,对于不预览、不比对内容而直接下载带来的问题本站不予受理。
- 2.下载的文档,不会出现我们的网址水印。
- 3、该文档所得收入(下载+内容+预览)归上传者、原创作者;如果您是本文档原作者,请点此认领!既往收益都归您。
下载文档到电脑,查找使用更方便
5000 积分 0人已下载
下载 | 加入VIP,交流精品资源 |
- 配套讲稿:
如PPT文件的首页显示word图标,表示该PPT已包含配套word讲稿。双击word图标可打开word文档。
- 特殊限制:
部分文档作品中含有的国旗、国徽等图片,仅作为作品整体效果示例展示,禁止商用。设计者仅对作品中独创性部分享有著作权。
- 关 键 词:
- ASTMF180119972004STANDARDPRACTICEFORCORROSIONFATIGUETESTINGOFMETALLICIMPLANTMATERIALS 金属 植入 腐蚀 疲劳 试验

链接地址:http://www.mydoc123.com/p-535698.html