ASTM E578-2007 Standard Test Method for Linearity of Fluorescence Measuring Systems《荧光测量系统的线性度用标准试验方法》.pdf
《ASTM E578-2007 Standard Test Method for Linearity of Fluorescence Measuring Systems《荧光测量系统的线性度用标准试验方法》.pdf》由会员分享,可在线阅读,更多相关《ASTM E578-2007 Standard Test Method for Linearity of Fluorescence Measuring Systems《荧光测量系统的线性度用标准试验方法》.pdf(3页珍藏版)》请在麦多课文档分享上搜索。
1、Designation: E 578 07Standard Test Method forLinearity of Fluorescence Measuring Systems1This standard is issued under the fixed designation E 578; the number immediately following the designation indicates the year oforiginal adoption or, in the case of revision, the year of last revision. A number
2、 in parentheses indicates the year of last reapproval. Asuperscript epsilon (e) indicates an editorial change since the last revision or reapproval.1. Scope1.1 This test method covers a procedure for evaluating thelimits of the linearity of response with fluorescence intensity offluorescence-measuri
3、ng systems under operating conditions.Particular attention is given to slit widths, filters, and samplecontainers. This test method can be used to test the overalllinearity under a wide variety of instrumental and samplingconditions. The results obtained apply only to the testedcombination of slit w
4、idth and filters, and the size, type andillumination of the sample cuvette, all of which must be statedin the report. The sources of nonlinearity may be the measuringelectronics, excessive absorption of either the exciting oremitted radiation, or both, and the sample handling technique,particularly
5、at low concentrations.1.2 This test method has been applied to fluorescence-measuring systems utilizing continuous and low-energy exci-tation sources (for example, an excitation source of 450-Welectrical input or less). There is no assurance that extremelyintense illumination will not cause photodec
6、omposition of thecompounds suggested in this test method.2For this reason it isrecommended that this test method not be indiscriminatelyemployed with high-intensity light sources. It is not a testmethod to determine the linearity of response of other materi-als. If this test method is extended to em
7、ploy other chemicalsubstances, the principles within can be applied, but newmaterial parameters, such as the concentration range of linear-ity, must be established. The user should be aware of thepossibility that these other substances may undergo decompo-sition, or adsorption onto containers.1.3 Th
8、is test method has been applied to fluorescence-measuring systems utilizing a single detector, that is, a photo-multiplier tube or a single photodiode. It has not been demon-strated if this method is effective for photo-array instrumentssuch as those using a CCD or a diode array detector.1.4 This te
9、st method is applicable to 10-mm pathlengthcuvette formats and instruments covering a wavelength rangewithin 190 to 900 nm. The use of other sample formats has notbeen established with this test method.1.5 This standard does not purport to address all of thesafety concerns, if any, associated with i
10、ts use. It is theresponsibility of the user of this standard to establish appro-priate safety and health practices and determine the applica-bility of regulatory limitations prior to use.2. Summary of Test Method2.1 This procedure is used for testing the linearity offluorescence-measuring systems by
11、 using solutions of quininesulfate dihydrate in sulfuric acid as standard test solutions.Other stable solutions which may be more suitable to the usercan be employed (Note 1). The standard used to determinelinearity should be stated in the report. The fluorescence of thetest solution is measured in
12、the measuring system with thecuvettes, slits, or filters that are to be employed in projecteduse.NOTE 1A substitute standard should have the following properties:(1) It should have a large quantum yield at very high dilution; (2) it shouldbe stable to the exciting radiation during spectral measureme
13、nts; (3) itsfluorescence and its absorption spectra overlap should be small; (4) itsquantum yield should not be strongly concentration dependent; and (5)itshould have a broad emission spectrum, so that little error is introducedwhen wide slits are used.32.2 Upper Limit of LinearityThe fluorescence i
14、ntensity ofa series of standard solutions is measured, the resultantinstrument readings are plotted against concentration on alog-log graph, and a smooth curve is drawn through the datapoints. The point (concentration) at which the upper end of thecurve deviates by more than 5 % of the signal from t
15、he straightline (defined by the center region of the curve) is taken as theupper limit of linearity. The limit is expressed in microgramsper millilitre of quinine sulfate dihydrate.NOTE 2Absorption of the exciting radiation at high solute concentra-tions is dependent on instrument geometry and pathl
16、ength, and can resultin fluorescence signal nonlinearity.1This test method is under the jurisdiction of ASTM Committee E13 onMolecular Spectroscopy and Separation Science and is the direct responsibility ofSubcommittee E13.01 on Ultra-Violet, Visible, and Luminescence Spectroscopy.Current edition ap
17、proved March 1, 2007. Published March 2007. Originallyapproved in 1976. Last previous edition approved in 2001 as E 578 01.2Lukasiewicz, R. J., and Fitzgerald, J. M., Analytical Chemistry, ANCHA, Vol45, 1973, p. 511.3Gill, J. E., Photochemistry and Photobiology, PHCBA, Vol 9, 1969, p. 313.1Copyright
18、 ASTM International, 100 Barr Harbor Drive, PO Box C700, West Conshohocken, PA 19428-2959, United States.2.3 Lower Limit of LinearityThe lower limit of linearity istaken as the point (concentration) at which the lower end of thecurve deviates from the straight line defined by the centralportion of t
19、he curve by more than twice the average percentdeviation of the points that determine the straight line.3. Significance and Use3.1 The range of concentration of a fluorescing substance insolution over which the fluorescence varies linearly with theconcentration is the range most useful for quantitat
20、ive analysis.This range is affected by properties of the solution underanalysis and by features of the measuring system. This testmethod provides a means of testing the performance of afluorescence measuring system and of determining the concen-tration range over which the system is suitable for mak
21、ing agiven quantitative analysis.3.2 This test method is not meant for comparing theperformance of different fluorescence measuring instruments.4. Apparatus4.1 Fluorescence-Measuring System, fully equipped forprojected use with a suitable UV source to cover the excitationwavelengths of quinine sulfa
22、te and a photodetector sensitive at450 nm.5. Standard Solutions5.1 Prepare a stock solution of quinine sulfate dihydrate bytransferring 0.100 g of crystalline dihydrate of quinine sulfate,(C20H24O2N2)2H2SO42H2O, National Institute of Standardsand Technology SRM 936 (or equivalent), into a 100-mLvolu
- 1.请仔细阅读文档,确保文档完整性,对于不预览、不比对内容而直接下载带来的问题本站不予受理。
- 2.下载的文档,不会出现我们的网址水印。
- 3、该文档所得收入(下载+内容+预览)归上传者、原创作者;如果您是本文档原作者,请点此认领!既往收益都归您。
下载文档到电脑,查找使用更方便
5000 积分 0人已下载
下载 | 加入VIP,交流精品资源 |
- 配套讲稿:
如PPT文件的首页显示word图标,表示该PPT已包含配套word讲稿。双击word图标可打开word文档。
- 特殊限制:
部分文档作品中含有的国旗、国徽等图片,仅作为作品整体效果示例展示,禁止商用。设计者仅对作品中独创性部分享有著作权。
- 关 键 词:
- ASTME5782007STANDARDTESTMETHODFORLINEARITYOFFLUORESCENCEMEASURINGSYSTEMS 荧光 测量 系统 线性 标准 试验 方法 PDF

链接地址:http://www.mydoc123.com/p-533075.html