ASTM E3029-2015 Standard Practice for Determining Relative Spectral Correction Factors for Emission Signal of Fluorescence Spectrometers《用于测定荧光光谱仪发射信号相对光谱校正因子的标准实施规程》.pdf
《ASTM E3029-2015 Standard Practice for Determining Relative Spectral Correction Factors for Emission Signal of Fluorescence Spectrometers《用于测定荧光光谱仪发射信号相对光谱校正因子的标准实施规程》.pdf》由会员分享,可在线阅读,更多相关《ASTM E3029-2015 Standard Practice for Determining Relative Spectral Correction Factors for Emission Signal of Fluorescence Spectrometers《用于测定荧光光谱仪发射信号相对光谱校正因子的标准实施规程》.pdf(5页珍藏版)》请在麦多课文档分享上搜索。
1、Designation: E3029 15Standard Practice forDetermining Relative Spectral Correction Factors forEmission Signal of Fluorescence Spectrometers1This standard is issued under the fixed designation E3029; the number immediately following the designation indicates the year oforiginal adoption or, in the ca
2、se of revision, the year of last revision. A number in parentheses indicates the year of last reapproval. Asuperscript epsilon () indicates an editorial change since the last revision or reapproval.1. Scope1.1 This practice (1)2describes three methods for determin-ing the relative spectral correctio
3、n factors for grating-basedfluorescence spectrometers in the ultraviolet-visible spectralrange. These methods are intended for instruments with a0/90 transmitting sample geometry. Each method uses dif-ferent types of transfer standards, including 1) a calibrated lightsource (CS), 2) a calibrated det
4、ector (CD) and a calibrateddiffuse reflector (CR), and 3) certified reference materials(CRMs). The wavelength region covered by the differentmethods ranges from 250 to 830 nm with some methods havinga broader range than others. Extending these methods to thenear infrared (NIR) beyond 830 nm will be
5、discussed briefly,where appropriate. These methods were designed for scanningfluorescence spectrometers with a single channel detector, butcan also be used with a multichannel detector, such as a diodearray or a CCD.1.2 The values stated in SI units are to be regarded asstandard. No other units of m
6、easurement are included in thisstandard.1.3 This standard does not purport to address all of thesafety concerns, if any, associated with its use. It is theresponsibility of the user of this standard to establish appro-priate safety and health practices and determine the applica-bility of regulatory
7、limitations prior to use.2. Referenced Documents2.1 ASTM Standards:3E131 Terminology Relating to Molecular SpectroscopyE388 Test Method for Wavelength Accuracy and SpectralBandwidth of Fluorescence SpectrometersE578 Test Method for Linearity of Fluorescence MeasuringSystemsE2719 Guide for Fluorescen
8、ceInstrument Calibration andQualification3. Significance and Use (Intro)3.1 Calibration of the responsivity of the detection systemfor emission (EM) as a function of EM wavelength (EM), alsoreferred to as spectral correction of emission, is necessary forsuccessful quantification when intensity ratio
9、s at different EMwavelengths are being compared or when the true shape orpeak maximum position of an EM spectrum needs to beknown. Such calibration methods are given here and summa-rized in Table 1. This type of calibration is necessary becausethe spectral responsivity of a detection system can chan
10、gesignificantly over its useful wavelength range (see Fig. 1). It ishighly recommended that the wavelength accuracy (see TestMethod E388) and the linear range of the detection system (seeGuide E2719 and Test Method E578) be determined beforespectral calibration is performed and that appropriate step
11、s aretaken to insure that all measured intensities during this cali-bration are within the linear range. For example, when usingwide slit widths in the monochromators, attenuators may beneeded to attenuate the excitation beam or emission, thereby,decreasing the fluorescence intensity at the detector
12、. Also notethat when using an EM polarizer, the spectral correction foremission is dependent on the polarizer setting. (2)Itisimportant to use the same instrument settings for all of thecalibration procedures mentioned here, as well as for subse-quent sample measurements.3.2 When using CCD or diode
13、array detectors with aspectrometer for EMselection, the spectral correction factorsare dependent on the grating position of the spectrometer.Therefore, the spectral correction profile versus EMmust bedetermined separately for each grating position used. (3)3.3 Instrument manufacturers often provide
14、an automatedprocedure and calculation for a spectral correction function foremission, or they may supply a correction that was determinedat the factory. This correction can often be applied duringspectral collection or as a post-collection correction. The user1This practice is under the jurisdiction
15、 of ASTM Committee E13 on MolecularSpectroscopy and Separation Science and is the direct responsibility of Subcom-mittee E13.01 on Ultra-Violet, Visible, and Luminescence Spectroscopy.Current edition approved Sept. 1, 2015. Published October 2015. DOI: 10.1520/E3029-152The boldface numbers in parent
16、heses refer to a list of references at the end ofthis standard.3For referenced ASTM standards, visit the ASTM website, www.astm.org, orcontact ASTM Customer Service at serviceastm.org. For Annual Book of ASTMStandards volume information, refer to the standards Document Summary page onthe ASTM websit
17、e.Copyright ASTM International, 100 Barr Harbor Drive, PO Box C700, West Conshohocken, PA 19428-2959. United States1should be advised to verify that the automated vendor proce-dure and calculation or supplied correction are performed anddetermined according to the guidelines given within thisstandar
18、d.4. Calibrated Optical Radiation Source (CS) Method (seeTest Method E578,(4-6, 14)4.1 Materials:4.1.1 A calibrated tungsten lamp is most commonly used asa CS in the visible region due to its high intensity and broad,featureless spectral profile. Its intensity falls off quickly in theultraviolet (UV
19、) region, but it can typically be used down to350 nm or so. It also displays a high intensity in the nearinfrared, peaking at about 1000 nm. Its intensity graduallydecreases beyond 1000 nm, but continues to have significantintensity out to about 2500 nm. A calibrated deuterium lampcan be used to ext
20、end farther into the UV with an effectiverange from about 200 to 380 nm. The effective range of a CSis dependent on the intensity of the CS and the sensitivity of thedetection system. This range can be determined by measuringthe low-signal regions where the signal profile of the light fromthe CS bec
21、omes flat or indistinguishable from the backgroundsignal, implying that the signal afforded by the CS is notmeasurable in these EMregions.4.1.2 A calibrated reflector (CR) is often used to reflect thelight from the CS into the emission detection system. A diffusereflector made of compressed or sinte
22、red polytetrafluoroethyl-ene (PTFE) is most commonly used as a CR, due to its nearlyLambertian reflectance, which prevents both polarization andspatial dependence of the reflectance. In addition, PTFEpossesses a reflectance profile that is nearly flat, changing byless than 10 % from 250 to 2500 nm.
23、For a CS and a CR,“calibrated” implies that the spectral radiance and the spectralreflectance, respectively, are known (calibrated wavelengthdependence of the spectral radiant factor including measure-ment uncertainty) and traceable to the SI (International Systemof Units). This is commonly done thr
24、ough certification of thesevalues by a national metrology institute (NMI). (15, 16, 7)4.2 Procedure:4.2.1 Direct the optical radiation from a CS into the EMdetection system by placing the CS at the sample position. Ifthe CS is too large to be placed at the sample position, place aCR at the sample po
- 1.请仔细阅读文档,确保文档完整性,对于不预览、不比对内容而直接下载带来的问题本站不予受理。
- 2.下载的文档,不会出现我们的网址水印。
- 3、该文档所得收入(下载+内容+预览)归上传者、原创作者;如果您是本文档原作者,请点此认领!既往收益都归您。
下载文档到电脑,查找使用更方便
5000 积分 0人已下载
下载 | 加入VIP,交流精品资源 |
- 配套讲稿:
如PPT文件的首页显示word图标,表示该PPT已包含配套word讲稿。双击word图标可打开word文档。
- 特殊限制:
部分文档作品中含有的国旗、国徽等图片,仅作为作品整体效果示例展示,禁止商用。设计者仅对作品中独创性部分享有著作权。
- 关 键 词:
- ASTME30292015STANDARDPRACTICEFORDETERMININGRELATIVESPECTRALCORRECTIONFACTORSFOREMISSIONSIGNALOFFLUORESCENCESPECTROMETERS

链接地址:http://www.mydoc123.com/p-532265.html