ASTM E2009-2008 Standard Test Method for Oxidation Onset Temperature of Hydrocarbons by Differential Scanning Calorimetry《用差别扫描热量测定法测定烃类氧化开始温度的标准试验方法》.pdf
《ASTM E2009-2008 Standard Test Method for Oxidation Onset Temperature of Hydrocarbons by Differential Scanning Calorimetry《用差别扫描热量测定法测定烃类氧化开始温度的标准试验方法》.pdf》由会员分享,可在线阅读,更多相关《ASTM E2009-2008 Standard Test Method for Oxidation Onset Temperature of Hydrocarbons by Differential Scanning Calorimetry《用差别扫描热量测定法测定烃类氧化开始温度的标准试验方法》.pdf(5页珍藏版)》请在麦多课文档分享上搜索。
1、Designation: E 2009 08Standard Test Method forOxidation Onset Temperature of Hydrocarbons byDifferential Scanning Calorimetry1This standard is issued under the fixed designation E 2009; the number immediately following the designation indicates the year oforiginal adoption or, in the case of revisio
2、n, the year of last revision. A number in parentheses indicates the year of last reapproval. Asuperscript epsilon () indicates an editorial change since the last revision or reapproval.1. Scope1.1 This test method describes the determination of theoxidative properties of hydrocarbons by differential
3、 scanningcalorimetry or pressure differential scanning calorimetry underlinear heating rate conditions and is applicable to hydrocar-bons, which oxidize exothermically in their analyzed form.1.2 Test Method AA differential scanning calorimeter(DSC) is used at ambient pressure, of one atmosphere ofox
4、ygen.1.3 Test Method BApressure DSC (PDSC) is used at highpressure, for example, 3.5 MPa (500 psig) oxygen.1.4 Test Method CA differential scanning calorimeter(DSC) is used at ambient pressure of one atmosphere of air.1.5 The values stated in SI units are to be regarded asstandard. No other units of
5、 measurement are included in thisstandard.1.6 This standard does not purport to address all of thesafety concerns, if any, associated with its use. It is theresponsibility of the user of this standard to establish appro-priate safety and health practices and determine the applica-bility of regulator
6、y limitations prior to use.2. Referenced Documents2.1 ASTM Standards:2D 3350 Specification for Polyethylene Plastics Pipe andFittings MaterialsD 3895 Test Method for Oxidative-Induction Time of Poly-olefins by Differential Scanning CalorimetryD 4565 Test Methods for Physical and Environmental Per-fo
7、rmance Properties of Insulations and Jackets for Tele-communications Wire and CableD 5483 Test Method for Oxidation Induction Time of Lu-bricating Greases by Pressure Differential Scanning Calo-rimetryE 473 Terminology Relating to Thermal Analysis and Rhe-ologyE 691 Practice for Conducting an Interl
8、aboratory Study toDetermine the Precision of a Test MethodE 967 Test Method for Temperature Calibration of Differ-ential Scanning Calorimeters and Differential ThermalAnalyzersE 1858 Test Method for Determining Oxidation InductionTime of Hydrocarbons by Differential Scanning Calorim-etry3. Terminolo
9、gy3.1 Definitions:3.1.1 oxidation (extrapolated) onset temperature (OOT)arelative measure of oxidative stability at the cited heating rateis determined from data recorded during a DSC scanningtemperature test. The temperature at which the onset to theobserved oxidation is taken as the OOT.3.1.2 For
10、definitions of terms used in this test method, referto Terminology E 473.4. Summary of Method4.1 The test specimen in an aluminum container and anempty reference aluminum container or pan are heated at aspecified constant heating rate in an oxygen (or air) environ-ment. Heat flow out of the specimen
11、 is monitored as a functionof temperature until the oxidative reaction is manifested byheat evolution on the thermal curve. The oxidation (extrapo-lated) onset temperature (OOT), a relative measure of oxida-tive stability at the cited heating rate, is determined from datarecorded during the scanning
12、 temperature test. The OOTmeasurement is initiated upon reaching the exothermic reactionand measuring the extrapolated onset temperature.4.2 For some particularly stable materials, the OOT may bequite high (300 C) at the specified heating rate of theexperiment. Under these circumstances, the OOT may
13、 bereduced by increasing the pressure of oxygen purge gas.Conversely, reducing the partial pressure of oxygen (such as bythe use of air) may retard reactions that proceed too rapidly,with a corresponding increase of the OOT. By admixing1This test method is under the jurisdiction ofASTM Committee E37
14、 on ThermalMeasurements and is the direct responsibility of Subcommittee E37.01 on ThermalTest Methods and Practices.Current edition approved Sept. 1, 2008. Published October 2008. Originallyapproved in 1999. Last previous edition approved in 2002 as E 200902.2For referenced ASTM standards, visit th
15、e ASTM website, www.astm.org, orcontact ASTM Customer Service at serviceastm.org. For Annual Book of ASTMStandards volume information, refer to the standards Document Summary page onthe ASTM website.1Copyright ASTM International, 100 Barr Harbor Drive, PO Box C700, West Conshohocken, PA 19428-2959,
16、United States.oxygen gas with a suitable diluent, for example, nitrogen, theOOT will be increased (see Specification D 3350 and TestMethods D 3895, D 4565, and D 5483).NOTE 1For some systems, the use of copper pans to catalyzeoxidation will reduce the oxidation onset temperature. The results,however
17、, will not necessarily correlate with non-catalyzed tests.5. Significance and Use5.1 Oxidation onset temperature is a relative measure of thedegree of oxidative stability of the material evaluated at a givenheating rate and oxidative environment, for example, oxygen;the higher the OOT value the more
18、 stable the material. TheOOT is described in Fig. 1. The OOT values can be used forcomparative purposes and are not an absolute measurement,like the oxidation induction time (OIT) at a constant tempera-ture (see Test Method E 1858). The presence or effectiveness ofantioxidants may be determined by t
19、his test method.5.2 Typical uses of this test method include the oxidativestability of edible oils and fats (oxidative rancidity), lubricants,greases, and polyolefins.6. Apparatus6.1 Differential Scanning Calorimeter (DSC) or PressureDifferential Scanning Calorimeter (PDSC)The essential in-strumenta
20、tion required to provide the minimum differentialscanning calorimetric capability for this test method includes:a DSC chamber composed of a furnace to provide uniformcontrolled heating of a specimen and a reference to a constantheating rate of at least 10 C/min within the applicabletemperature range
21、 for this test method; a temperature sensor toprovide an indication of the specimen temperature to 6 0.1 C;a differential sensor to detect heat flow (power) differencebetween the specimen and the reference to 0.1 mW; and theinstrument should have the capability of measuring heat flowof at least 6 mW
22、, with provision for less sensitive ranges.NOTE 2In certain cases when the sample under study is of highvolatility (for example, low molecular weight hydrocarbons), the use ofpressures in excess of 0.1 MPa (1 atmosphere) is needed. The operator iscautioned to verify (with apparatus designer) the max
23、imum oxygenpressure at which the apparatus may be safely operated. A PDSC is usedin Method B.6.2 A Data Collection Device, to provide a means ofacquiring, storing, and displaying measured or calculatedsignals, or both. The minimum output signals required for DSCare heat flow, temperature and time.6.
24、3 A high-pressure gas regulator or similar device to adjustthe applied pressure in the test chamber to less than 6 5%,including any temperature dependence on the transducer, isused in Method B.NOTE 3Gas delivery tubing should be kept as short as possible tominimize dead volume. The link between the
- 1.请仔细阅读文档,确保文档完整性,对于不预览、不比对内容而直接下载带来的问题本站不予受理。
- 2.下载的文档,不会出现我们的网址水印。
- 3、该文档所得收入(下载+内容+预览)归上传者、原创作者;如果您是本文档原作者,请点此认领!既往收益都归您。
下载文档到电脑,查找使用更方便
5000 积分 0人已下载
下载 | 加入VIP,交流精品资源 |
- 配套讲稿:
如PPT文件的首页显示word图标,表示该PPT已包含配套word讲稿。双击word图标可打开word文档。
- 特殊限制:
部分文档作品中含有的国旗、国徽等图片,仅作为作品整体效果示例展示,禁止商用。设计者仅对作品中独创性部分享有著作权。
- 关 键 词:
- ASTME20092008STANDARDTESTMETHODFOROXIDATIONONSETTEMPERATUREOFHYDROCARBONSBYDIFFERENTIALSCANNINGCALORIMETRY

链接地址:http://www.mydoc123.com/p-530045.html