ASTM E1944-1998(2013) Standard Practice for Describing and Measuring Performance of Laboratory Fourier Transform Near-Infrared (FT-NIR) Spectrometers Level Zero and Level One Tests.pdf
《ASTM E1944-1998(2013) Standard Practice for Describing and Measuring Performance of Laboratory Fourier Transform Near-Infrared (FT-NIR) Spectrometers Level Zero and Level One Tests.pdf》由会员分享,可在线阅读,更多相关《ASTM E1944-1998(2013) Standard Practice for Describing and Measuring Performance of Laboratory Fourier Transform Near-Infrared (FT-NIR) Spectrometers Level Zero and Level One Tests.pdf(5页珍藏版)》请在麦多课文档分享上搜索。
1、Designation: E1944 98 (Reapproved 2013)Standard Practice forDescribing and Measuring Performance of LaboratoryFourier Transform Near-Infrared (FT-NIR) Spectrometers:Level Zero and Level One Tests1This standard is issued under the fixed designation E1944; the number immediately following the designat
2、ion indicates the year oforiginal adoption or, in the case of revision, the year of last revision. A number in parentheses indicates the year of last reapproval. Asuperscript epsilon () indicates an editorial change since the last revision or reapproval.1. Scope1.1 This practice covers two levels of
3、 tests to measure theperformance of laboratory Fourier transform near infrared(FT-NIR) spectrometers. This practice applies to the short-wave near infrared region, approximately 800 nm (12 500 cm-1) to 1100 nm (9090.91 cm-1); and the long-wavelength nearinfrared region, approximately 1100 nm (9090.9
4、1 cm-1)to2500 nm (4000 cm-1). This practice is intended mainly fortransmittance measurements of gases and liquids, although it isbroadly applicable for reflectance measurements.1.2 The values stated in SI units are to be regarded asstandard. No other units of measurement are included in thisstandard
5、.1.3 This standard does not purport to address all of thesafety concerns, if any, associated with its use. It is theresponsibility of the user of this standard to establish appro-priate safety and health practices and determine the applica-bility of regulatory limitations prior to use.2. Referenced
6、Documents2.1 ASTM Standards:2E131 Terminology Relating to Molecular SpectroscopyE168 Practices for General Techniques of Infrared Quanti-tative AnalysisE932 Practice for Describing and Measuring Performance ofDispersive Infrared SpectrometersE1252 Practice for General Techniques for Obtaining Infra-
7、red Spectra for Qualitative AnalysisE1421 Practice for Describing and Measuring Performanceof Fourier Transform Mid-Infrared (FT-MIR) Spectrom-eters: Level Zero and Level One Tests3. Terminology3.1 For definitions of terms used in this practice, refer toTerminology E131. All identifications of spect
8、ral regions andabsorbance band positions are given in nanometers (nm), andwavenumbers (cm-1); and spectral energy, transmittance,reflectance, and absorbance are signified by the letters E, T, Rand A respectively. A subscripted number signifies a spectralposition in nanometers, with wavenumbers in pa
9、renthesis (forexample, A1940(5154.64), denotes the absorbance at 1940 nm or5154.64 cm-1).4. Significance and Use4.1 This practice permits an analyst to compare the generalperformance of a laboratory instrument on any given day withthe prior performance of that instrument. This practice is notintende
10、d for comparison of different instruments with eachother, nor is it directly applicable to dedicated process FT-NIRanalyzers. This practice requires the use of a check samplecompatible with the instrument under test as described in 5.3.5. Test Conditions5.1 Operating ConditionsIn obtaining spectroph
11、otometricdata for the check sample, the analyst must select the properinstrumental operating conditions in order to realize satisfac-tory instrument performance. Operating conditions for indi-vidual instruments are best obtained from the manufacturersinstructional literature due to the variations wi
12、th instrumentdesign. It should be noted that many FT-NIR instruments aredesigned to work best if left in standby mode when they are notin use. A record should be kept to document the operatingconditions selected during a test so that they can be duplicatedfor future tests. Note that spectrometers ar
13、e to be tested onlywithin their respective recommended measurement wavelength(wavenumber) ranges.5.2 Instrumental characteristics can influence these mea-surements in several ways. Vignetting of the beam (that is, theaperture of the sample cell is smaller than the diameter of thenear infrared beam a
14、t the focus) reduces the transmittancevalue measured in nonabsorbing regions, and on most instru-ments can change the apparent wavelength (or wavenumber)scale by a small amount, usually less than 0.01 nm (0.1 cm-1).1This practice is under the jurisdiction of ASTM Committee E13 on MolecularSpectrosco
15、py and Separation Science and is the direct responsibility of Subcom-mittee E13.03 on Infrared and Near Infrared Spectroscopy.Current edition approved Jan. 1, 2013. Published January 2013. Originallyapproved in 1998. Last previous edition approved in 2007 as E1944 98 (2007).DOI: 10.1520/E1944-98R13.
16、2For referenced ASTM standards, visit the ASTM website, www.astm.org, orcontact ASTM Customer Service at serviceastm.org. For Annual Book of ASTMStandards volume information, refer to the standards Document Summary page onthe ASTM website.Copyright ASTM International, 100 Barr Harbor Drive, PO Box C
17、700, West Conshohocken, PA 19428-2959. United States1Focus changes can also change transmittance values, so thesample should be positioned in the same location in the samplecompartment for each measurement. The angle of acceptance(established by the f number) of the optics between the sampleand dete
18、ctor significantly affects apparent transmittance. Heat-ing of the sample by the beam or by the higher temperatureswhich exist inside most spectrometers changes absorbancessomewhat, and even changes band ratios and locations slightly.Allow the sample to come to thermal equilibrium prior tomeasuremen
19、t.5.3 The recommended check sample should meet the fol-lowing requirements: the check sample should be fully com-patible with the requirements for repeatable sample presenta-tion to the measuring spectrophotometer. The check sampleshould consist of a single pure compound or precisely knownmixture of
20、 compounds which is spectroscopically stable overmonths or years. The spectra obtained from such a checksample should be known to indicate changes in thespectrophotometer, not the check sample itself. It is recom-mended that independent verification of the integrity of thecheck sample be used prior
21、to test measurement. The checksample should be measured under precisely the sample mea-surement conditions of temperature, humidity, and instrumentset up configuration. Suggested check samples may include,but are not limited to the following: for gases, water vapor at5.89 Torr and 1 atmosphere ina2m
22、gascell, or methane at 18psig pressure in a 10 cm gas cell; for liquids, pure spectro-scopic grade hydrocarbon compounds (for example, toluene,decane, isooctane, etc.), or precise mixtures of these purecompounds; for reflectance measurements of solids, rare earthoxides mixed with white halon powder,
23、 or Spectralon3-basedrare earth oxide reflectance standards. Reference reflectancestandards yielding a featureless, near 100 % reflectance spec-trum are pure powdered sulfur, halon, or Spectralon.6. Level Zero Tests6.1 Nature of TestsRoutine checks of instrument perfor-mance can be performed within
24、a few minutes. They aredesigned to uncover malfunctions or other changes in instru-ment operation but not to specifically diagnose or quantita-tively assess any malfunction. For Level Zero tests, a resolu-tion of 4 cm-1and a nominal measurement time of 30 s isrecommended. Resolution and measurement
- 1.请仔细阅读文档,确保文档完整性,对于不预览、不比对内容而直接下载带来的问题本站不予受理。
- 2.下载的文档,不会出现我们的网址水印。
- 3、该文档所得收入(下载+内容+预览)归上传者、原创作者;如果您是本文档原作者,请点此认领!既往收益都归您。
下载文档到电脑,查找使用更方便
5000 积分 0人已下载
下载 | 加入VIP,交流精品资源 |
- 配套讲稿:
如PPT文件的首页显示word图标,表示该PPT已包含配套word讲稿。双击word图标可打开word文档。
- 特殊限制:
部分文档作品中含有的国旗、国徽等图片,仅作为作品整体效果示例展示,禁止商用。设计者仅对作品中独创性部分享有著作权。
- 关 键 词:
- ASTME194419982013STANDARDPRACTICEFORDESCRIBINGANDMEASURINGPERFORMANCEOFLABORATORYFOURIERTRANSFORMNEARINFRAREDFTNIRSPECTROMETERSLEVELZEROANDLEVELONETESTSPDF

链接地址:http://www.mydoc123.com/p-529880.html