1、Designation: E1303 95 (Reapproved 2017)Standard Practice forRefractive Index Detectors Used in Liquid Chromatography1This standard is issued under the fixed designation E1303; the number immediately following the designation indicates the year oforiginal adoption or, in the case of revision, the yea
2、r of last revision. A number in parentheses indicates the year of last reapproval. Asuperscript epsilon () indicates an editorial change since the last revision or reapproval.1. Scope1.1 This practice covers tests used to evaluate the perfor-mance and to list certain descriptive specifications of ar
3、efractive index (RI) detector used as the detection componentof a liquid chromatographic (LC) system.1.2 This practice is intended to describe the performance ofthe detector both independent of the chromatographic system(static conditions, without flowing solvent) and with flowingsolvent (dynamic co
4、nditions).1.3 The values stated in SI units are to be regarded as thestandard.1.4 This standard does not purport to address all of thesafety concerns, if any, associated with its use. It is theresponsibility of the user of this standard to establish appro-priate safety, health, and environmental pra
5、ctices and deter-mine the applicability of regulatory limitations prior to use.1.5 This international standard was developed in accor-dance with internationally recognized principles on standard-ization established in the Decision on Principles for theDevelopment of International Standards, Guides a
6、nd Recom-mendations issued by the World Trade Organization TechnicalBarriers to Trade (TBT) Committee.2. Referenced Documents2.1 ASTM Standards:2E386 Practice for Data Presentation Relating to High-Resolution Nuclear Magnetic Resonance (NMR) Spec-troscopy (Withdrawn 2015)33. Significance and Use3.1
7、Although it is possible to observe and measure each ofseveral characteristics of a detector under different and uniqueconditions, it is the intent of this practice that a complete set ofdetector test results should be obtained under the same oper-ating conditions. It should also be noted that to spe
8、cifycompletely a detectors capability, its performance should bemeasured at several sets of conditions within the useful rangeof the detector.3.2 The objective of this practice is to test the detector underspecified conditions and in a configuration without an LCcolumn. This is a separation independ
9、ent test. In certaincircumstances it might also be necessary to test the detector inthe separation mode with an LC column in the system, and theappropriate concerns are also mentioned. The terms and testsdescribed in this practice are sufficiently general so that theymay be adapted for use at whatev
10、er conditions may be chosenfor other reasons.4. Noise, Drift, and Flow Sensitivity4.1 Descriptions of Terms Specific to This Standard:4.1.1 short term noisethis noise is the mean amplitude inrefractive index units (RIU) for random variations of thedetector signal having a frequency of one or more cy
11、cles perminute. Short term noise limits the smallest signal detectableby an RI detector, limits the precision attainable, and sets thelower limit on the dynamic range. This noise corresponds toobserved noise of the RI detector only. (The actual noise of theLC system may be larger or smaller than the
12、 observed value,depending upon the method of data collection, or signalmonitoring of the detector, since observed noise is a function ofthe frequency, speed of response and the band width of therecorder or other electronic circuit measuring the detectorsignal.)4.1.2 long term noisethis noise is the
13、maximum amplitudein RIU for random variations of the detector signal withfrequencies between 6 and 60 cycles per h (0.1 and 1.0 cyclesper min). It represents noise that may be mistaken for alate-eluting peak. This noise corresponds to the observed noiseonly and may not always be present.4.1.3 driftt
14、he average slope of the long term noise enve-lope expressed in RIU per hour as measured over a period of1h.4.1.4 staticrefers to the noise and drift measured underconditions of no flow.4.1.5 dynamicrefers to the noise and drift measured at aflow rate of 1.0 mL/min.1This practice is under the jurisdi
15、ction of ASTM Committee E13 on MolecularSpectroscopy and Separation Science and is the direct responsibility of Subcom-mittee E13.19 on Separation Science.Current edition approved Oct. 1, 2017. Published October 2017. Originallyapproved in 1989. Last previous edition approved in 2010 as E1303 95 (20
16、10).DOI: 10.1520/E1303-95R17.2For referenced ASTM standards, visit the ASTM website, www.astm.org, orcontact ASTM Customer Service at serviceastm.org. For Annual Book of ASTMStandards volume information, refer to the standards Document Summary page onthe ASTM website.3The last approved version of th
17、is historical standard is referenced onwww.astm.org.Copyright ASTM International, 100 Barr Harbor Drive, PO Box C700, West Conshohocken, PA 19428-2959. United StatesThis international standard was developed in accordance with internationally recognized principles on standardization established in th
18、e Decision on Principles for theDevelopment of International Standards, Guides and Recommendations issued by the World Trade Organization Technical Barriers to Trade (TBT) Committee.14.1.6 flow sensitivitythe rate of change of signal displace-ment (in RIU) vs flow rate (in mL/min) resulting from ste
19、pchanges in flow rate calculated at 1 mL/min as described in4.3.12.4.2 Test Conditions:4.2.1 The same test solvent must be used in both sample andreference cells. The test solvent used and its purity should bespecified. Water equilibrated with the laboratory atmospherecontaining minimum impurities i
20、s the preferred test solvent formeasuring noise and drift. Water for this purpose (preferablypurified by distillation, deionization, or reverse osmosis)should be drawn, filtered through a 0.45-m filter, and allowedto stand in a loosely covered container for several hours atambient temperature in the
21、 laboratory in which testing is to becarried out. This will ensure complete equilibration of thewater with the gases in the laboratory atmosphere.NOTE 1It is essentially impossible to maintain a constant RI value ofde-gassed water and of very dilute samples in de-gassed water. This is dueto the fact
22、 that the difference in refractive index between completelyde-gassed water and atmosphere-equilibrated water is 1.5 106RIU.4Thus, small differences in the concentration of dissolved gases betweenthe sample and the trapped reference can lead to significant errors inmeasurement of solutions where the
23、expected difference in RI due tosolute is of the order of 106RIU or less. Therefore, in order to minimizeerror in determining samples with small RIU differences between them,atmosphere-equilibrated water (5.2.1) is recommended as the solvent fordetermining linearity and minimum detectability (Sectio
24、n 5).4.2.2 The detector should be located at the test site andswitched on at least 24 h prior to the start of testing. Somedetectors provide an oven to thermostat the optics assembly.The oven should be set at a suitable temperature, following themanufacturers recommendations, and this temperature sh
- 1.请仔细阅读文档,确保文档完整性,对于不预览、不比对内容而直接下载带来的问题本站不予受理。
- 2.下载的文档,不会出现我们的网址水印。
- 3、该文档所得收入(下载+内容+预览)归上传者、原创作者;如果您是本文档原作者,请点此认领!既往收益都归您。
下载文档到电脑,查找使用更方便
5000 积分 0人已下载
下载 | 加入VIP,交流精品资源 |
- 配套讲稿:
如PPT文件的首页显示word图标,表示该PPT已包含配套word讲稿。双击word图标可打开word文档。
- 特殊限制:
部分文档作品中含有的国旗、国徽等图片,仅作为作品整体效果示例展示,禁止商用。设计者仅对作品中独创性部分享有著作权。
- 关 键 词:
- ASTME1303199520174375STANDARDPRACTICEFORREFRACTIVEINDEXDETECTORSUSEDINLIQUIDCHROMATOGRAPHY 色谱 折射率 检测器

链接地址:http://www.mydoc123.com/p-528367.html