ASTM E1217-2005 Standard Practice for Determination of the Specimen Area Contributing to the Detected Signal in Auger Electron Spectrometers and Some X-Ray Photoelectron Spectromet.pdf
《ASTM E1217-2005 Standard Practice for Determination of the Specimen Area Contributing to the Detected Signal in Auger Electron Spectrometers and Some X-Ray Photoelectron Spectromet.pdf》由会员分享,可在线阅读,更多相关《ASTM E1217-2005 Standard Practice for Determination of the Specimen Area Contributing to the Detected Signal in Auger Electron Spectrometers and Some X-Ray Photoelectron Spectromet.pdf(7页珍藏版)》请在麦多课文档分享上搜索。
1、Designation: E 1217 05Standard Practice forDetermination of the Specimen Area Contributing to theDetected Signal in Auger Electron Spectrometers and SomeX-Ray Photoelectron Spectrometers1This standard is issued under the fixed designation E 1217; the number immediately following the designation indi
2、cates the year oforiginal adoption or, in the case of revision, the year of last revision. A number in parentheses indicates the year of last reapproval. Asuperscript epsilon (e) indicates an editorial change since the last revision or reapproval.1. Scope1.1 This practice describes methods for deter
3、mining thespecimen area contributing to the detected signal in Augerelectron spectrometers and some types of X-ray photoelectronspectrometers when this area is defined by the electroncollection lens and aperture system of the electron energyanalyzer. The practice is applicable only to those X-raypho
4、toelectron spectrometers in which the specimen area ex-cited by the incident X-ray beam is larger than the specimenarea viewed by the analyzer, in which the photoelectrons travelin a field-free region from the specimen to the analyzerentrance, and in which an auxiliary electron gun can bemounted to
5、produce an electron beam of variable energy on thespecimen.1.2 This practice is recommended as a useful means fordetermining the specimen area viewed by the analyzer fordifferent conditions of spectrometer operation, for verifyingadequate specimen and beam alignment, and for characterizingthe imagin
6、g properties of the electron energy analyzer.1.3 This standard does not purport to address all of thesafety concerns, if any, associated with its use. It is theresponsibility of the user of this standard to establish appro-priate safety and health practices and determine the applica-bility of regula
7、tory limitations prior to use.2. Referenced Documents2.1 ASTM Standards:2E 673 Terminology Relating to Surface AnalysisE 1016 Guide for Literature Describing Properties of Elec-trostatic Electron Spectrometers2.2 ISO Standards:ISO 18115:2001 Surface Chemical AnalysisVocabulary33. Terminology3.1 Defi
8、nitionsSee Terminology E 673 for terms used inAuger electron spectroscopy and X-ray photoelectron spectros-copy.4. Summary of Practice4.1 An electron beam with a selected energy is scannedacross the surface of a test specimen. The beam may bescanned once, that is, a line scan, or in a pattern, that
9、is,rastered. As the electron beam is deflected across the specimensurface, measurements are made of the intensities detected bythe electron energy analyzer as a function of the beam positionfor selected conditions of analyzer operation. The measuredintensities may be due to electrons elastically sca
10、ttered by thespecimen surface, to electrons inelastically scattered by thespecimen, or to Auger electrons emitted by the specimen. Theintensity distributions for a particular detected electron energycan be plotted as a function of beam position in several waysand can be utilized to obtain informatio
11、n on the specimen areacontributing to the detected signal and on analyzer perfor-mance for the particular conditions of operation. This informa-tion can be used to determine the analysis area (see Terminol-ogy E 673 or ISO 18115).5. Significance and Use5.1 Auger electron spectroscopy and X-ray photo
12、electronspectroscopy are used extensively for the surface analysis ofmaterials. This practice summarizes methods for determiningthe specimen area contributing to the detected signal forinstruments in which a focused electron beam can be scannedover a region with dimensions greater than the dimension
13、s ofthe specimen area viewed by the analyzer.5.2 This practice is intended as a means for determining theobserved specimen area for selected conditions of operation ofthe electron energy analyzer. The observed specimen area1This practice is under the jurisdiction of ASTM Committee E42 on SurfaceAnal
14、ysis and is the direct responsibility of Subcommittee E42.03 on Auger ElectronSpectroscopy and X-Ray Photoelectron Spectroscopy.Current edition approved Nov. 1, 2005. Published December 2005. Originallyapproved in 1987. Last previous edition approved in 2000 as E 1217 00.2For referenced ASTM standar
15、ds, visit the ASTM website, www.astm.org, orcontact ASTM Customer Service at serviceastm.org. For Annual Book of ASTMStandards volume information, refer to the standards Document Summary page onthe ASTM website.3Available from American National Standards Institute (ANSI), 25 W. 43rd St.,4th Floor, N
16、ew York, NY 10036.1Copyright ASTM International, 100 Barr Harbor Drive, PO Box C700, West Conshohocken, PA 19428-2959, United States.depends on whether or not the electrons are retarded beforeenergy analysis, the analyzer pass energy or retarding ratio ifthe electrons are retarded before energy anal
17、ysis, the size ofselected slits or apertures, and the value of the electron energyto be measured. The observed specimen area depends on theseselected conditions of operation and also can depend on theadequacy of alignment of the specimen with respect to theelectron energy analyzer.5.3 Any changes in
18、 the observed specimen area as afunction of measurement conditions, for example, electronenergy or analyzer pass energy, may need to be known if thespecimen materials in regular use have lateral inhomogeneitieswith dimensions comparable to the dimensions of the specimenarea viewed by the analyzer.5.
19、4 This practice can give useful information on the imagingproperties of the electron energy analyzer for particular con-ditions of operation. This information can be helpful incomparing analyzer performance with manufacturers specifi-cations.5.5 Examples of the application of the methods described i
20、nthis practice have been published (1-6).45.6 An ISO Technical Report provides guidance on deter-minations of lateral resolution, analysis area, and sample areaviewed by the analyzer in AES and XPS (7). Baer andEngelhard have used well-defined dots of a material on asubstrate to determine the area o
21、f a specimen contributing tothe measured signal of a small-area XPS measurement (8).This area could be as much as ten times the area estimatedsimply from the lateral resolution of the instrument. Theamount of intensity in fringe or tail regions could also behighly dependent on lens operation and the
22、 adequacy ofspecimen alignment. In scanningAuger microscopy, the area ofanalysis is determined more by the radial extent of backscat-tered electrons than by the radius of the primary beam (9, 10).6. Apparatus6.1 Test Specimen, preferably a conductor, is required and ismounted in the Auger electron o
23、r X-ray photoelectron spec-trometer in the usual position for surface analysis. It isrecommended that the test specimen be a metallic foil withlateral dimensions larger than the dimensions of the field ofview of the electron energy analyzer. The test specimen shouldbe polycrystalline and have grain
24、dimensions much less thanthe expected spatial resolution of the analyzer or the width ofthe incident beam on the specimen in order to avoid artifactsdue to channeling or diffraction effects. The specimen surfaceshould be smooth and be free of scratches and similar defectsthat are observable with the
- 1.请仔细阅读文档,确保文档完整性,对于不预览、不比对内容而直接下载带来的问题本站不予受理。
- 2.下载的文档,不会出现我们的网址水印。
- 3、该文档所得收入(下载+内容+预览)归上传者、原创作者;如果您是本文档原作者,请点此认领!既往收益都归您。
下载文档到电脑,查找使用更方便
5000 积分 0人已下载
下载 | 加入VIP,交流精品资源 |
- 配套讲稿:
如PPT文件的首页显示word图标,表示该PPT已包含配套word讲稿。双击word图标可打开word文档。
- 特殊限制:
部分文档作品中含有的国旗、国徽等图片,仅作为作品整体效果示例展示,禁止商用。设计者仅对作品中独创性部分享有著作权。
- 关 键 词:
- ASTME12172005STANDARDPRACTICEFORDETERMINATIONOFTHESPECIMENAREACONTRIBUTINGTOTHEDETECTEDSIGNALINAUGERELECTRONSPECTROMETERSANDSOMEXRAYPHOTOELECTRONSPECTROMETPDF

链接地址:http://www.mydoc123.com/p-528190.html