《ASTM D5839-2015 3955 Standard Test Method for Trace Element Analysis of Hazardous Waste Fuel by Energy-Dispersive X-Ray Fluorescence Spectrometry《采用能量发散X射线荧光光谱法分析危险废油中微量元素的标准试验方法》.pdf》由会员分享,可在线阅读,更多相关《ASTM D5839-2015 3955 Standard Test Method for Trace Element Analysis of Hazardous Waste Fuel by Energy-Dispersive X-Ray Fluorescence Spectrometry《采用能量发散X射线荧光光谱法分析危险废油中微量元素的标准试验方法》.pdf(4页珍藏版)》请在麦多课文档分享上搜索。
1、Designation: D5839 15Standard Test Method forTrace Element Analysis of Hazardous Waste Fuel byEnergy-Dispersive X-Ray Fluorescence Spectrometry1This standard is issued under the fixed designation D5839; the number immediately following the designation indicates the year oforiginal adoption or, in th
2、e case of revision, the year of last revision. A number in parentheses indicates the year of last reapproval. Asuperscript epsilon () indicates an editorial change since the last revision or reapproval.1. Scope1.1 This test method applies to the determination of traceelement concentrations by energy
3、-dispersive X-ray fluores-cence (EDXRF) spectrometry in typical liquid hazardous wastefuels (LHWF) used by industrial furnaces.1.2 This test method has been used successfully on numer-ous samples of LHWF that are mixtures of solvents, oils,paints, and pigments for the determination of the followinge
4、lements: Ag, As, Ba, Cd, Cr, Hg, Ni, Pb, Sb, Se, and Tl.1.3 This test method also may be applicable to elements notlisted above and to the analysis of trace metals in organicliquids other than those used as LHWF.1.4 This standard does not purport to address all of thesafety concerns, if any, associa
5、ted with its use. It is theresponsibility of the user of this standard to establish appro-priate safety and health practices and determine the applica-bility of regulatory limitations prior to use.2. Referenced Documents2.1 ASTM Standards:2C982 Guide for Selecting Components for Energy-Dispersive X-
6、Ray Fluorescence (XRF) Systems (With-drawn 2008)33. Summary of Test Method3.1 A weighed portion of the sample is transferred to aporcelain evaporating dish and placed on a hot plate. Thesample is heated for 15 to 30 min at a temperature notexceeding 105C to evaporate highly volatile components. Thed
7、ish is removed from the hot plate and allowed to cool.Graphite powder is blended with the evaporated sample until ahomogeneous paste is produced and the sample weight isrecorded. The blended sample is inserted in a disposablesample cup and placed in the X-ray spectrometer for analysis.3.2 The K Spec
8、tral X-ray emission lines are used forAg,As,Ba, Cd, Cr, Ni, Sb, and Se.3.3 The L spectral X-ray emission lines are used for Hg, Pb,and Tl.4. Significance and Use4.1 The analysis of trace elements is often a regulatory andprocess specific requirement for facilities utilizing LHWF.With proper instrume
9、nt standardization, set-up, and qualitycontrol, this test method provides the user an accurate, rapid,nondestructive method for trace element determinations.5. Interferences5.1 Spectral Overlaps:5.1.1 Samples containing a mixture of elements often ex-hibit X-ray emission line overlap. Modern Si (Li)
10、 detectorsgenerally provide adequate resolution to minimize the effectsof spectrum overlaps on the analytical results of the LHWF. Inthose cases where direct emission line overlap exists, spectraldeconvolution methods extract corrected analyte emission lineintensities. Table 1 lists the significant
11、line overlaps observedfor the elements analyzed in LHWF. Follow the EDXRFmanufacturers recommendations concerning spectral decon-volution of the emission lines.NOTE 1Not all possible interferences are listed in Table 1. The LHWFsamples to be analyzed may have other emission line interferences notmen
12、tioned. Consult the manufacturers recommendations for optimumdeconvolution methods.5.2 Matrix Interferences:5.2.1 Large concentration variations of metal or matrixcomponents, or both, in LHWF samples can result in non-linear metal X-ray intensity response at increasing metalconcentrations. Untreated
13、 matrix interactions may have adeleterious effect on metal determination accuracy. Matrixinteractions, if exhibited by the LHWF samples, must beaccounted for by method calibration.1This guide is under the jurisdiction of ASTM Committee D34 on WasteManagement and is the direct responsibility of Subco
14、mmittee D34.01.06 onAnalytical Methods.Current edition approved May 1, 2015. Published May 2015. Originallyapproved in 1996. Last previous edition approved in 2006 as D5839 - 96 (2006),which was withdrawn in January 2015 and reinstated in May 2015. DOI:10.1520/D5839-15.2For referenced ASTM standards
15、, visit the ASTM website, www.astm.org, orcontact ASTM Customer Service at serviceastm.org. For Annual Book of ASTMStandards volume information, refer to the standards Document Summary page onthe ASTM website.3The last approved version of this historical standard is referenced onwww.astm.org.Copyrig
16、ht ASTM International, 100 Barr Harbor Drive, PO Box C700, West Conshohocken, PA 19428-2959. United States16. Apparatus6.1 Energy Dispersive X-Ray Spectrometer, capable of mea-suring the wavelengths listed in Table 1. Refer to Guide C982for system specifications.6.2 Hot Plate, with temperature contr
17、ol to 105C.6.3 Analytical Balance, capable of weighing to 0.001 g.6.4 Porcelain Evaporating Dishes, 70 to 150 mL capacity.7. Reagents and Materials7.1 Purity of ReagentsUse reagent grade chemicals in alltests. Unless otherwise indicated, all reagents will conform tothe specifications of the Committe
18、e on Analytical Reagents ofthe American Chemical Society where such specifications areavailable.4Other grades may be used, provided it is firstascertained that the reagent is of sufficient high purity to permitits use without lessening the accuracy of the determination.7.2 Use graphite powder, mixin
19、g grade, 44 m, 325 mesh.57.3 Use oil-based Ag, As, Ba, Cd, Cr, Hg, Ni, Pb, Sb and Sestandards, 10 to 10 000 mg/kg depending on the usersanalytical requirements. If the results of this test method are tobe used for compliance purposes, standards or a commercialsource must be traceable to NIST or othe
20、r certifying body.Quality control samples for analyses done for compliancepurposes may need to be prepared with standards from adifferent vendor or lot number.7.4 Use oil or solvent-based Tl standard, 10 to 1000 mg/kgdepending on the users analytical requirements. If the resultsof this test method a
21、re to be used for compliance purposes,standards or a commercial source must be traceable to NIST orother certifying body.67.5 Use paraffinic base oil.7.6 Use sample cups, vented or unvented.7.7 Use thin-film sample support.NOTE 2The user should select a thin-film support that provides formaximum tra
22、nsmittance and is resistant to typical components in LHWF.The thin-film supports used in the development of this test method were4 m Prolene.8. Sampling8.1 Because stratification or layering of liquid samples ispossible, the laboratory sample should be thoroughly mixed byshaking prior to withdrawing
23、 a portion for testing, or use asampling method that assures all portions of the sample arerepresented.9. Preparation of Apparatus9.1 Follow the manufacturers instructions for set-up,conditioning, preparation, and maintenance of the XRF spec-trometer.9.2 Obtain reference spectra from pure element st
24、andardsfor all analytes and interelement correction metals.9.3 Address spectral interferences, as listed in 5.1.1,inaccordance with manufacturers recommendations.10. Calibration and Standardization10.1 Calibrate the spectrometer to an appropriate referenceelement at the minimum frequency specified b
- 1.请仔细阅读文档,确保文档完整性,对于不预览、不比对内容而直接下载带来的问题本站不予受理。
- 2.下载的文档,不会出现我们的网址水印。
- 3、该文档所得收入(下载+内容+预览)归上传者、原创作者;如果您是本文档原作者,请点此认领!既往收益都归您。
下载文档到电脑,查找使用更方便
5000 积分 0人已下载
下载 | 加入VIP,交流精品资源 |
- 配套讲稿:
如PPT文件的首页显示word图标,表示该PPT已包含配套word讲稿。双击word图标可打开word文档。
- 特殊限制:
部分文档作品中含有的国旗、国徽等图片,仅作为作品整体效果示例展示,禁止商用。设计者仅对作品中独创性部分享有著作权。
- 关 键 词:
- ASTMD583920153955STANDARDTESTMETHODFORTRACEELEMENTANALYSISOFHAZARDOUSWASTEFUELBYENERGYDISPERSIVEXRAYFLUORESCENCESPECTROMETRY

链接地址:http://www.mydoc123.com/p-520832.html