ASTM F1841-97(2017) Standard Practice for Assessment of Hemolysis in Continuous Flow Blood Pumps.pdf
《ASTM F1841-97(2017) Standard Practice for Assessment of Hemolysis in Continuous Flow Blood Pumps.pdf》由会员分享,可在线阅读,更多相关《ASTM F1841-97(2017) Standard Practice for Assessment of Hemolysis in Continuous Flow Blood Pumps.pdf(5页珍藏版)》请在麦多课文档分享上搜索。
1、Designation: F1841 97 (Reapproved 2017)Standard Practice forAssessment of Hemolysis in Continuous Flow BloodPumps1This standard is issued under the fixed designation F1841; the number immediately following the designation indicates the year oforiginal adoption or, in the case of revision, the year o
2、f last revision. A number in parentheses indicates the year of last reapproval. Asuperscript epsilon () indicates an editorial change since the last revision or reapproval.INTRODUCTIONThe goal of blood pump development is to replace or supplement the function of the human heart.As a result, continuo
3、us flow blood pumps, including roller pumps and centrifugal pumps, arecommonly used in clinical extracorporeal circulation. They are used not only for cardiopulmonarybypass in routine cardiac surgery but also for ventricular assist, percutaneous cardiopulmonarysupport, and extracorporeal membrane ox
4、ygenation.Many investigators have attempted to develop an atraumatic blood pump. Hemolysis is one of themost important parameters of blood trauma induced by blood pumps. However, comparative in vitroevaluation of the reported results of hemolysis are difficult due to the lack of uniformity of the te
5、stmethods employed. Thus, it is necessary to standardize the method of performing in vitro hemolysistests for the evaluation of continuous flow blood pumps.1. Scope1.1 This practice covers a protocol for the assessment of thehemolytic properties of continuous flow blood pumps used inextracorporeal o
6、r implantable circulatory assist. An assessmentis made based on the pumps effects on the erythrocytes overa certain period of time. For this assessment, a recirculation testis performed with a pump for 6 h.1.2 The values stated in either SI units or inch-pound unitsare to be regarded separately as s
7、tandard. The values stated ineach system may not be exact equivalents; therefore, eachsystem shall be used independently of the other. Combiningvalues from the two systems may result in non-conformancewith the standard.1.3 This standard does not purport to address all of thesafety concerns, if any,
8、associated with its use. It is theresponsibility of the user of this standard to establish appro-priate safety and health practices and determine the applica-bility of regulatory limitations prior to use.1.4 This international standard was developed in accor-dance with internationally recognized pri
9、nciples on standard-ization established in the Decision on Principles for theDevelopment of International Standards, Guides and Recom-mendations issued by the World Trade Organization TechnicalBarriers to Trade (TBT) Committee.2. Referenced Documents2.1 ASTM Standards:2F1830 Practice for Selection o
10、f Blood for in vitro Evaluationof Blood Pumps3. Terminology3.1 Definitions:3.1.1 continuous flow blood pumpa blood pump thatproduces continuous blood flow due to its rotary motion.3.1.2 free plasma hemoglobinthe amount of hemoglobin(iron or heme-containing protein) in plasma.3.1.3 hemolysisdamage to
11、 erythrocytes resulting in theliberation of hemoglobin into the plasma.3.1.4 Index of Hemolysis3.1.4.1 normalized index of hemolysisadded grams ofplasma free hemoglobin per 100 L of blood pumped, correctedfor plasma volume using hematocrit and normalized by flowrate and circulation time.3.1.4.2 norm
12、alized milligram index of hemolysisnormalized index of hemolysis expressed by milligram valueof free plasma hemoglobin.1This practice is under the jurisdiction ofASTM Committee F04 on Medical andSurgical Materials and Devices and is the direct responsibility of SubcommitteeF04.30 onCardiovascular St
13、andards.Current edition approved Sept. 1, 2017. Published September 2017. Originallyapproved in 1997. Last previous edition approved in 2013 as F1841 97 (2013).DOI: 10.1520/F1841-97R17.2For referenced ASTM standards, visit the ASTM website, www.astm.org, orcontact ASTM Customer Service at serviceast
14、m.org. For Annual Book of ASTMStandards volume information, refer to the standards Document Summary page onthe ASTM website.Copyright ASTM International, 100 Barr Harbor Drive, PO Box C700, West Conshohocken, PA 19428-2959. United StatesThis international standard was developed in accordance with in
15、ternationally recognized principles on standardization established in the Decision on Principles for theDevelopment of International Standards, Guides and Recommendations issued by the World Trade Organization Technical Barriers to Trade (TBT) Committee.13.1.4.3 modified index of hemolysismass of he
16、moglobinreleased into plasma normalized by the total amount ofhemoglobin pumped through the loop.4. Formulas4.1 Normalized Index of Hemolysis (N.I.H.) (1, 2, 3, 4)3:N.I.H. g/100l 5 freeHb 3V 3100 2 Ht1003100Q 3T(1)free Hb = increase of plasma free hemoglobin concentra-tion (g/L) over the sampling ti
17、me interval,where:V = circuit volume (L),Q = flow rate (L/min),Ht = hematocrit (%), andT = sampling time interval (min).4.2 Normalized Milligram Index of Hemolysis. (mg.N.I.H.)(2, 3, 4):2mg.N.I.H.mg/100l 5 freeHb 3V 3100 2 Ht1003100Q 3T(2)4.3 Modified Index of Hemolysis (M.I.H.):4.3.1 Modified index
18、 of hemolysis (M.I.H.) (5, 6) that canbe written with no units or as (milligram of hemoglobinreleased into plasma/mg of total hemoglobin pumped throughthe loop):M.I.H. 5 freeHb 3V 3100 2 Ht1003106Q 3T 3Hb(3)where:Hb = total blood hemoglobin concentration at timezero (mg/L), andfree Hb = increase of
19、plasma free hemoglobin concentra-tion (mg/L) over the sampling time interval.4.3.2 Among these indices, M.I.H. is recommended as anindex to express the degree of hemolysis caused by a bloodpump in a recirculating system. N.I.H. was proposed to accountfor the plasma volume based on the hemotocrit. Re
20、centdevelopment of less hemolytic blood pumps has since made itconvenient to use mg. N.I.H. rather than N.I.H. However, boththe N.I.H. and the mg N.I.H. vary with hematocrit of the blood(6). M.I.H. is the recommended index to express the degree ofhemolysis caused by a blood pump in a recirculating s
21、ystem.The M.I.H. equation corrects for differences in blood hemo-globin concentration and hematocrit directly (5).4.4 Testing BloodBecause the level of trauma-inducedhemolysis is different based on the source of blood, it isnecessary to identify the source of blood and its respectiveindex of hemolys
22、is. Human, bovine, or porcine blood arerecommended as the primary sources of testing blood (seePractice F1830). It is preferable that the blood collected at astandard slaughter house not be used due to the risk of beingcontaminated with fluids other than blood, unless the blood isobtained by control
23、led venipuncture. Although animal blood isused in the development stage of a pump, it is suggested thatpre-clinical evaluation tests be repeated with human blood.5. Summary of Practice5.1 BloodThe blood is obtained from human volunteers,cattle or pigs having normal body temperatures, no physicalsign
24、s of disease, including diarrhea or rhinorrhea, and anacceptable range of hemotological profiles. The blood shouldbe collected by vascular puncture using a needle (14G orlarger) and collected into the standard 5002000 mL bagscontaining citrate phospate dextrose adenine (CPDA-1) anti-coagulant soluti
- 1.请仔细阅读文档,确保文档完整性,对于不预览、不比对内容而直接下载带来的问题本站不予受理。
- 2.下载的文档,不会出现我们的网址水印。
- 3、该文档所得收入(下载+内容+预览)归上传者、原创作者;如果您是本文档原作者,请点此认领!既往收益都归您。
下载文档到电脑,查找使用更方便
5000 积分 0人已下载
下载 | 加入VIP,交流精品资源 |
- 配套讲稿:
如PPT文件的首页显示word图标,表示该PPT已包含配套word讲稿。双击word图标可打开word文档。
- 特殊限制:
部分文档作品中含有的国旗、国徽等图片,仅作为作品整体效果示例展示,禁止商用。设计者仅对作品中独创性部分享有著作权。
- 关 键 词:
- ASTM F1841 97 2017 Standard Practice for Assessment of Hemolysis in Continuous Flow Blood Pumps

链接地址:http://www.mydoc123.com/p-287170.html