1、COMMERCIAL ITEM DESCRIPTION SUPPORT, CERVICAL The General Services Administration has authorized the use of this Commercial Item Description. This Commercial Item Description covers one type and four sizes of a two-piece, plastic and foam, radiolucent, cervical extrication collar, suitable for immob
2、ilization and maintenance of the cervical spine in a neutral position in order to prevent spinal cord injury during patient extrication. Salient characteristics: The support provided shall be in accordance with the size specified below: Size Description 1 Pediatric size, two-piece, cervical extricat
3、ion support, fits neck circumference 8-16 inches, height 2-1/2 inches. 2 Small size, two-piece, cervical extrication support, fits neck circumference 10-20 inches, height 4 inches. 3 Medium size, two-piece, cervical extrication support, fits neck circumference 12-22 inches, height 4-3/4 inches. 4 La
4、rge size, two-piece, cervical extrication support, fits neck circumference 14-24 inches, height 5-1/2 inches . General. The cervical support collar shall be two-pieces made of lightweight, flexible, 1/8 inch thick low density polyethylene, padded on the interior with 1/4 inch thick charcoal-colored,
5、 closed-cell foam padding to facilitate cleaning and promote reuse. The foam padding shall extend AMSC N/A FSC 6515 DISTRIBUTION STATEMENT A. Approved for public release; distribution is unlimited. A-A-54215 65 777777Li 0056514 8 I A-A-5 4 2 15 31 July 1990 Provided by IHSNot for ResaleNo reproducti
6、on or networking permitted without license from IHS-,-,-A- A-54215 e past the borders of the polyethylene on all parts of the collar. The foam shall be held to the rigid polyethylene by means of 5/8 inch diameter nylon rivets. The two-pieces of the cervical collar shall be adjustably coupled togethe
7、r by means of woven nylon hook and pile closures, located on both sides of the neck. The anterior portion of the collar shall consist of two polyethylene panels. This anterior piece shall have a diamond shaped tracheal hole, a minimum of 2-1/2 inches in length and 171/2 inches in height which may be
8、 used for access to monitor carotid pulse or for airway management. Each anterior piece shall have a semi-circular, polyethylene chin piece designed to fold down flat for easy, compact storage. A strap of woven nylon pile shall extend from the side of the anterior panel and shall be used to hold the
9、 front piece in position while the posterior piece is applied. A strip of woven nylon hook shall extend across the length of the front panel. The posterior portion of the collar shall be composed of two preformed polyethylene pieces riveted together in a minimum of three places. The smaller piece of
10、 the posterior panel shall be fully padded. The larger piece shall be lined along the bottom and sides with padding. Two pieces (straps) of woven nylon pile shall be riveted to the posterior panel on either side and shall be designed to attach the anterior piece to the posterior piece. When coupled
11、together, the support shall immobilize the cervical area by enclosing the neck, mandible, occiput, and the chin. The low density polyethylene construction shall permit the collar to conform to the patient. The collar shall provide rigid flexion, extension, axial and rotational immobilization. The ce
12、rvical support shall maintain the spine in a neutral orientation during extrication and transport to prevent cervical spinal cord injury. The collar shall have clinical studies documenting capacity to immobilize. The cervical support shall be packaged flat for ease of storage and may be reused. Each
13、 cervical support shall be furnished with an instruction sheet illustrating the sizing and application of the collar and describing the cleaning instructions. 2 Provided by IHSNot for ResaleNo reproduction or networking permitted without license from IHS-,-,-Workmanship. The cervical supports shall
14、be free from defects which detract from their appearance or impair their serviceability. specified, constitutes one unit. Each individual package shall come with “Directions for Use“. Unit. Each (EA). One individual package cervical support, as Quality Assurance Provisions. the contract or purchase
15、order, the contractor is responsible for the performance of all inspection requirements as specified herein. Except as otherwise specified in the contract or purchase order, the contractor may use his own or any facilities suitable for the performance of the inspection requirements specified herein,
16、 unless disapproved by the Government. The Government reserves the right to perform any of the inspections set forth in the specification where such inspections are deemed necessary to assure supplies and services conform to prescribed requirements. Responsibility for inspection. Unless otherwise sp
- 1.请仔细阅读文档,确保文档完整性,对于不预览、不比对内容而直接下载带来的问题本站不予受理。
- 2.下载的文档,不会出现我们的网址水印。
- 3、该文档所得收入(下载+内容+预览)归上传者、原创作者;如果您是本文档原作者,请点此认领!既往收益都归您。
下载文档到电脑,查找使用更方便
10000 积分 0人已下载
下载 | 加入VIP,交流精品资源 |
- 配套讲稿:
如PPT文件的首页显示word图标,表示该PPT已包含配套word讲稿。双击word图标可打开word文档。
- 特殊限制:
部分文档作品中含有的国旗、国徽等图片,仅作为作品整体效果示例展示,禁止商用。设计者仅对作品中独创性部分享有著作权。
- 关 键 词:
- DODAA542151990SUPPORTCERVICAL 子宫颈 支撑 PDF