BS ISO 16129-2012 Surface chemical analysis X-ray photoelectron spectroscopy Procedures for assessing the day-to-day performance of an X-ray photoelectron spectrometer《表面化学分析 X射线光电.pdf
《BS ISO 16129-2012 Surface chemical analysis X-ray photoelectron spectroscopy Procedures for assessing the day-to-day performance of an X-ray photoelectron spectrometer《表面化学分析 X射线光电.pdf》由会员分享,可在线阅读,更多相关《BS ISO 16129-2012 Surface chemical analysis X-ray photoelectron spectroscopy Procedures for assessing the day-to-day performance of an X-ray photoelectron spectrometer《表面化学分析 X射线光电.pdf(30页珍藏版)》请在麦多课文档分享上搜索。
1、raising standards worldwideNO COPYING WITHOUT BSI PERMISSION EXCEPT AS PERMITTED BY COPYRIGHT LAWBSI Standards PublicationBS ISO 16129:2012Surface chemical analysis X-ray photoelectron spectroscopy Procedures for assessing the day-to-day performance of an X-ray photoelectron spectrometerBS ISO 16129
2、:2012 BRITISH STANDARDNational forewordThis British Standard is the UK implementation of ISO 16129:2012. The UK participation in its preparation was entrusted to T e c h n i c a l Committee CII/60, Surface chemical analysis.A list of organizations represented on this committee can be obtained on req
3、uest to its secretary.This publication does not purport to include all the necessary provisions of a contract. Users are responsible for its correct application. The British Standards Institution 2012. Published by BSI Standards Limited 2012.ISBN 978 0 580 72060 4 ICS 71.040.40 Compliance with a Bri
4、tish Standard cannot confer immunity from legal obligations.This British Standard was published under the authority of the Standards Policy and Strategy Committee on 31 December 2012.Amendments issued since publicationDate T e x t a f f e c t e dBS ISO 16129:2012 ISO 2012Surface chemical analysis X-
5、ray photoelectron spectroscopy Procedures for assessing the day-to-day performance of an X-ray photoelectron spectrometerAnalyse chimique des surfaces Spectroscopie de photolectrons X Modes opratoires dvaluation de la performance au jour le jour dun spectromtre de photolectrons XINTERNATIONAL STANDA
6、RDISO16129First edition2012-04-15Reference numberISO 16129:2012(E)BS ISO 16129:2012ISO 16129:2012(E)ii ISO 2012 All rights reservedCOPYRIGHT PROTECTED DOCUMENT ISO 2012All rights reserved. Unless otherwise specified, no part of this publication may be reproduced or utilized in any form or by any mea
7、ns, electronic or mechanical, including photocopying and microfilm, without permission in writing from either ISO at the address below or ISOs member body in the country of the requester.ISO copyright officeCase postale 56 CH-1211 Geneva 20Tel. + 41 22 749 01 11Fax + 41 22 749 09 47E-mail copyrighti
8、so.orgWeb www.iso.orgPublished in SwitzerlandBS ISO 16129:2012ISO 16129:2012(E) ISO 2012 All rights reserved iiiContents PageForeword ivIntroduction v1 Scope 12 Terms and definitions . 13 Initial approach 14 Initial instrument calibration, alignment and assessment . 15 Test specimen selection 25.1 G
9、eneral information 25.2 The conductive specimen . 25.3 The non-conductive specimen 25.4 Specimen for assessing the X-ray source 46 Collection of reference data . 46.1 General information 46.2 Rapid test of the instrument using a conductive specimen 46.3 Rapid test of the instrument using a non-condu
10、ctive specimen . 86.4 Rapid test of the X-ray source using a phosphor specimen . 86.5 Rapid test of the X-ray source using a uniform conductive specimen 87 Collection of subsequent performance data . 98 Analysis of the performance data . 98.1 General information 98.2 Survey spectrum . 98.3 High-reso
11、lution spectrum from the conductive specimen 108.4 High-resolution spectrum from the non-conductive specimen 108.5 Images from the phosphor specimen 108.6 Images from the uniform conductive specimen .108.7 Spectrum ratios .109 Control charts .15Bibliography .18BS ISO 16129:2012ISO 16129:2012(E)Forew
12、ordISO (the International Organization for Standardization) is a worldwide federation of national standards bodies (ISO member bodies). The work of preparing International Standards is normally carried out through ISO technical committees. Each member body interested in a subject for which a technic
13、al committee has been established has the right to be represented on that committee. International organizations, governmental and non-governmental, in liaison with ISO, also take part in the work. ISO collaborates closely with the International Electrotechnical Commission (IEC) on all matters of el
14、ectrotechnical standardization.International Standards are drafted in accordance with the rules given in the ISO/IEC Directives, Part 2.The main task of technical committees is to prepare International Standards. Draft International Standards adopted by the technical committees are circulated to the
15、 member bodies for voting. Publication as an International Standard requires approval by at least 75 % of the member bodies casting a vote.Attention is drawn to the possibility that some of the elements of this document may be the subject of patent rights. ISO shall not be held responsible for ident
16、ifying any or all such patent rights.ISO 16129 was prepared by Technical Committee ISO/TC 201, Surface chemical analysis, Subcommittee SC 7, Electron spectroscopies.iv ISO 2012 All rights reservedBS ISO 16129:2012ISO 16129:2012(E)IntroductionXPS instruments are complex, and unsatisfactory performanc
17、e is not always obvious to an operator. It is therefore necessary to provide a test for the correct operation of the system that can be performed regularly and frequently without interfering excessively with the normal work of the laboratory.A full diagnostic test can require many hours or even days
18、; such a test can be appropriate only when the instrument is known to have a fault that needs to be remedied or following a major maintenance procedure. Data acquired before a problem is uncovered become suspect if the spectrometer has not been routinely tested, leading to a loss of confidence in th
19、ose data. If a regular check of the instrument is made, changes in performance can be monitored and corrective action taken in good time to ensure that the data supplied are fit-for-purpose. In the event that a serious fault is uncovered, then only the data since the last check can be in doubt and n
20、eed to be repeated.The purpose of this document is to provide a user with a procedure which is not excessively time-consuming so that it is capable of being completed on a regular and frequent basis, daily if required. The user will then gain an awareness of the current characteristics of the instru
21、ment so that a decision can be made as to whether or not a more complete and time-consuming action is required to return the instrument to a satisfactory level of performance.This procedure is intended to be applied to an XPS instrument that has been correctly calibrated and aligned in accordance wi
22、th ISO standards or manufacturers instructions. It is designed to highlight aspects of the instruments characteristics that differ significantly from those that were measured immediately following the calibration procedure. The procedure does not show how the instrument can be returned to its origin
23、al state. Instead, it guides the user to possible areas of concern. The procedure provides data that can be used in control charts, allowing trends to be observed and acted upon before data quality deteriorates to an unacceptable level for the needs of the analyst. ISO 2012 All rights reserved vBS I
24、SO 16129:2012BS ISO 16129:2012Surface chemical analysis X-ray photoelectron spectroscopy Procedures for assessing the day-to-day performance of an X-ray photoelectron spectrometer1 ScopeThis International Standard is designed to allow the user to assess, on a regular basis, several key parameters of
- 1.请仔细阅读文档,确保文档完整性,对于不预览、不比对内容而直接下载带来的问题本站不予受理。
- 2.下载的文档,不会出现我们的网址水印。
- 3、该文档所得收入(下载+内容+预览)归上传者、原创作者;如果您是本文档原作者,请点此认领!既往收益都归您。
下载文档到电脑,查找使用更方便
10000 积分 0人已下载
下载 | 加入VIP,交流精品资源 |
- 配套讲稿:
如PPT文件的首页显示word图标,表示该PPT已包含配套word讲稿。双击word图标可打开word文档。
- 特殊限制:
部分文档作品中含有的国旗、国徽等图片,仅作为作品整体效果示例展示,禁止商用。设计者仅对作品中独创性部分享有著作权。
- 关 键 词:
- BSISO161292012SURFACECHEMICALANALYSISXRAYPHOTOELECTRONSPECTROSCOPYPROCEDURESFORASSESSINGTHEDAYTODAYPERFORMANCEOFANXRAYPHOTOELECTRONSPECTROMETER

链接地址:http://www.mydoc123.com/p-584948.html