ASTM F3172-2015 Standard Guide for Design Verification Device Size and Sample Size Selection for Endovascular Devices《血管内器械的设计验证装置尺寸和样品尺寸选择的标准指南》.pdf
《ASTM F3172-2015 Standard Guide for Design Verification Device Size and Sample Size Selection for Endovascular Devices《血管内器械的设计验证装置尺寸和样品尺寸选择的标准指南》.pdf》由会员分享,可在线阅读,更多相关《ASTM F3172-2015 Standard Guide for Design Verification Device Size and Sample Size Selection for Endovascular Devices《血管内器械的设计验证装置尺寸和样品尺寸选择的标准指南》.pdf(14页珍藏版)》请在麦多课文档分享上搜索。
1、Designation: F3172 15Standard Guide forDesign Verification Device Size and Sample Size Selectionfor Endovascular Devices1This standard is issued under the fixed designation F3172; the number immediately following the designation indicates the year oforiginal adoption or, in the case of revision, the
2、 year of last revision. A number in parentheses indicates the year of last reapproval. Asuperscript epsilon () indicates an editorial change since the last revision or reapproval.1. Scope1.1 This guide provides guidance for selecting an appropri-ate device size(s) and determining an appropriate samp
3、lesize(s) (that is, number of samples) for design verificationtesting of endovascular devices. A methodology is presented todetermine which device size(s) should be selected for testing toverify the device design adequately for each design inputrequirement (that is, test characteristic). Additionall
4、y, differentstatistical approaches are presented and discussed to help guidethe developer to determine and justify sample size(s) for thedesign input requirement being verified. Alternate methodolo-gies for determining device size selection and sample sizeselection may be acceptable for design verif
5、ication.1.2 This guide applies to physical design verification test-ing. This guide addresses in-vitro testing; in-vivo/animal stud-ies are outside the scope of this guide. This guide does notdirectly address design validation; however, the methodologiespresented may be applicable to in-vitro design
6、 validationtesting. Guidance for sampling related to computational simu-lation (for example, sensitivity analysis and tolerance analysis)is not provided. Guidance for using models, such as design ofexperiments (DOE), for design verification testing is notprovided. This guide does not address samplin
7、g across multiplemanufacturing lots as this is typically done as process valida-tion. Special considerations are to be given to certain tests suchas fatigue (see Practice E739) and shelf life testing (see Section8).1.3 Regulatory guidance may exist for endovascular devicesthat should be considered f
8、or design verification device sizeand sample size selection.1.4 UnitsThe values stated in SI units are to be regardedas the standard. No other units of measurement are included inthis standard.1.5 This standard does not purport to address all of thesafety concerns, if any, associated with its use. I
9、t is theresponsibility of the user of this standard to establish appro-priate safety and health practices and determine the applica-bility of regulatory limitations prior to use.2. Referenced Documents2.1 ASTM Standards:2E739 Practice for StatisticalAnalysis of Linear or LinearizedStress-Life (S-N)
10、and Strain-Life (-N) Fatigue DataF2914 Guide for Identification of Shelf-life Test Attributesfor Endovascular Devices2.2 ISO Standards:3ISO 14971:2012 Medical devicesApplication of risk man-agement to medical devices3. Terminology3.1 Definitions:3.1.1 attribute data, ndata that identify the presence
11、 orabsence of a characteristic (for example, good/bad or pass/fail).3.1.2 design input requirements, nphysical and perfor-mance requirements of a device that are used as a basis fordevice design (typically defined as test characteristics such asballoon burst pressure, shaft tensile strength, and so
12、forth).3.1.3 design output, nfeatures of the device (that is,dimensions, materials, and so forth) that define the design andmake it capable of achieving design input requirements.3.1.4 design subgroup, nset defined by the device sizeswithin the device matrix in which the essential design outputsdo n
13、ot vary for a specified design input requirement (that is,device sizes that share the same design for a specified designinput requirement).3.1.5 design validation, nestablishing by objective evi-dence that the device conforms to defined user needs andintended use(s).1This guide is under the jurisdic
14、tion of ASTM Committee F04 on Medical andSurgical Materials and Devices and is the direct responsibility of SubcommitteeF04.30 on Cardiovascular Standards.Current edition approved Dec. 15, 2015. Published February 2016. DOI:10.1520/F317215.2For referenced ASTM standards, visit the ASTM website, www.
15、astm.org, orcontact ASTM Customer Service at serviceastm.org. For Annual Book of ASTMStandards volume information, refer to the standards Document Summary page onthe ASTM website.3Available from American National Standards Institute (ANSI), 25 W. 43rd St.,4th Floor, New York, NY 10036, http:/www.ans
16、i.org.Copyright ASTM International, 100 Barr Harbor Drive, PO Box C700, West Conshohocken, PA 19428-2959. United States13.1.6 design verification, nconfirmation by examinationand provision of objective evidence that the device design(design output) fulfills the specified requirements (designinput).3
17、.1.7 device matrix, nentire range of available models/sizes for the device family.3.1.8 device size, nindividual model/size (for example, 6mm diameter by 25 mm length balloon on 135 cm lengthcatheter or a 6Fr 100 cm length guide catheter).3.1.9 endovascular device, ndevice used to treat vascularcond
18、itions from within the vessel.3.1.10 essential design output, EDO, ndesign feature(s) orcharacteristic(s) of the device that affects its ability to achievethe design input requirements (that is, design output(s) that hasa relevant effect on the test results).3.1.11 process validation, nestablishment
19、 by objectiveevidence that a process consistently produces a result or deviceachieving its predetermined requirements.3.1.12 safety factor, nratio of the device performance tothe specification requirement (for example, how much strongerthe device is than it needs to be to meet its specificationrequi
20、rement).3.1.13 sample size, nquantity of individual specimens of adevice tested.3.1.14 variables data, ndata that measure the numericalmagnitude of a characteristic (how good/how bad).4. Significance and Use4.1 The purpose of this guide is to provide guidance forselecting appropriate device size(s)
21、and determining appropri-ate sample size(s) for design verification of endovasculardevices. The device size(s) and sample size(s) for each designinput requirement should be determined before testing. Thedevice size(s) selected for verification testing should establishthat the entire device matrix is
22、 able to achieve the design inputrequirements. If testing is not performed on all device sizes,justification should be provided.4.2 The sample size justification and statistical proceduresused to analyze the data should be based on sound scientificprinciples and should be suitable for reaching a jus
23、tifiableconclusion. Insufficient sample size may lead to erroneousconclusions more often than desired.4.3 Guidance regarding methodologies for determining de-vice size selection and appropriate sample size is provided inSections 5 and 6.5. Selection of Device Size(s)5.1 Design input requirements are
24、 the physical and perfor-mance requirements of a device that are used as a basis fordevice design. Once the device design is defined, testing istypically performed to verify that the design input requirementsare met. The appropriate device size(s) for verification testingshould be determined for eac
- 1.请仔细阅读文档,确保文档完整性,对于不预览、不比对内容而直接下载带来的问题本站不予受理。
- 2.下载的文档,不会出现我们的网址水印。
- 3、该文档所得收入(下载+内容+预览)归上传者、原创作者;如果您是本文档原作者,请点此认领!既往收益都归您。
下载文档到电脑,查找使用更方便
5000 积分 0人已下载
下载 | 加入VIP,交流精品资源 |
- 配套讲稿:
如PPT文件的首页显示word图标,表示该PPT已包含配套word讲稿。双击word图标可打开word文档。
- 特殊限制:
部分文档作品中含有的国旗、国徽等图片,仅作为作品整体效果示例展示,禁止商用。设计者仅对作品中独创性部分享有著作权。
- 关 键 词:
- ASTMF31722015STANDARDGUIDEFORDESIGNVERIFICATIONDEVICESIZEANDSAMPLESIZESELECTIONFORENDOVASCULARDEVICES

链接地址:http://www.mydoc123.com/p-540103.html