1、Designation: F 1357 99 (Reapproved 2009)Standard Specification forArticulating Total Wrist Implants1This standard is issued under the fixed designation F 1357; the number immediately following the designation indicates the year oforiginal adoption or, in the case of revision, the year of last revisi
2、on. A number in parentheses indicates the year of last reapproval. Asuperscript epsilon () indicates an editorial change since the last revision or reapproval.1. Scope1.1 This specification describes total wrist implants, includ-ing solid ceramic implants, used to provide functioning articu-lation b
3、y employing radial carpal components.1.2 This specification excludes those implants with ceramic-coated or porous-coated surfaces, one piece elastomeric im-plants (with or without grommets), and those devices used forcustom applications.1.3 The values stated in SI units are to be regarded asstandard
4、. No other units of measurement are included in thisstandard.1.4 This standard does not purport to address all of thesafety concerns, if any, associated with its use. It is theresponsibility of the user of this standard to establish appro-priate safety and health practices and determine the applica-
5、bility of regulatory limitations prior to use.2. Referenced Documents2.1 ASTM Standards:2F67 Specification for Unalloyed Titanium, for SurgicalImplant Applications (UNS R50250, UNS R50400, UNSR50550, UNS R50700)F75 Specification for Cobalt-28 Chromium-6 MolybdenumAlloy Castings and Casting Alloy for
6、 Surgical Implants(UNS R30075)F86 Practice for Surface Preparation and Marking of Me-tallic Surgical ImplantsF90 Specification for Wrought Cobalt-20Chromium-15Tungsten-10Nickel Alloy for Surgical Implant Applica-tions (UNS R30605)F 136 Specification for Wrought Titanium-6Aluminum-4Vanadium ELI (Extr
7、a Low Interstitial) Alloy for SurgicalImplant Applications (UNS R56401)F 562 Specification for Wrought 35Cobalt-35Nickel-20Chromium-10Molybdenum Alloy for Surgical ImplantApplications (UNS R30035)F 563 Specification for Wrought Cobalt-20Nickel-20Chromium-3.5Molybdenum-3.5Tungsten-5Iron Alloyfor Surg
8、ical Implant Applications (UNS R30563)3F 601 Practice for Fluorescent Penetrant Inspection of Me-tallic Surgical ImplantsF 603 Specification for High-Purity Dense Aluminum Ox-ide for Medical Application3F 629 Practice for Radiography of Cast Metallic SurgicalImplantsF 648 Specification for Ultra-Hig
9、h-Molecular-Weight Poly-ethylene Powder and Fabricated Form for Surgical Im-plantsF 748 Practice for Selecting Generic Biological Test Meth-ods for Materials and DevicesF 799 Specification for Cobalt-28Chromium-6MolybdenumAlloy Forgings for Surgical Implants (UNS R31537,R31538, R31539)F 981 Practice
10、 for Assessment of Compatibility of Bioma-terials for Surgical Implants with Respect to Effect ofMaterials on Muscle and BoneF 983 Practice for Permanent Marking of Orthopaedic Im-plant ComponentsF 1108 Specification for Titanium-6Aluminum-4VanadiumAlloy Castings for Surgical Implants (UNS R56406)F
11、1537 Specification for Wrought Cobalt-28Chromium-6MolybdenumAlloys for Surgical Implants (UNS R31537,UNS R31538, and UNS R31539)3. Terminology3.1 Definitions:3.1.1 carpal componentarticulating member inserted intoor through the carpal bones.3.1.2 radial componentarticulating member inserted intothe
12、radius for articulation with the carpal component.3.1.3 total wrist replacementprosthetic parts substitutedfor the native opposing radial and carpal articulating surfaces.4. Classification4.1 ConstrainedA constrained joint prosthesis is used forjoint replacement and prevents dislocation of the prost
13、hesis in1This specification is under the jurisdiction of ASTM Committee F04 onMedical and Surgical Materials and Devices and is the direct responsibility ofSubcommittee F04.22 on Arthroplasty.Current edition approved June 15, 2009. Published July 2009. Originallyapproved in 1991. Last previous editi
14、on approved in 2004 as F 1357 - 99 (2004).2For referenced ASTM standards, visit the ASTM website, www.astm.org, orcontact ASTM Customer Service at serviceastm.org. For Annual Book of ASTMStandards volume information, refer to the standards Document Summary page onthe ASTM website.3Withdrawn. The las
15、t approved version of this historical standard is referencedon www.astm.org.1Copyright ASTM International, 100 Barr Harbor Drive, PO Box C700, West Conshohocken, PA 19428-2959, United States.more than one anatomical plane and consists of either a single,flexible, across-the-joint component, or more
16、than one compo-nent linked together or affined.4.2 Partially ConstrainedA semi-constrained joint pros-thesis is used for partial or total joint replacement and limitstranslation and rotation of the prosthesis in one or more planesvia the geometry of its articulating surfaces. It has no across-the-jo
17、int linkages.4.3 UnconstrainedAn unconstrained joint prosthesis isused for partial or total joint replacement and restricts mini-mally prosthesis movement in one or more planes. Its compo-nents have no across-the-joint linkages.5. Materials and Manufacture5.1 Proper material selection is necessary,
18、but insufficient toensure suitable function of a device.5.2 All metal implant components shall conform to one ofthe following specifications for implant materials: Specifica-tion F67, F75, F90, F 136, F 562, F 563 (nonbearing useonly), F 799, F 1108,orF 1537.5.3 All polymeric components shall confor
19、m to the follow-ing specification for implant materials: Specification F 648.5.4 All solid ceramic components shall conform to Specifi-cation F 603 for implant materials.5.5 BiocompatibilityArticulating implants shall be manu-factured from the materials listed in 5.2-5.4. Before implantscan be manuf
20、actured from other materials, their biocompatibil-ity will be considered suitable only if they produce anacceptable response after testing in accordance with PracticeF 981.5.6 When required for metallic implants, fluorescent pen-etrant inspection shall be performed in accordance with Prac-tice F 601
21、.5.7 When required for cast metallic implants, radiographyshall be performed in accordance with Practice F 629.6. Performance Requirements6.1 Polymeric Creep (Cold Flow)Ultra-high molecularweight polyethylene in implant form must conform to therequirements detailed in Specification F 648. When creep
22、occurs, it must not impair the function or stability of theinterface.6.2 Wear of Alternative MaterialsIt is important to under-stand the wear performance for articulating surfaces. Any newor different material couple should not exceed the wear rates ofthe following material couple when tested under
23、physiologicalconditions. The current wear couple is CoCrMo alloy (Speci-fication F75) against ultra high molecular weight polyethyl-ene. This is an industry wide referenced wear couple and isconsidered by some to be the minimum. It has been proven toprovide clinically acceptable results.NOTE 1In sit
24、uations where the pin-on-flat test may not be consideredappropriate, other test methods may be considered.6.3 Range of Motion of the Device Before ImplantationThe implant shall be evaluated to determine the maximumdorsiflexion, palmar flexion, radial deviation, and ulnar devia-tion possible before s
- 1.请仔细阅读文档,确保文档完整性,对于不预览、不比对内容而直接下载带来的问题本站不予受理。
- 2.下载的文档,不会出现我们的网址水印。
- 3、该文档所得收入(下载+内容+预览)归上传者、原创作者;如果您是本文档原作者,请点此认领!既往收益都归您。
下载文档到电脑,查找使用更方便
5000 积分 0人已下载
下载 | 加入VIP,交流精品资源 |
- 配套讲稿:
如PPT文件的首页显示word图标,表示该PPT已包含配套word讲稿。双击word图标可打开word文档。
- 特殊限制:
部分文档作品中含有的国旗、国徽等图片,仅作为作品整体效果示例展示,禁止商用。设计者仅对作品中独创性部分享有著作权。
- 关 键 词:
- ASTMF135719992009STANDARDSPECIFICATIONFORARTICULATINGTOTALWRISTIMPLANTSPDF

链接地址:http://www.mydoc123.com/p-534730.html