ASTM E968-2002(2014) 6189 Standard Practice for Heat Flow Calibration of Differential Scanning Calorimeters《差分扫描量热仪的热流量校准标准实施规程》.pdf
《ASTM E968-2002(2014) 6189 Standard Practice for Heat Flow Calibration of Differential Scanning Calorimeters《差分扫描量热仪的热流量校准标准实施规程》.pdf》由会员分享,可在线阅读,更多相关《ASTM E968-2002(2014) 6189 Standard Practice for Heat Flow Calibration of Differential Scanning Calorimeters《差分扫描量热仪的热流量校准标准实施规程》.pdf(5页珍藏版)》请在麦多课文档分享上搜索。
1、Designation: E968 02 (Reapproved 2014)Standard Practice forHeat Flow Calibration of Differential Scanning Calorimeters1This standard is issued under the fixed designation E968; the number immediately following the designation indicates the year oforiginal adoption or, in the case of revision, the ye
2、ar of last revision. A number in parentheses indicates the year of last reapproval. Asuperscript epsilon () indicates an editorial change since the last revision or reapproval.1. Scope1.1 This practice covers the heat flow calibration of differ-ential scanning calorimeters over the temperature range
3、from 130C to +800C.1.2 The values stated in SI units are to be regarded asstandard. No other units of measurement are included in thisstandard.1.3 Computer or electronic based instruments, techniques ordata manipulation equivalent to this practice may also be used.1.4 This standard does not purport
4、to address all of thesafety concerns, if any, associated with its use. It is theresponsibility of whoever uses this standard to consult andestablish appropriate safety and health practices and deter-mine the applicability of regulatory limitations prior to use.See also Section 7.2. Referenced Docume
5、nts2.1 ASTM Standards:2E473 Terminology Relating to Thermal Analysis and Rhe-ologyE793 Test Method for Enthalpies of Fusion and Crystalliza-tion by Differential Scanning CalorimetryE967 Test Method for Temperature Calibration of Differen-tial Scanning Calorimeters and Differential Thermal Ana-lyzers
6、E1142 Terminology Relating to Thermophysical Properties3. Terminology3.1 DefinitionsSpecific technical terms used in this prac-tice are in accordance with Terminologies E473 and E1142.3.2 Definitions of Terms Specific to This Standard:3.2.1 coeffcient of variation, na measure of relative pre-cision
7、calculated as the standard deviation of a series of valuesdivided by their average. It is usually multiplied by 100 andexpressed as a percentage.NOTE 1The term quantitative differential thermal analysis refers todifferential thermal analyzers that are designed to obtain quantitative orsemiquantitati
8、ve heat flow results. This procedure may also be used tocalibrate such apparatus.4. Summary of Practice4.1 Differential scanning calorimeters measure heat flow(power) into or out of a test specimen and provide a signaloutput proportional to this measurement. This signal often isrecorded as a functio
9、n of a second signal proportional totemperature or time. If this heat flow signal is integrated overtime, the resultant value is proportional to energy (or enthalpyor heat).To obtain the desired energy information, the observedinstrument response (such as the area under the curve scribed)must be mul
10、tiplied by a proportionality constant that convertsthe units of instrument output into the desired energy units.This proportionality constant is called the instrument calibra-tion coefficient (E). The value and dimensions (units) of Edepend upon the particular differential scanning calorimeterand re
11、cording system being used and, moreover, may vary withtemperature.4.2 This practice consists of calibrating the heat flowresponse of a differential scanning calorimeter (that is, deter-mining the calibration coefficient) by recording the meltingendotherm of a high-purity standard material (where the
12、 heat offusion is known to better than 61.5 % (rel) as a function oftime. The peak is then integrated (over time) to yield an areameasurement proportional to the enthalpy of melting of thestandard material.4.3 Calibration of the instrument is extended to tempera-tures other than that of the melting
13、point of the standardmaterial through the recording of the specific heat capacity ofa (second) standard material over the temperature range ofinterest. The ratio of the measured specific heat capacity at thetemperature of interest to that of the temperature of calibrationprovides an instrument calib
14、ration coefficient at the newtemperature.4.4 Once the calibration coefficient at a given temperature isdetermined, it may be used to determine the desired energyvalue associated with an enthalpic transition in an unknownspecimen at that temperature (see Test Method E793).1This practice is under the
15、jurisdiction of ASTM Committee E37 on ThermalMeasurements and is the direct responsibility of Subcommittee E37.01 on Calo-rimetry and Mass Loss.Current edition approved March 15, 2014. Published April 2014. Originallyapproved in 1983. Last previous edition approved in 2008 as E968 02 (2008).DOI: 10.
16、1520/E0968-02R14.2For referenced ASTM standards, visit the ASTM website, www.astm.org, orcontact ASTM Customer Service at serviceastm.org. For Annual Book of ASTMStandards volume information, refer to the standards Document Summary page onthe ASTM website.Copyright ASTM International, 100 Barr Harbo
17、r Drive, PO Box C700, West Conshohocken, PA 19428-2959. United States15. Significance and Use5.1 Differential scanning calorimetry is used to determinethe heat or enthalpy of transition. For this information to bemeaningful in an absolute sense, heat flow calibration of theapparatus or comparison of
18、 the resulting data to that of aknown standard is required.5.2 This practice is useful in calibrating the heat flow axis ofdifferential scanning calorimeters or quantitative differentialthermal analyzers for subsequent use in the measurement oftransition energies and specific heat capacities of unkn
19、owns.6. Apparatus6.1 Differential Scanning Calorimeter (DSC)The essentialinstrumentation required to provide the minimum differentialscanning calorimetric capability for this method includes:6.1.1 A DSC test chamber, composed of the following:6.1.1.1 A furnace(s) to provide uniform controlled heatin
20、g(cooling) of a specimen and reference to a constant temperatureor at a constant rate with the temperature range of 100 to600C.NOTE 2This temperature range may be extended to higher and lowertemperatures depending upon the capabilities of the apparatus.6.1.1.2 Atemperature sensor, to provide an indi
21、cation of thespecimen/furnace temperature to 60.01 K.6.1.1.3 A differential sensor, to detect a heat flow (power)difference between the specimen and reference equivalent to 1W.6.1.1.4 A means of sustaining a test chamber environment,of an inert purge gas at a purge gas rate of 10 to 100 mL/min6 5 mL
22、/min.NOTE 3Typically, 99.9+ % pure nitrogen, argon or helium areemployed when oxidation in air is a concern. Unless effects of moistureare to be studied, use of dry purge gas is recommended and is essential foroperation at subambient temperatures.6.1.2 A temperature controller, capable of executing
23、aspecific temperature program by operating the furnace(s)between selected temperature limits at a rate of temperaturechange of between 1 and 35 K/min constant to 61 % and at anisothermal temperature constant to 60.1 K.6.1.3 Arecording device, either digital or analog, capable ofrecording and display
24、ing the heat flow (DSC curve) signalversus temperature, displaying any fraction including thesignal noise.6.1.4 Containers, (pans, crucibles, vials, etc. and associatedlids), that are inert to the specimen and reference materials andthat are of suitable structural shape and integrity to contain thes
- 1.请仔细阅读文档,确保文档完整性,对于不预览、不比对内容而直接下载带来的问题本站不予受理。
- 2.下载的文档,不会出现我们的网址水印。
- 3、该文档所得收入(下载+内容+预览)归上传者、原创作者;如果您是本文档原作者,请点此认领!既往收益都归您。
下载文档到电脑,查找使用更方便
5000 积分 0人已下载
下载 | 加入VIP,交流精品资源 |
- 配套讲稿:
如PPT文件的首页显示word图标,表示该PPT已包含配套word讲稿。双击word图标可打开word文档。
- 特殊限制:
部分文档作品中含有的国旗、国徽等图片,仅作为作品整体效果示例展示,禁止商用。设计者仅对作品中独创性部分享有著作权。
- 关 键 词:
- ASTME968200220146189STANDARDPRACTICEFORHEATFLOWCALIBRATIONOFDIFFERENTIALSCANNINGCALORIMETERS 扫描 量热仪 流量

链接地址:http://www.mydoc123.com/p-533868.html