ASTM E968-2002 Standard Practice for Heat Flow Calibration of Differential Scanning Calorimeters《差分扫描量热计的热流量校准标准实施规程》.pdf
《ASTM E968-2002 Standard Practice for Heat Flow Calibration of Differential Scanning Calorimeters《差分扫描量热计的热流量校准标准实施规程》.pdf》由会员分享,可在线阅读,更多相关《ASTM E968-2002 Standard Practice for Heat Flow Calibration of Differential Scanning Calorimeters《差分扫描量热计的热流量校准标准实施规程》.pdf(5页珍藏版)》请在麦多课文档分享上搜索。
1、Designation: E 968 02Standard Practice forHeat Flow Calibration of Differential Scanning Calorimeters1This standard is issued under the fixed designation E 968; the number immediately following the designation indicates the year oforiginal adoption or, in the case of revision, the year of last revis
2、ion. A number in parentheses indicates the year of last reapproval. Asuperscript epsilon (e) indicates an editorial change since the last revision or reapproval.1. Scope1.1 This practice covers the heat flow calibration of differ-ential scanning calorimeters over the temperature rangefrom 130C to +
3、800C.1.2 Values given in SI units are to be regarded as thestandard.1.3 Computer or electronic based instruments, techniques ordata manipulation equivalent to this practice may also be used.1.4 This standard does not purport to address all of thesafety concerns, if any, associated with its use. It i
4、s theresponsibility of whoever uses this standard to consult andestablish appropriate safety and health practices and deter-mine the applicability of regulatory limitations prior to use.See also Section 7.2. Referenced Documents2.1 ASTM Standards:E 473 Terminology Relating to Thermal Analysis2E 793
5、Test Method for Heats of Fusion and Crystallizationby Differential Scanning Calorimetry2E 967 Practice for Temperature Calibration of DifferentialScanning Calorimeters and Differential Thermal Analyz-ers23. Terminology3.1 DefinitionsSpecific technical terms used in this prac-tice are in accordance w
6、ith Terminologies E 474 and E 1142.3.2 Definitions of Terms Specific to This Standard:3.2.1 coeffcient of variation, na measure of relativeprecision calculated as the standard deviation of a series ofvalues divided by their average. It is usually multiplied by 100and expressed as a percentage.NOTE 1
7、The term quantitative differential thermal analysis refers todifferential thermal analyzers that are designed to obtain quantitative orsemiquantitative heat flow results. This procedure may also be used tocalibrate such apparatus.4. Summary of Practice4.1 Differential scanning calorimeters measure h
8、eat flow1This practice is under the jurisdiction of ASTM Committee E37 on ThermalMeasurements and is the direct responsibility of Subcommittee E37.01 on ThermalAnalysis Methods.Current edition approved March 10, 2002. Published June 2002. Originallypublished as E 968 83. Last previous edition E 968
9、99.2Annual Book of ASTM Standards, Vol 14.02.TABLE 1 Sapphire (a Al2O3) Specific Heat CapacityATemperature,KSpecific HeatCapacity, J/gKTemperature, KSpecific HeatCapacity, J/gK140 0.2739 630 1.1184150 0.3133 640 1.1228160 0.3525 650 1.1272170 0.3912 660 1.1313180 0.4290 670 1.1353190 0.4659 680 1.13
10、93200 0.5014 690 1.1431210 0.5356 700 1.1467220 0.5684 710 1.1503230 0.5996 720 1.1538240 0.6294 730 1.1572250 0.6577 740 1.1604260 0.6846 750 1.1637270 0.7102 760 1.1667280 0.7344 770 1.1698290 0.7574 780 1.1727300 0.7792 790 1.1756310 0.7999 800 1.1784320 0.8194 810 1.1811330 0.8380 820 1.1839340
11、0.8555 830 1.1864350 0.8721 840 1.1890360 0.8878 850 1.1914370 0.9027 860 1.1939380 0.9168 870 1.1962390 0.9302 880 1.1986400 0.9429 890 1.2008410 0.9550 900 1.2031420 0.9665 910 1.2053430 0.9775 920 1.2074440 0.9879 930 1.2095450 0.9978 940 1.2115460 1.0073 950 1.2135470 1.0164 960 1.2155480 1.0250
12、 970 1.2174490 1.0332 980 1.2194500 1.0411 990 1.2212510 1.0486 1000 1.2230520 1.0559 1010 1.2249530 1.0628 1020 1.2266540 1.0694 1030 1.2284550 1.0758 1040 1.2301560 1.0819 1050 1.2318570 1.0877 1060 1.2335580 1.0934 1070 1.2351590 1.0988 1080 1.2367600 1.1039 1090 1.2383610 1.1090 1100 1.2400620 1
13、.1138AArcher, D.G., J. Phys. Chem. Ref. Data, Vol 22, No. 8, pp. 14411453 (1993).1Copyright ASTM International, 100 Barr Harbor Drive, PO Box C700, West Conshohocken, PA 19428-2959, United States.(power) into or out of a test specimen and provide a signaloutput proportional to this measurement. This
14、 signal often isrecorded as a function of a second signal proportional totemperature or time. If this heat flow signal is integrated overtime, the resultant value is proportional to energy (or enthalpyor heat). To obtain the desired energy information, the observedinstrument response (such as the ar
15、ea under the curve scribed)must be multiplied by a proportionality constant that convertsthe units of instrument output into the desired energy units.This proportionality constant is called the instrument calibra-tion coefficient ( E). The value and dimensions (units) of Edepend upon the particular
16、differential scanning calorimeterand recording system being used and, moreover, may vary withtemperature.4.2 This practice consists of calibrating the heat flowresponse of a differential scanning calorimeter (that is, deter-mining the calibration coefficient) by recording the meltingendotherm of a h
17、igh-purity standard material (where the heat offusion is known to better than 6 1.5 % (rel) as a function oftime. The peak is then integrated (over time) to yield an areameasurement proportional to the enthalpy of melting of thestandard material.4.3 Calibration of the instrument is extended to tempe
18、ra-tures other than that of the melting point of the standardmaterial through the recording of the specific heat capacity ofa (second) standard material over the temperature range ofinterest. The ratio of the measured specific heat capacity at thetemperature of interest to that of the temperature of
19、 calibrationprovides an instrument calibration coefficient at the newtemperature.4.4 Once the calibration coefficient at a given temperature isdetermined, it may be used to determine the desired energyvalue associated with an enthalpic transition in an unknownspecimen at that temperature (see Method
20、 E 793).5. Significance and Use5.1 Differential scanning calorimetry is used to determinethe heat or enthalpy of transition. For this information to bemeaningful in an absolute sense, heat flow calibration of theapparatus or comparison of the resulting data to that of aknown standard is required.5.2
21、 This practice is useful in calibrating the heat flow axis ofdifferential scanning calorimeters or quantitative differentialthermal analyzers for subsequent use in the measurement oftransition energies and specific heat capacities of unknowns.6. Apparatus6.1 Differential Scanning Calorimeter (DSC)Th
22、e essen-tial instrumentation required to provide the minimum differen-tial scanning calorimetric capability for this method includes:6.1.1 A DSC test chamber, composed of the following:6.1.1.1 A furnace(s) to provide uniform controlled heating(cooling) of a specimen and reference to a constant tempe
23、ratureor at a constant rate with the temperature range of 100 to600C.NOTE 2This temperature range may be extended to higher and lowertemperatures depending upon the capabilities of the apparatus.6.1.1.2 A temperature sensor, to provide an indication of thespecimen/furnace temperature to 6 0.01 K.6.1
24、.1.3 A differential sensor, to detect a heat flow (power)difference between the specimen and reference equivalent to 1W.6.1.1.4 A means of sustaining a test chamber environment,of an inert purge gas at a purge gas rate of 10 to 100 mL/min6 5 mL/min.NOTE 3Typically, 99.9+ % pure nitrogen, argon or he
- 1.请仔细阅读文档,确保文档完整性,对于不预览、不比对内容而直接下载带来的问题本站不予受理。
- 2.下载的文档,不会出现我们的网址水印。
- 3、该文档所得收入(下载+内容+预览)归上传者、原创作者;如果您是本文档原作者,请点此认领!既往收益都归您。
下载文档到电脑,查找使用更方便
5000 积分 0人已下载
下载 | 加入VIP,交流精品资源 |
- 配套讲稿:
如PPT文件的首页显示word图标,表示该PPT已包含配套word讲稿。双击word图标可打开word文档。
- 特殊限制:
部分文档作品中含有的国旗、国徽等图片,仅作为作品整体效果示例展示,禁止商用。设计者仅对作品中独创性部分享有著作权。
- 关 键 词:
- ASTME9682002STANDARDPRACTICEFORHEATFLOWCALIBRATIONOFDIFFERENTIALSCANNINGCALORIMETERS 扫描 量热计 流量 校准 标准

链接地址:http://www.mydoc123.com/p-533866.html