ASTM E967-2018 6250 Standard Test Method for Temperature Calibration of Differential Scanning Calorimeters and Differential Thermal Analyzers《差示扫描量热计和差示热分析仪温度校准的标准试验方法》.pdf
《ASTM E967-2018 6250 Standard Test Method for Temperature Calibration of Differential Scanning Calorimeters and Differential Thermal Analyzers《差示扫描量热计和差示热分析仪温度校准的标准试验方法》.pdf》由会员分享,可在线阅读,更多相关《ASTM E967-2018 6250 Standard Test Method for Temperature Calibration of Differential Scanning Calorimeters and Differential Thermal Analyzers《差示扫描量热计和差示热分析仪温度校准的标准试验方法》.pdf(5页珍藏版)》请在麦多课文档分享上搜索。
1、Designation: E967 18Standard Test Method forTemperature Calibration of Differential ScanningCalorimeters and Differential Thermal Analyzers1This standard is issued under the fixed designation E967; the number immediately following the designation indicates the year oforiginal adoption or, in the cas
2、e of revision, the year of last revision. A number in parentheses indicates the year of last reapproval. Asuperscript epsilon () indicates an editorial change since the last revision or reapproval.1. Scope1.1 This test method describes the temperature calibrationof differential thermal analyzers and
3、 differential scanningcalorimeters over the temperature range from 40Cto +2000C.1.2 The values stated in SI units are to be regarded asstandard. No other units of measurement are included in thisstandard.1.3 This test method is similar to ISO standard 113571.1.4 This standard does not purport to add
4、ress all of thesafety concerns, if any, associated with its use. It is theresponsibility of the user of this standard to establish appro-priate safety, health, and environmental practices and deter-mine the applicability of regulatory limitations prior to use.Specific precautionary statements are gi
5、ven in Section 7.1.5 This international standard was developed in accor-dance with internationally recognized principles on standard-ization established in the Decision on Principles for theDevelopment of International Standards, Guides and Recom-mendations issued by the World Trade Organization Tec
6、hnicalBarriers to Trade (TBT) Committee.2. Referenced Documents2.1 ASTM Standards:2E473 Terminology Relating to Thermal Analysis and Rhe-ologyE1142 Terminology Relating to Thermophysical Properties2.2 ISO Standards:3113571 Plastics-Differential Scanning Calorimetry (DSC)-Part 1: General Principles3.
7、 Terminology3.1 DefinitionsSpecific technical terms used in this testmethod are defined in Terminologies E473 and E1142.4. Summary of Test Method4.1 This test method consists of heating the calibrationmaterials at a controlled rate in a controlled atmospherethrough a region of known thermal transiti
8、on. The heat flowinto the calibration material or the difference of temperaturebetween the calibration material and a reference sample ismonitored and continuously recorded. A transition is markedby the absorption of energy by the specimen resulting in acorresponding endothermic peak in the heating
9、curve.NOTE 1Heat flow calibrations are sometimes determined in conjunc-tion with temperature calibration. Some differential scanning calorimeterspermit both heat flow and temperature calibrations to be obtained from thesame experimental procedure.5. Significance and Use5.1 Differential scanning calo
10、rimeters and differential ther-mal analyzers are used to determine the transition temperaturesof materials. For this information to be meaningful in anabsolute sense, temperature calibration of the apparatus andcomparison of the resulting data to that of known standardmaterials is required.5.2 This
11、test method is useful in calibrating the temperatureaxis of differential scanning calorimeters and differential ther-mal analyzers.6. Apparatus6.1 Apparatus shall be of either type listed below:6.1.1 Differential Scanning Calorimeter (DSC), capable ofheating a test specimen and a reference material
12、at a controlledrate and of automatically recording the differential heat flowbetween the sample and the reference material to the requiredsensitivity and precision.6.1.1.1 A Furnace(s), to provide uniform controlled heatingor cooling of a specimen and reference to a constant tempera-ture or at a con
13、stant rate within the applicable temperaturerange of this test method.6.1.1.2 A Temperature Sensor, to provide an indication ofthe specimen temperature.1This test method is under the jurisdiction ofASTM Committee E37 on ThermalMeasurements and is the direct responsibility of Subcommittee E37.01 on C
14、alo-rimetry and Mass Loss.Current edition approved March 15, 2018. Published March 2018. Originallyapproved in 1983. Last previous edition approved in 2014 as E967 08 (2014).DOI: 10.1520/E0967-18.2For referenced ASTM standards, visit the ASTM website, www.astm.org, orcontact ASTM Customer Service at
15、 serviceastm.org. For Annual Book of ASTMStandards volume information, refer to the standards Document Summary page onthe ASTM website.3Available from International Organization for Standardization (ISO), 1, ch. dela Voie-Creuse, CP 56, CH-1211 Geneva 20, Switzerland, http:/www.iso.org.Copyright AST
16、M International, 100 Barr Harbor Drive, PO Box C700, West Conshohocken, PA 19428-2959. United StatesThis international standard was developed in accordance with internationally recognized principles on standardization established in the Decision on Principles for theDevelopment of International Stan
17、dards, Guides and Recommendations issued by the World Trade Organization Technical Barriers to Trade (TBT) Committee.16.1.1.3 Differential Sensors, to detect a heat flow (power)difference between the specimen and reference.6.1.1.4 Test Chamber Environment, a means of sustaining atest chamber environ
18、ment of nitrogen or other inert purge gasat a purge rate of 10 to 50 mL/min.6.1.1.5 A Temperature Controller, capable of executing aspecific temperature program by operating the furnace(s)between selected temperature limits at a rate of temperaturechange of 10C/min.6.1.1.6 Data Collection Device, to
19、 provide a means ofacquiring, storing, and displaying measured or calculatedsignals, or both. The minimum output signals required for DSCare heat flow, temperature, and time.6.1.2 Differential Thermal Analyzer (DTA), capable of heat-ing a test specimen and reference material at a controlled rateand
20、of automatically recording the differential temperaturebetween sample and reference material both to the requiredsensitivity and precision.6.2 Containers (pans, crucibles, vials, lids, closures, seals,etc.), that are inert to the specimen and reference materials andthat are of suitable structural sh
21、ape and integrity to contain thespecimen and reference in accordance with the specific require-ments of this test method.6.3 Nitrogen, or other inert purge gas supply.6.4 A Balance, to weigh specimens or containers (pans,crucibles, vials, etc.), or both to 60.1 mg. The balance shouldhave a capacity
22、greater than 20 mg.7. Precautions7.1 Toxic or corrosive effluents, or both, may be releasedwhen heating some material and could be harmful to personneland to apparatus.7.2 This test method assumes linear temperature indication.Care must be taken in the application of this test method toensure that c
23、alibration points are taken sufficiently closetogether so that linear temperature indication may be approxi-mated. Linear temperature indications means that there exists alinear, or first order, dependence on the temperature determinedby the instruments temperature sensor on the true temperatureof t
24、he sample material in its container and that this relation isadequately expressed by Eq 1.8. Calibration Materials8.1 For the temperature range covered by manyapplications, the melting transition of 99.99 % pure materialsin Table 1 may be used for calibration.9. Procedure9.1 Two Point Calibration:9.
- 1.请仔细阅读文档,确保文档完整性,对于不预览、不比对内容而直接下载带来的问题本站不予受理。
- 2.下载的文档,不会出现我们的网址水印。
- 3、该文档所得收入(下载+内容+预览)归上传者、原创作者;如果您是本文档原作者,请点此认领!既往收益都归您。
下载文档到电脑,查找使用更方便
5000 积分 0人已下载
下载 | 加入VIP,交流精品资源 |
- 配套讲稿:
如PPT文件的首页显示word图标,表示该PPT已包含配套word讲稿。双击word图标可打开word文档。
- 特殊限制:
部分文档作品中含有的国旗、国徽等图片,仅作为作品整体效果示例展示,禁止商用。设计者仅对作品中独创性部分享有著作权。
- 关 键 词:
- ASTME96720186250STANDARDTESTMETHODFORTEMPERATURECALIBRATIONOFDIFFERENTIALSCANNINGCALORIMETERSANDDIFFERENTIALTHERMALANALYZERS

链接地址:http://www.mydoc123.com/p-533865.html