ASTM E579-2004(2015) 0980 Standard Test Method for Limit of Detection of Fluorescence of Quinine Sulfate in Solution《溶液中硫酸奎宁的荧光检测限值的标准试验方法》.pdf
《ASTM E579-2004(2015) 0980 Standard Test Method for Limit of Detection of Fluorescence of Quinine Sulfate in Solution《溶液中硫酸奎宁的荧光检测限值的标准试验方法》.pdf》由会员分享,可在线阅读,更多相关《ASTM E579-2004(2015) 0980 Standard Test Method for Limit of Detection of Fluorescence of Quinine Sulfate in Solution《溶液中硫酸奎宁的荧光检测限值的标准试验方法》.pdf(3页珍藏版)》请在麦多课文档分享上搜索。
1、Designation: E579 04 (Reapproved 2015)Standard Test Method forLimit of Detection of Fluorescence of Quinine Sulfate inSolution1This standard is issued under the fixed designation E579; the number immediately following the designation indicates the year oforiginal adoption or, in the case of revision
2、, the year of last revision. A number in parentheses indicates the year of last reapproval. Asuperscript epsilon () indicates an editorial change since the last revision or reapproval.1. Scope1.1 This test method employs the signal-to-noise ratio todetermine the sensitivity of a fluorescence measuri
3、ng system intesting for the limit of detection (LOD) of quinine sulfatedihydrate in solution. The results obtained with quinine sulfatedihydrate in solution are suitable for specifying instrumentperformance on samples having excitation and fluorescencebands wider than 10 nm at or near room temperatu
4、re.1.1.1 This test method is not intended to be used as (1)arigorous test of performance of instrumentation, or (2), tointercompare the quantitative performance of instruments ofdifferent design. Intercomparison of the LOD between instru-ments is commonly expressed as the ratio of the water Ramanpea
5、k intensity to the root-mean-square (rms) noise as measuredon a fluorometer using an excitation wavelength of 350 nmThis test method uses the excitation and emission peakwavelengths for quinine sulfate dihydrate in solution, whichare approximately 350 nm and 450 nm, respectively.1.2 This test method
6、 has been applied to fluorescence-measuring systems utilizing non-laser, low-energy excitationsources. There is no assurance that extremely intense illumi-nation will not cause photodecomposition2of the compoundsuggested in this test method. For this reason, it is recom-mended that this test method
7、not be indiscriminately employedwith high intensity light sources. This test method is notintended to determine minimum detectable amounts of othermaterials. If this test method is extended to employ otherchemical substances, the user should be aware of the possibil-ity that these other substances m
8、ay undergo decomposition oradsorption onto containers.1.3 A typical LOD for conventional fluorometers using thistest method is 1 ng of quinine sulfate per mL.1.4 The suggested shelf life of a 1 mg/mL stock solution ofquinine sulfate dihydrate is three months, when stored in thedark in a stoppered gl
9、ass bottle.1.5 The values stated in SI units are to be regarded asstandard. No other units of measurement are included in thisstandard.1.6 This standard does not purport to address all of thesafety problems, if any, associated with its use. It is theresponsibility of the user of this standard to est
10、ablish appro-priate safety and health practices and determine the applica-bility of regulatory limitations prior to use.2. Referenced Documents2.1 ASTM Standards:3E578 Test Method for Linearity of Fluorescence MeasuringSystems3. Summary of Test Method3.1 To measure the concentration corresponding to
11、 theLOD, the fluorescence intensity scale and gain on the detectorare adjusted such that noise observed with pure solvent in thesample cell is large enough to measure. The test solution is thendiluted until readings on both the test solution and pure solventcan be read at the same intensity, scale,
12、and instrument settings.The concentration corresponding to the limit of detection isthat at which the noise intensity, multiplied by three, is equal tothe signal intensity.3.2 This test for limit of detection requires an instrument tomeet the following conditions: stable, free of extraneous noise,el
13、ectrical pickup, and internal stray light. The sample spacemust be covered to exclude room light. The instrument shouldbe operated according to the manufacturers recommendations,or, if they are modified, the modifications must be appliedconsistently to the test for limit of detection and to the anal
14、ysisfor which the test is a requirement, so that levels of perfor-mance are comparable for both. All modifications must beincluded in the report outlined in Section 8.1This test method is under the jurisdiction of ASTM Committee E13 onMolecular Spectroscopy and Separation Science and is the direct r
15、esponsibility ofSubcommittee E13.01 on Ultra-Violet, Visible, and Luminescence Spectroscopy.Current edition approved May 1, 2015. Published June 2015. Originallypublished in 1976. Last previous edition approved in 2009 as E579 04 (2009).DOI: 10.1520/E0579-04R15.2Lukasiewicz, R. J., and Fitzgerald, J
16、. M., Analytical Chemistry, ANCHA, Vol45, 1973, p. 511.3For referenced ASTM standards, visit the ASTM website, www.astm.org, orcontact ASTM Customer Service at serviceastm.org. For Annual Book of ASTMStandards volume information, refer to the standards Document Summary page onthe ASTM website.Copyri
17、ght ASTM International, 100 Barr Harbor Drive, PO Box C700, West Conshohocken, PA 19428-2959. United States1NOTE 1To obtain the lowest reading (the best instrumental response)for the limit of detection of fluorescent material, a number of precautionsmust be taken. The quality, condition, and positio
18、n of the sample cell aremost important. The cell must be made of fused silica that does notfluoresce at the excitation wavelength and be free of scratches and marksthat scatter light into the fluorescence detection system. Only spectralgrade chemicals and solvents (including water) that do not fluor
19、esceshould be used.4Dilute solutions of quinine sulfate dihydrate should bemade, just before use, from concentrated stock solutions.All samples usedmust be maintained at the same temperature to obviate effects due totemperature fluctuations. The average temperature coefficient for fluores-cence inte
20、nsity in the temperature range from 16 35C is 0.62 % C at450 nm for 1 g/mL quinine sulfate dihydrate in 0.1 mol/L HClO4.54. Significance and Use4.1 When determining the limiting detectable concentrationof a fluorescent substance, it is usually necessary to increasethe readout scale of a photoelectri
21、c instrument to a point wherenoise (that is, random fluctuations of the system) becomesapparent. This noise will be superimposed upon the signal fromthe sample.4.2 In molecular fluorescence spectroscopy, the limit ofdetection for the sample will be determined by the limitingsignal-to-noise ratio, S/
22、N, where the signal, S, is the differencebetween readings obtained with the sample and blanksolutions, and N is the total root-mean-square (rms) noise. Thelimit of detection for the sample will be given by theinstrument readings that give a signal equal to three times therms value of the noise.NOTE
23、2Factors other than noise affecting the sample concentrationcorresponding to the limit of detection include: the spectral bandwidths ofthe excitation and emission monochromators, the intensity of the excitinglight that can be concentrated on the sample, the fraction of thefluorescence collected by t
24、he detection system, the response time of thedetection system, and the purity of the solvent. The size and arrangementof the sample container with respect to the light beams are also important,as they affect both the desired signal and the extraneous signal that onlycontributes noise.NOTE 3The value
- 1.请仔细阅读文档,确保文档完整性,对于不预览、不比对内容而直接下载带来的问题本站不予受理。
- 2.下载的文档,不会出现我们的网址水印。
- 3、该文档所得收入(下载+内容+预览)归上传者、原创作者;如果您是本文档原作者,请点此认领!既往收益都归您。
下载文档到电脑,查找使用更方便
5000 积分 0人已下载
下载 | 加入VIP,交流精品资源 |
- 配套讲稿:
如PPT文件的首页显示word图标,表示该PPT已包含配套word讲稿。双击word图标可打开word文档。
- 特殊限制:
部分文档作品中含有的国旗、国徽等图片,仅作为作品整体效果示例展示,禁止商用。设计者仅对作品中独创性部分享有著作权。
- 关 键 词:
- ASTME579200420150980STANDARDTESTMETHODFORLIMITOFDETECTIONOFFLUORESCENCEOFQUININESULFATEINSOLUTION 溶液

链接地址:http://www.mydoc123.com/p-533079.html