ASTM E545-2014 red 7511 Standard Test Method for Determining Image Quality in Direct Thermal Neutron Radiographic Examination《测定直接热中子射线照相检验中成像质量的标准试验方法》.pdf
《ASTM E545-2014 red 7511 Standard Test Method for Determining Image Quality in Direct Thermal Neutron Radiographic Examination《测定直接热中子射线照相检验中成像质量的标准试验方法》.pdf》由会员分享,可在线阅读,更多相关《ASTM E545-2014 red 7511 Standard Test Method for Determining Image Quality in Direct Thermal Neutron Radiographic Examination《测定直接热中子射线照相检验中成像质量的标准试验方法》.pdf(7页珍藏版)》请在麦多课文档分享上搜索。
1、Designation: E545 05 (Reapproved 2010)E545 14Standard Test Method forDetermining Image Quality in Direct Thermal NeutronRadiographic Examination1This standard is issued under the fixed designation E545; the number immediately following the designation indicates the year oforiginal adoption or, in th
2、e case of revision, the year of last revision. A number in parentheses indicates the year of last reapproval. Asuperscript epsilon () indicates an editorial change since the last revision or reapproval.This standard has been approved for use by agencies of the U.S. Department of Defense.1. Scope Sco
3、pe*1.1 This test method covers the use of an Image Quality Indicator (IQI) system to determine the relative2 quality of radiographicimages produced by direct, thermal neutron radiographic examination. The requirements expressed in this test method are notintended to control the quality level of mate
4、rials and components.1.2 This standard does not purport to address the safety concerns, if any, associated with its use. It is the responsibility of theuser of this standard to establish appropriate safety and health practices and determine the applicability of regulatory limitationsprior to use.1.3
5、 The values stated in SI units are regarded to be standard.2. Referenced Documents2.1 ASTM Standards:3E543 Specification for Agencies Performing Nondestructive TestingE748 Practices for Thermal Neutron Radiography of MaterialsE803 Test Method for Determining the L/D Ratio of Neutron Radiography Beam
6、sE1316 Terminology for Nondestructive ExaminationsE2003 Practice for Fabrication of the Neutron Radiographic Beam Purity IndicatorsE2023 Practice for Fabrication of Neutron Radiographic Sensitivity Indicators3. Terminology3.1 DefinitionsFor definitions of terms used in this test method, see Terminol
7、ogy E1316, Section H.4. Summary of Test Method4.1 The judgment of the quality of a neutron radiograph is based upon the evaluation of images obtained from indicators thatare exposed along with the test object. In cases of limited film size or extended object size, the indicators may be exposed onano
8、ther film immediately prior to or following exposure of the test object under exactly the same conditions (refer to ProcessControl Radiographs, Section 10). The IQI values must be determined from films with an optical density between 2.0 to 3.0. Twotypes of IQIs are used.4.1.1 Beam Purity Indicator
9、(BPI)The BPI is a device used for quantitative determination of radiographic quality. It is apolytetrafluoroethylene block containing two boron nitride disks, two lead disks, and two cadmium wires.Akey feature of the BPIis the ability to make a visual analysis of its image for subjective information
10、, such as image unsharpness and film and processingquality. Densitometric measurements of the image of the device permit quantitative determination of the effective value for thethermal neutron content, gamma content, pair production content, and scattered neutron content. The BPI shall be construct
11、ed inaccordance with Practice E2003. Optionally, any BPI fabricated prior to publication of Practice E2003 which conforms to TestMethod E545 - 81 through 91 may be used.1 This test method is under the jurisdiction of ASTM Committee E07 on Nondestructive Testing and is the direct responsibility of Su
12、bcommittee E07.05 on Radiology(Neutron) Method.Current edition approved June 1, 2010June 1, 2014. Published November 2010June 2014. Originally approved in 1975. Last previous edition approved in 20052010 asE545 - 05.E545 - 05(2010). DOI: 10.1520/E0545-05R10.10.1520/E0545-14.2 The numerical values ob
13、tained in the calculations described herein may vary between different film processing systems, film types, and within one processing systemif processing variables changes.3 For referencedASTM standards, visit theASTM website, www.astm.org, or contactASTM Customer Service at serviceastm.org. For Ann
14、ual Book of ASTM Standardsvolume information, refer to the standards Document Summary page on the ASTM website.This document is not an ASTM standard and is intended only to provide the user of an ASTM standard an indication of what changes have been made to the previous version. Becauseit may not be
15、 technically possible to adequately depict all changes accurately, ASTM recommends that users consult prior editions as appropriate. In all cases only the current versionof the standard as published by ASTM is to be considered the official document.*A Summary of Changes section appears at the end of
16、 this standardCopyright ASTM International, 100 Barr Harbor Drive, PO Box C700, West Conshohocken, PA 19428-2959. United States14.1.2 Sensitivity Indicator (SI)The SI is one of several devices used for qualitative determination of the sensitivity of detailvisible on a neutron radiograph. The SI is a
17、 step-wedge device containing gaps and holes of known dimensions. Visual inspectionof the image of this device provides subjective information regarding total radiographic sensitivity with respect to the step-blockmaterial. The SI shall be in accordance with Practice E2023. Optionally, any SI fabric
18、ated prior to publication of Practice E2023which conforms to Test Method E545-81 through 91 may be used.4.2 Neutron radiography practices are discussed in Practices E748.5. Significance and Use5.1 The BPI is designed to yield quantitative information concerning neutron beam and image system paramete
19、rs that contributeto film exposure and thereby affect overall image quality. In addition, the BPI can be used to verify the day-to-day consistency ofthe neutron radiographic quality. Gadolinium conversion screens and single-emulsion silver-halide films, exposed together in theneutron imaging beam, w
20、ere used in the development and testing of the BPI. Use of alternative detection systems may producedensitometric readings that are not valid for the equations used in Section 9.5.2 The only truly valid sensitivity indicator is a reference standard part. A reference standard part is a material or co
21、mponentthat is the same as the object being neutron radiographed except with a known standard discontinuity, inclusion, omission, or flaw.The sensitivity indicators were designed to substitute for the reference standard and provide qualitative information on hole andgap sensitivity.5.3 The number of
22、 areas or objects to be radiographed and the film acceptance standard used should be specified in the contract,purchase order, specification, or drawings.6. Basis of Application6.1 Qualification of Nondestructive AgenciesIf specified in the contractual agreement, NDT agencies shall be qualified ande
23、valuated in accordance with Practice E543. The applicable revision of Practice E543 shall be specified in the contractualagreement.6.2 Procedures and TechniquesThe procedures and techniques to be utilized shall be as described in this test method unlessotherwise specified. Specific techniques may be
24、 specified in contractual documents.6.3 Extent of ExaminationThe extent of examination shall be in accordance with Section 7 unless otherwise specified.6.4 Reporting Criteria/Acceptance CriteriaReporting criteria for the examination results shall be in accordance with Section11 unless otherwise spec
- 1.请仔细阅读文档,确保文档完整性,对于不预览、不比对内容而直接下载带来的问题本站不予受理。
- 2.下载的文档,不会出现我们的网址水印。
- 3、该文档所得收入(下载+内容+预览)归上传者、原创作者;如果您是本文档原作者,请点此认领!既往收益都归您。
下载文档到电脑,查找使用更方便
5000 积分 0人已下载
下载 | 加入VIP,交流精品资源 |
- 配套讲稿:
如PPT文件的首页显示word图标,表示该PPT已包含配套word讲稿。双击word图标可打开word文档。
- 特殊限制:
部分文档作品中含有的国旗、国徽等图片,仅作为作品整体效果示例展示,禁止商用。设计者仅对作品中独创性部分享有著作权。
- 关 键 词:
- ASTME5452014RED7511STANDARDTESTMETHODFORDETERMININGIMAGEQUALITYINDIRECTTHERMALNEUTRONRADIOGRAPHICEXAMINATION

链接地址:http://www.mydoc123.com/p-533010.html