ASTM E1026-2013 red 1875 Standard Practice for Using the Fricke Dosimetry System《使用弗里克剂量测定系统的标准实施规范》.pdf
《ASTM E1026-2013 red 1875 Standard Practice for Using the Fricke Dosimetry System《使用弗里克剂量测定系统的标准实施规范》.pdf》由会员分享,可在线阅读,更多相关《ASTM E1026-2013 red 1875 Standard Practice for Using the Fricke Dosimetry System《使用弗里克剂量测定系统的标准实施规范》.pdf(14页珍藏版)》请在麦多课文档分享上搜索。
1、Designation: E1026 041E1026 13 An American National StandardStandard Practice forUsing the Fricke Reference-Standard Dosimetry System1This standard is issued under the fixed designation E1026; the number immediately following the designation indicates the year oforiginal adoption or, in the case of
2、revision, the year of last revision. A number in parentheses indicates the year of last reapproval. Asuperscript epsilon () indicates an editorial change since the last revision or reapproval.1 NOTEEquations 3 and 4 were corrected editorially in August 2005.1. Scope1.1 This practice covers the proce
3、dures for preparation, testing and using the acidic aqueous ferrous ammonium sulfate solutiondosimetry system to measure absorbed dose to water when exposed to ionizing radiation. The system consists of a dosimeter andappropriate analytical instrumentation. The system will be referred to as the Fric
4、ke system. It is classified as a reference-standarddosimetry system (see ISO/ASTM 51261).dosimetry system. The Fricke dosimetry system may be used as either a referencestandard dosimetry system or a routine dosimetry system.1.2 This practice is one of a set of standards that provides recommendations
5、 for properly implementing dosimetry in radiationprocessing, and describes a means of achieving compliance with the requirements of Practice E2628 for the Fricke dosimetrysystem. It is intended to be read in conjunction with Practice E2628.1.3 The practice describes the spectrophotometric analysis p
6、rocedures for the Fricke dosimeter.dosimetry system.1.4 This practice applies only to gamma rays, x-raysradiation, X-radiation (bremsstrahlung), and high-energy electrons.1.5 This practice applies provided the following are satisfied:1.5.1 The absorbed dose range shall be from 20 to 400 Gy (1).21.5.
7、2 The absorbed-dose rate does not exceed 106 Gys1 (2).1.5.3 For radioisotope gamma-raygamma sources, the initial photon energy is greater than 0.6 MeV. For x-raysX-radiation(bremsstrahlung), the initial energy of the electrons used to produce the photons is equal to or greater than 2 MeV. For electr
8、onbeams, the initial electron energy is greater than 8 MeV (see ICRU Reports 34 and 35). MeV.NOTE 1The lower energy limits given are appropriate for a cylindrical dosimeter ampoule of 12-mm outside12 mm diameter. Corrections for dosegradients across an ampoule of that diameter or less are notdisplac
9、ement effects and dose gradient across the ampoule may be required for electron beamsrequired.(3). The Fricke dosimetry system may be used at lower energies by employing thinner (in the beam direction) dosimeter containers (see ICRUReport 35).1.5.4 The irradiation temperature of the dosimeter should
10、 be within the range of 10 to 60C.1.6 This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibilityof the user of this standard to establish appropriate safety and health practices and determine the applicability of regulatorylimitati
11、ons prior to use.2. Referenced Documents2.1 ASTM Standards:3C912 Practice for Designing a Process for Cleaning Technical GlassesD1193 Specification for Reagent WaterE170 Terminology Relating to Radiation Measurements and DosimetryE178 Practice for Dealing With Outlying ObservationsE275 Practice for
12、Describing and Measuring Performance of Ultraviolet and Visible SpectrophotometersE666 Practice for Calculating Absorbed Dose From Gamma or X Radiation1 This practice is under the jurisdiction ofASTM Committee E61 on Radiation Processing and is the direct responsibility of Subcommittee E61.02 on Dos
13、imetry Systems.Current edition approved Jan. 1, 2004Jan. 1, 2013. Published February 2004March 2013. Originally approved in 1984. Last previous edition approved in 20032004 asE1026 03.E1026 041. DOI: 10.1520/E1026-04E01.10.1520/E1026-13.2 The boldface numbers that appear in parentheses refer to a li
14、st of references at the end of this practice.3 For referenced ASTM and ISO/ASTM standards, visit the ASTM webiste, www.astm.org, or contact ASTM Customer Service at serviceastm.org. For Annual Bookof ASTM Standards volume information, refer to the standards Document Summary page on the ASTM website.
15、This document is not an ASTM standard and is intended only to provide the user of an ASTM standard an indication of what changes have been made to the previous version. Becauseit may not be technically possible to adequately depict all changes accurately, ASTM recommends that users consult prior edi
16、tions as appropriate. In all cases only the current versionof the standard as published by ASTM is to be considered the official document.Copyright ASTM International, 100 Barr Harbor Drive, PO Box C700, West Conshohocken, PA 19428-2959. United States1E668 Practice for Application of Thermoluminesce
17、nce-Dosimetry (TLD) Systems for Determining Absorbed Dose inRadiation-Hardness Testing of Electronic DevicesE925 Practice for Monitoring the Calibration of Ultraviolet-Visible Spectrophotometers whose Spectral Bandwidth does notExceed 2 nmE958 Practice for Estimation of the Spectral Bandwidth of Ult
18、raviolet-Visible SpectrophotometersE2628 Practice for Dosimetry in Radiation Processing2.2 ISO/ASTM Standards:ISO/ASTM 51205 Method for Using the Ceric-Cerous Sulfate Dosimetry System3ISO/ASTM 51261 GuidePractice for Selection and Calibration of Routine Dosimetry Systems for Radiation ProcessingISO/
19、ASTM 51707 Guide for Estimating Uncertainties in Dosimetry for Radiation Processing2.3 ISO/IEC Standard:ISO/IEC 17025 General requirements for the competence of testing and calibration laboratories42.4 International Commission on Radiation Units and Measurements (ICRU) Reports:5ICRU Report 3414 The
20、Dosimetry of Pulsed RadiationRadiation Dosimetry: X Rays and Gamma Rays with Maximum PhotonEnergies Between 0.6 and 50 MeVICRU Report 35 Radiation Dosimetry: Electrons with Initial Energies Between 1 and 50 MeVICRU Report 60 Fundamental Quantities and Units for Ionizing Radiation4ICRU Report 64 Dosi
21、metry of High-Energy Photon Beams based on Standards of Absorbed Dose to WaterICRU Report 80 Dosimetry Systems for Use in Radiation ProcessingICRU Report 85a Fundamental Quantities and Units for Ionizing Radiation2.5 Joint Committee for Guides in Metrology (JCGM) Reports:6JCGM 100:2008 GUM 1995 , wi
22、th minor corrections, Evaluation of measurement data Guide to the Expression of Uncertaintyin Measurement2.6 National Research Council Canada (NRCC):PIRS-0815 The IRS Fricke Dosimetry System73. Terminology3.1 Definitions:3.1.1 Fricke Dosimetry Systemapproved laboratoryconsists of a liquid chemical d
23、osimeter (composed of ferrous sulfate orferrous ammonium sulfate in aqueous sulfuric acid solution), a spectrophotometer (to measure optical absorbance) and itsassociated reference standards, and procedures for its use.laboratory that is a recognized national metrology institute; or has beenformally
24、 accredited to ISO/IEC 17025; or has a quality system consistent with the requirements of ISO/IEC 17025.3.1.1.1 DiscussionThe Fricke dosimetry system is considered a reference-standard dosimetry system. Sodium chloride is usually added to dosimetricsolution to minimize the effects of organic impurit
- 1.请仔细阅读文档,确保文档完整性,对于不预览、不比对内容而直接下载带来的问题本站不予受理。
- 2.下载的文档,不会出现我们的网址水印。
- 3、该文档所得收入(下载+内容+预览)归上传者、原创作者;如果您是本文档原作者,请点此认领!既往收益都归您。
下载文档到电脑,查找使用更方便
5000 积分 0人已下载
下载 | 加入VIP,交流精品资源 |
- 配套讲稿:
如PPT文件的首页显示word图标,表示该PPT已包含配套word讲稿。双击word图标可打开word文档。
- 特殊限制:
部分文档作品中含有的国旗、国徽等图片,仅作为作品整体效果示例展示,禁止商用。设计者仅对作品中独创性部分享有著作权。
- 关 键 词:
- ASTME10262013RED1875STANDARDPRACTICEFORUSINGTHEFRICKEDOSIMETRYSYSTEM 使用 弗里克 剂量 测定 系统 标准 实施 规范 PDF

链接地址:http://www.mydoc123.com/p-527766.html