ASTM D7962-2017 red 9375 Standard Practice for Determination of Minimum Immersion Depth and Assessment of Temperature Sensor Measurement Drift《测定温度传感器最小浸入深度以及评估测得漂移的标准实施规程》.pdf
《ASTM D7962-2017 red 9375 Standard Practice for Determination of Minimum Immersion Depth and Assessment of Temperature Sensor Measurement Drift《测定温度传感器最小浸入深度以及评估测得漂移的标准实施规程》.pdf》由会员分享,可在线阅读,更多相关《ASTM D7962-2017 red 9375 Standard Practice for Determination of Minimum Immersion Depth and Assessment of Temperature Sensor Measurement Drift《测定温度传感器最小浸入深度以及评估测得漂移的标准实施规程》.pdf(4页珍藏版)》请在麦多课文档分享上搜索。
1、Designation: D7962 16D7962 17Standard Practice forDetermination of Minimum Immersion Depth andAssessment of Temperature Sensor Measurement Drift1This standard is issued under the fixed designation D7962; the number immediately following the designation indicates the year oforiginal adoption or, in t
2、he case of revision, the year of last revision. A number in parentheses indicates the year of last reapproval. Asuperscript epsilon () indicates an editorial change since the last revision or reapproval.1. Scope*1.1 This practice describes a two procedures for use with temperature measurement device
3、s. Methodology is described fordetermining minimum immersion depth for thermal sensors, in particular RTDs or similar temperature sensors. Included is aprocedure for consistently preparing a reference bath for the purpose of monitoring measurement drift of thermal sensors such asliquid-in-glass or d
4、igital contact thermometers.1.2 This practice focuses on temperature measurement drift in a laboratory. If the user requires greater measurement accuracy,then they should follow the instructions in Practice E563.1.3 The values stated in SI units are to be regarded as standard. No other units of meas
5、urement are included in this standard.1.4 This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibilityof the user of this standard to establish appropriate safety and health practices and determine the applicability of regulatorylimi
6、tations prior to use.1.5 This international standard was developed in accordance with internationally recognized principles on standardizationestablished in the Decision on Principles for the Development of International Standards, Guides and Recommendations issuedby the World Trade Organization Tec
7、hnical Barriers to Trade (TBT) Committee.2. Referenced Documents2.1 ASTM Standards:2E563 Practice for Preparation and Use of an Ice-Point Bath as a Reference Temperature3. Terminology3.1 Definitions:3.1.1 digital contact thermometer (DCT), nan electronic device consisting of a digital display and as
8、sociated temperaturesensing temperature-sensing probe.3.1.1.1 DiscussionThis device consists of a temperature sensor connected to a measuring instrument; this instrument measures the temperature-dependent quantity of the sensor, computes the temperature from the measured quantity, and provides a dig
9、ital output. This digitaloutput goes to a digital display and/or recording device that may be internal or external to the device. These devices are sometimesreferred to as a “digital thermometer”.thermometer.”3.1.1.2 DiscussionPortable electronic thermometers (PET) is an acronym sometimes used to re
10、fer to a subset of the devices covered by this definition.1 This practice is under the jurisdiction of ASTM Committee D02 on Petroleum Products, Liquid Fuels, and Lubricants and is the direct responsibility of SubcommitteeD02.91 on Coordinating Subcommittee on Thermometry.Current edition approved De
11、c. 1, 2016May 1, 2017. Published January 2017May 2017. Originally approved in 2015. Last previous edition approved in 20152016 asD7962 15a.D7962 16. DOI: 10.1520/D7962-16.10.1520/D7962-17.2 For referencedASTM standards, visit theASTM website, www.astm.org, or contactASTM Customer Service at servicea
12、stm.org. For Annual Book of ASTM Standardsvolume information, refer to the standards Document Summary page on the ASTM website.This document is not an ASTM standard and is intended only to provide the user of an ASTM standard an indication of what changes have been made to the previous version. Beca
13、useit may not be technically possible to adequately depict all changes accurately, ASTM recommends that users consult prior editions as appropriate. In all cases only the current versionof the standard as published by ASTM is to be considered the official document.*A Summary of Changes section appea
14、rs at the end of this standardCopyright ASTM International, 100 Barr Harbor Drive, PO Box C700, West Conshohocken, PA 19428-2959. United States13.1.2 ice-point bath, nphysical system containing ice and water assembled to realize the ice point as a reference temperature,or to establish a constant tem
15、perature near 0 C.3.1.3 minimum immersion depth, ndepth that a thermometer should be immersed, in a uniform temperature environment, suchthat further immersion does not produce a change in indicated temperature greater than the specified tolerance.4. Summary of Practice4.1 This practice describes a
16、procedure for consistently preparing an ice bath that is an intimate mixture of crushed ice or iceparticles and water in a thermally insulating vessel open to the atmosphere.4.1.1 Cautionwhen the ice bath is not made from distilled water, its temperature will differ from the natural fixed-pointtempe
17、rature by a consistent amount, typically less than 0.02 C. If the user needs a more accurate ice point, then they should usePractice E563 to prepare the ice bath.4.2 This practice includes a procedure for determining the minimum immersion depth of the temperature sensor using an icebath.4.2.1 This p
18、rocedure determines minimum immersion depth with a 25 C differential between sensed temperature and ambienttemperature. If the probe is subjected to a greater differential temperature, a larger immersion depth may be required to correctlymeasure the temperature.5. Significance and Use5.1 This practi
19、ce provides a means for the users of ASTM Committee D02 standards to monitor the drift in sensed temperatureof liquid-in-glass thermometer (LiG), and digital contact thermometers (DCT). Digital contact thermometers are sometimesreferred to as portable electronic thermometers (PET) or simply digital
20、thermometers.5.2 This practice is not suitable for determining the accuracy or calibration of a temperature measuring temperature-measuringdevice as the error in the ice bath temperature can be greater than 0.02 C. For greater accuracy, the user should use Practice E563to prepare the ice bath.5.3 Th
21、e ice point is a common practical industrial reference point of thermometry. The ice point is relatively simple to realizeand provides a readily available natural fixed-point reference temperature.5.4 This practice only checks the measurement drift at a single temperature. It will not detect a chang
22、e in measurement responsewith change in temperature. Temperature measuring Temperature-measuring devices should be recalibrated at set intervals. Seedevice supplier for recommendations.5.5 This practice provides a technique to determine minimum immersion depth of the sensing probe of the thermometer
23、 usingan ice bath. The minimum immersion depth determined by this practice may change when the differential temperature differssignificantly from the conditions described. A greater differential will likely increase the minimum immersion depth.6. Reagents and Equipment6.1 WaterGood quality drinking
24、distilled or deionized water from a consistent source.6.2 IceCrushed ice prepared from the water in 6.1.6.3 Insulated containerContaineranAn insulated container large enough to hold the thermometer and provide severalcentimeters of water-ice slush around the measuring portion of the thermometer.7. P
- 1.请仔细阅读文档,确保文档完整性,对于不预览、不比对内容而直接下载带来的问题本站不予受理。
- 2.下载的文档,不会出现我们的网址水印。
- 3、该文档所得收入(下载+内容+预览)归上传者、原创作者;如果您是本文档原作者,请点此认领!既往收益都归您。
下载文档到电脑,查找使用更方便
5000 积分 0人已下载
下载 | 加入VIP,交流精品资源 |
- 配套讲稿:
如PPT文件的首页显示word图标,表示该PPT已包含配套word讲稿。双击word图标可打开word文档。
- 特殊限制:
部分文档作品中含有的国旗、国徽等图片,仅作为作品整体效果示例展示,禁止商用。设计者仅对作品中独创性部分享有著作权。
- 关 键 词:
- ASTMD79622017RED9375STANDARDPRACTICEFORDETERMINATIONOFMINIMUMIMMERSIONDEPTHANDASSESSMENTOFTEMPERATURESENSORMEASUREMENTDRIFT

链接地址:http://www.mydoc123.com/p-526575.html