《ASTM D7731-2011 1250 Standard Test Method for Determination of Dipropylene Glycol Monobutyl Ether and Ethylene Glycol Monobutyl Ether in Sea Water by Liquid Chromatography Tandem M.pdf》由会员分享,可在线阅读,更多相关《ASTM D7731-2011 1250 Standard Test Method for Determination of Dipropylene Glycol Monobutyl Ether and Ethylene Glycol Monobutyl Ether in Sea Water by Liquid Chromatography Tandem M.pdf(9页珍藏版)》请在麦多课文档分享上搜索。
1、Designation: D7731 11Standard Test Method forDetermination of Dipropylene Glycol Monobutyl Ether andEthylene Glycol Monobutyl Ether in Sea Water by LiquidChromatography/Tandem Mass Spectrometry (LC/MS/MS)1This standard is issued under the fixed designation D7731; the number immediately following the
2、 designation indicates the year oforiginal adoption or, in the case of revision, the year of last revision. A number in parentheses indicates the year of last reapproval. Asuperscript epsilon () indicates an editorial change since the last revision or reapproval.1. Scope1.1 This procedure covers the
3、 determination of DipropyleneGlycol Monobutyl Ether (DPGBE) and Ethylene GlycolMonobutyl Ether (EGBE) in sea water by direct injection usingliquid chromatography (LC) and detection with tandem massspectrometry (MS/MS). This analyte is qualitatively and quan-titatively determined by this method. This
4、 method adheres toselected reaction monitoring (SRM) mass spectrometry.1.2 The Detection Verification Level (DVL) and ReportingRange for DPGBE and EGBE are listed in Table 1.1.2.1 The DVL is required to be at a concentration at least3 times below the Reporting Limit (RL) and have a signal/noise rati
5、o greater than 3:1. Fig. 1 and Fig. 2 display thesignal/noise ratio of the single reaction monitoring (SRM)transition.1.2.2 The reporting limit is the concentration of the Level 1calibration standard as shown in Table 4 for DPGBE andEGBE, taking into account the 20% sample preparation dilu-tion fact
6、or.1.3 The values stated in SI units are to be regarded asstandard. No other units of measurement are included in thisstandard.1.4 This standard does not purport to address all of thesafety concerns, if any, associated with its use. It is theresponsibility of the user of this standard to establish a
7、ppro-priate safety and health practices and determine the applica-bility of regulatory limitations prior to use.2. Referenced Documents2.1 ASTM Standards:2D1193 Specification for Reagent WaterD2777 Practice for Determination of Precision and Bias ofApplicable Test Methods of Committee D19 on Water2.
8、2 Other Standards:3EPApublication SW-846 Test Methods for Evaluating SolidWaste, Physical/Chemical Methods3. Terminology3.1 Definitions:3.1.1 detection verification level, DVL, na concentrationthat has a signal/noise ratio greater than 3:1 and is at least 3times below the Reporting Limit (RL).3.1.2
9、reporting limit, RL, nthe concentration of thelowest-level calibration standard used for quantification.3.1.2.1 DiscussionIn this test method, a 20 mL samplealiquot is diluted to a 25 mL final volume after thoroughlyrinsing the collection vial with acetonitrile for quantitativetransfer. In this case
10、, the lowest calibration level of 100 ppb forEGBE would allow for a reporting limit of 125 ppb to beachieved.3.2 Abbreviations:3.2.1 ppbparts per billion, g/L3.2.2 pptparts per trillion, ng/L3.2.3 mMmillimolar,1x10-3moles/L3.2.4 NAno addition3.2.5 NDnon-detect1This test method is under the jurisdict
11、ion of ASTM Committee D19 on Waterand is the direct responsibility of Subcommittee D19.06 on Methods forAnalysis forOrganic Substances in Water.Current edition approved May 1, 2011. Published May, 2011. DOI: 10.1520/D773111.2For referenced ASTM standards, visit the ASTM website, www.astm.org, orcont
12、act ASTM Customer Service at serviceastm.org. For Annual Book of ASTMStandards volume information, refer to the standards Document Summary page onthe ASTM website.3Available from from National Technical Information Service (NTIS), U.S.Department of Commerce, 5285 Port Royal Road, Springfield, VA, 22
13、161 or athttp:/www.epa.gov/epawaste/hazard/testmethods/index.htmTABLE 1 Detection Verification Level and Reporting RangeAnalyte DVL (g/L) Reporting Range (g/L)DPGBE 0.2 110EGBE 25 12512501Copyright ASTM International, 100 Barr Harbor Drive, PO Box C700, West Conshohocken, PA 19428-2959, United State
14、s.FIG. 1 Detection Verification Level Signal/Noise Ratio.FIG. 2 Reporting Level (Calibra tion standard) Signal/Noise Ratio.D7731 1124. Summary of Test Method4.1 This is a performance based method, and modificationsare allowed to improve performance.4.2 For DPGBE and EGBE analysis, samples a re shipp
15、ed tothe lab between 0C and 6C and analyzed within 5 days ofcollection. The DOW MSDS sheet on DOWANOL* DPNBglycol ether (DPGBE) Issue Date: 06/18/2010 lists that thematerial is readily biodegradable. The Organisation for Eco-nomic Co-Operation and Development (OECD) 302B Test lists96% biodegradation
16、 in 28 days.4.3 In the lab, the entire collected 20 mL sample is spikedwith surrogate and brought to a volume of 25 mL withacetonitrile. This prepared sample is then filtered using asyringe driven filter unit, and analyzed by LC/MS/MS. Ifvisible oil is present, the prepared sample is allowed to sett
17、leresulting in an oil layer at the top of the 25 mL solution. Aportion of the aqueous (bottom) layer is filtered, leaving the oillayer behind, through a syringe driven filter assembly andanalyzed by LC/MS/MS.4.4 DPGBE, EGBE and surrogate are identified by retentiontime and one SRM transition. The ta
18、rget analytes and surrogateare quantitated using the SRM transitions utilizing an externalcalibration. The final report issued for each sample lists theconcentration of DPGBE, EGBE and the surrogate recovery.5. Significance and Use5.1 DPGBE and EGBE have a variety of residential andindustrial applic
19、ations such as, cleaning formulations, surfacecoatings, inks and cosmetics. These analytes may be releasedinto the environment at levels that may be harmful to aquaticlife.5.2 This method has been investigated for use with reagentand sea water.6. Interferences6.1 Method interferences may be caused b
20、y contaminants insolvents, reagents, glassware, and other apparatus producingdiscrete artifacts or elevated baselines. All of these materialsare demonstrated to be free from interferences by analyzinglaboratory reagent blanks under the same conditions assamples.6.2 All glassware is washed in hot wat
21、er with detergent andrinsed in hot water followed by distilled water. Detergentscontaining DPGBE or EGBE must not be used. The glasswareis then dried and heated in an oven at 250C for 15 to 30minutes. All glassware is subsequently cleaned with acetonefollowed by methanol.6.3 All reagents and solvent
22、s should be pesticide residuepurity or higher to minimize interference problems.6.4 Matrix interferences may be caused by contaminants inthe sample. The extent of matrix interferences can varyconsiderably from sample source depending on variations ofthe sample matrix.7. Apparatus7.1 LC/MS/MS System7
23、.1.1 Liquid Chromatography SystemA complete LC sys-tem is needed in order to analyze samples.4Any system that iscapable of performing at the flows, pressures, controlledtemperatures, sample volumes, and requirements of the stan-dard may be used.7.1.2 Analytical ColumnWaters- XBridgey, 2.1 x 150mm, 3
24、.5 m particle size was used to develop this test method.Any column that achieves baseline resolution of these analytesmay be used. Baseline resolution simplifies data analysis andcan reduce the chance of ion suppression, leading to higherlimits of detection. The retention times and order of elutionm
- 1.请仔细阅读文档,确保文档完整性,对于不预览、不比对内容而直接下载带来的问题本站不予受理。
- 2.下载的文档,不会出现我们的网址水印。
- 3、该文档所得收入(下载+内容+预览)归上传者、原创作者;如果您是本文档原作者,请点此认领!既往收益都归您。
下载文档到电脑,查找使用更方便
5000 积分 0人已下载
下载 | 加入VIP,交流精品资源 |
- 配套讲稿:
如PPT文件的首页显示word图标,表示该PPT已包含配套word讲稿。双击word图标可打开word文档。
- 特殊限制:
部分文档作品中含有的国旗、国徽等图片,仅作为作品整体效果示例展示,禁止商用。设计者仅对作品中独创性部分享有著作权。
- 关 键 词:
- ASTMD773120111250STANDARDTESTMETHODFORDETERMINATIONOFDIPROPYLENEGLYCOLMONOBUTYLETHERANDETHYLENEGLYCOLMONOBUTYLETHERINSEAWATERBYLIQUIDCHROMATOGRAPHYTANDEMMPDF

链接地址:http://www.mydoc123.com/p-526205.html