《ASTM D7639-2010(2014) 9214 Standard Test Method for Determination of Zirconium Treatment Weight or Thickness on Metal Substrates by X-Ray Fluorescence《使用X射线荧光法测定金属基体上的锆处理重量或厚度的标准试验.pdf》由会员分享,可在线阅读,更多相关《ASTM D7639-2010(2014) 9214 Standard Test Method for Determination of Zirconium Treatment Weight or Thickness on Metal Substrates by X-Ray Fluorescence《使用X射线荧光法测定金属基体上的锆处理重量或厚度的标准试验.pdf(3页珍藏版)》请在麦多课文档分享上搜索。
1、Designation: D7639 10 (Reapproved 2014)Standard Test Method forDetermination of Zirconium Treatment Weight or Thicknesson Metal Substrates by X-Ray Fluorescence1This standard is issued under the fixed designation D7639; the number immediately following the designation indicates the year oforiginal a
2、doption or, in the case of revision, the year of last revision. A number in parentheses indicates the year of last reapproval. Asuperscript epsilon () indicates an editorial change since the last revision or reapproval.1. Scope1.1 This test method covers the use of X-ray fluorescence(XRF) spectromet
3、ry for the determination of the mass ofzirconium (Zr) coating weight per unit area of metal substrates.1.2 Coating treatments can also be expressed in units oflinear thickness provided that the density of the coating isknown, or provided that a calibration curve has been estab-lished for thickness d
4、etermination using standards with treat-ment matching this of test specimens to be analyzed. Forsimplicity, the method will subsequently refer to the determi-nation expressed as coating weight.1.3 XRF is applicable for the determination of the coatingweight as zirconium or total coating weight of a
5、zirconiumcontaining treatment, or both, on a variety of metal substrates.1.4 The maximum measurable coating weight for a givencoating is that weight beyond which the intensity of thecharacteristic X-ray radiation from the coating or the substrateis no longer sensitive to small changes in weight.1.5
6、The values stated in SI units are to be regarded asstandard. No other units of measurement are included in thisstandard.1.6 This standard does not purport to address all of thesafety concerns, if any, associated with its use. It is theresponsibility of the user of this standard to establish appro-pr
7、iate safety and health practices and determine the applica-bility of regulatory limitations prior to use.2. Referenced Documents2.1 ASTM Standards:2E177 Practice for Use of the Terms Precision and Bias inASTM Test MethodsE691 Practice for Conducting an Interlaboratory Study toDetermine the Precision
8、 of a Test Method3. Summary of Test Method3.1 The test specimen is placed in the X-ray beam, and theresultant peak intensity of the zirconium Ka line (at 0.0786 nmor 15.747 keV) or the zirconium La line (at 0.606 nm or 2.042keV) is measured. The intensity (in counts or counts persecond) is then comp
9、ared to a previously prepared calibrationcurve or equation to obtain the coating weight of zirconiumtreatment in mg/m2or mg/ft2(or m or nm).3.2 The exact relationship between the measured number ofcounts and the corresponding coating weight (or coatingthickness) must be established for each individu
10、al combinationof substrate and zirconium-containing treatment. Usually de-termined by the treatment supplier, this relationship is estab-lished by using primary standards having known amounts ofthe same treatment applied to the same substrate compositionas the test specimens to be measured.4. Signif
11、icance and Use4.1 The procedure described in this test method is designedto provide a method by which the coating weight of zirconiumtreatments on metal substrates may be determined.4.2 This test method is applicable for determination of thetotal coating weight and the zirconium coating weight of az
12、irconium-containing treatment.5. Apparatus5.1 X-Ray Fluorescence Spectrometer, capable of measuringthe intensity of zirconium Ka or La line, and establish therelationship between peak intensity and coating weight. Thespectrometers design must include, as a minimum, the follow-ing features:5.1.1 Sour
13、ce of X-Ray Excitation, X-ray tube with excitationabove 2.55 keV if measuring the zirconium La line, or above18 keV if measuring the zirconium Ka line.5.1.2 X-Ray Detector, with high sensitivity and capable ofdiscriminating between zirconium La or Ka radiation and otherX-rays of higher or lower ener
14、gies.5.1.2.1 In the case of wavelength dispersive X-ray fluores-cence (WDXRF), this can be an analyzing crystal (for example,1This test method is under the jurisdiction of ASTM Committee D01 on Paintand Related Coatings, Materials, and Applications and is the direct responsibility ofSubcommittee D01
15、.53 on Coil Coated Metal.Current edition approved June 15, 2014. Published June 2014. Originallyapproved in 2010. Last previous edition approved in 2010 as D7639 10.DOI:10.1520/D7639-10R14.2For referenced ASTM standards, visit the ASTM website, www.astm.org, orcontact ASTM Customer Service at servic
16、eastm.org. For Annual Book of ASTMStandards volume information, refer to the standards Document Summary page onthe ASTM website.Copyright ASTM International, 100 Barr Harbor Drive, PO Box C700, West Conshohocken, PA 19428-2959. United States1fixed channel, goniometer) setup to detect the zirconiumX-
17、rays (La or Ka line). Germanium 111 has been found to beacceptable for the Zirconium La line and LiF220 or LiF200 forthe zirconium Ka line.5.1.2.2 In the case of energy dispersive X-ray fluorescence(EDXRF), it can be a proportional counter, or a semiconductorsuch as a PIN diode or a silicon-drift de
18、tector.5.1.3 Pulse-Height Analyzer, or other means of energydiscrimination.5.1.4 Optical Path, specified by manufacturer. A helium orvacuum path is recommended when measuring the zirconiumLa line in order to minimize the attenuation of the X-rays bythe air in the optical path. The zirconium Ka line
19、has a higherenergy and its intensity will not be affected by air.5.1.5 Signal Conditioning and Data Handling System,whereby a coating weight versus X-ray counts curve may beestablished within the system for the direct readout of coatingweight.5.1.6 Sample Spinner (optional), to reduce the effects of
20、coating weight variation across the test specimen.6. Calibration Standards and Test Specimens6.1 Calibration Standards should be specimens for whichthe coating weight has been well characterized by otheranalytical procedures such as x-ray photoelectronspectroscopy, Auger emission spectroscopy, glow
21、dischargeoptical emission spectrometry, weigh-strip-weigh method, orother depth-profiling analytical technique.6.2 Blank (bare and untreated) Specimen (optional), shouldbe of the same metal substrate on which the treatment coatingweight is to be determined. It may be necessary to prepare ablank spec
22、imen from a treated specimen if an untreatedspecimen is not available. To best imitate a bare, untreatedblank, abrade a treated specimen that is from the same metalspecimen as the test specimen using a small abrasive pad.NOTE 1The first abrading is made parallel with the rolling direction ofthe meta
23、l, the second abrading is made perpendicular to the rollingdirection of the metal, and the third abrading is made parallel with therolling direction of the metal. This procedure should be repeated untilconstant readings are obtained. Always use the same side of the metalsubstrate from which the read
24、ings of the treated specimen will be taken.6.3 Calibration Standards and Test Specimens shall be cutto the required size, if necessary, for measurement by theinstrument.6.4 All calibration standards and test specimens shall be flatin the area of measurement and free of burrs and distortionsthat woul
- 1.请仔细阅读文档,确保文档完整性,对于不预览、不比对内容而直接下载带来的问题本站不予受理。
- 2.下载的文档,不会出现我们的网址水印。
- 3、该文档所得收入(下载+内容+预览)归上传者、原创作者;如果您是本文档原作者,请点此认领!既往收益都归您。
下载文档到电脑,查找使用更方便
5000 积分 0人已下载
下载 | 加入VIP,交流精品资源 |
- 配套讲稿:
如PPT文件的首页显示word图标,表示该PPT已包含配套word讲稿。双击word图标可打开word文档。
- 特殊限制:
部分文档作品中含有的国旗、国徽等图片,仅作为作品整体效果示例展示,禁止商用。设计者仅对作品中独创性部分享有著作权。
- 关 键 词:
- ASTMD7639201020149214STANDARDTESTMETHODFORDETERMINATIONOFZIRCONIUMTREATMENTWEIGHTORTHICKNESSONMETALSUBSTRATESBYXRAYFLUORESCENCE

链接地址:http://www.mydoc123.com/p-526001.html