1、Designation: D6318 03 (Reapproved 2014)Standard Practice forCalibrating a Fathometer Using a Bar Check Method1This standard is issued under the fixed designation D6318; the number immediately following the designation indicates the year oforiginal adoption or, in the case of revision, the year of la
2、st revision. A number in parentheses indicates the year of last reapproval. Asuperscript epsilon () indicates an editorial change since the last revision or reapproval.1. Scope1.1 This practice provides the user with procedures used inmanually calibrating the fathometer or electronic depthsounder. T
3、his narrative describes calibration terminology, de-scribes acceptable environmental conditions for calibration,and describes the calibration procedures.1.2 The references cited contain useful information in theconstruction and the correct operation of the calibration equip-ment.1.3 Any references c
4、ited in this narrative to specific prod-ucts or brand names are made for information only, and isintended to be descriptive, but not restrictive, of products thatwill perform satisfactorily.1.4 The values stated in inch-pound units are to be regardedas standard. The values given in parentheses are m
5、athematicalconversions to SI units that are provided for information onlyand are not considered standard.1.5 This standard does not purport to address all of thesafety concerns, if any, associated with its use. It is theresponsibility of the user of this standard to establish appro-priate safety and
6、 health practices and determine the applica-bility of regulatory limitations prior to use.2. Referenced Documents2.1 ASTM Standards:2D1129 Terminology Relating to WaterD5073 Practice for Depth Measurement of Surface Water3. Terminology3.1 DefinitionsRefer to Terminology D1129 for termsused in this g
7、uide.3.2 Definitions of Terms Specific to This Standard:3.2.1 bara section of metallic channel, I-beam, T-beam,pipe, plate, or ball that will reflect sound waves produced by afathometer.3.2.2 bar-checka method for calibrating a fathometer bysetting a sound or accoustic reflector (bar) below a survey
8、vessel to a known depth below a sounding transducer.3.2.3 draft (transducer draft)the vertical distance fromthe bottom of the transducer to the surface of the water.3.2.4 fathometerAn electronic device for registeringdepths of water by measuring the time required for thetransmission and reflection o
9、f sound waves between a sonictransducer and the lake or river bottom.3.2.5 soundto determine the depth of water.3.2.6 sounding scrollthe chart record of an underwatercross section or profile of the bottom.3.2.7 transducera device for translating electrical energyto acoustical energy and acoustical e
10、nergy back to electricalenergy.4. Significance and Use4.1 The accuracy of depth measurements made by a fath-ometer or echo sounder requires a number of correctionsbecause of the variability of sound or acoustic velocity in waterwith changes in temperature, salinity, and depth of water. Inaddition in
11、stability of the equipment can also result in signifi-cant errors. For additional information see Practice D5073.4.2 Calibration of echo sounding instruments is absolutelycritical in assuring the adequacy of depth measurements. Whenan echo sounder has been accurately calibrated, any observed(recorde
12、d) depth can be related to the true depth of water. Sincethe intended purpose of echo sounding is to measure the “true”depth, an independent “true” reference must be used.4.3 A bar-check is the most wide-spread, easiest toconstruct, and most economical mechanical method to deter-mine corrections for
13、 instrument and velocity errors.4.4 This procedure explains the calibration of a fathometeror electronic depth sounder using a bar-check.4.5 Bar-checking techniques and equipment are general innature and may need to be modified for use in specific fieldconditions.1This practice is under the jurisdic
14、tion of ASTM Committee D19 on Water andis the direct responsibility of Subcommittee D19.07 on Sediments, Geomorphology,and Open-Channel Flow.Current edition approved Jan. 1, 2014. Published March 2014. Originallyapproved in 1998. Last previous edition approved in 2008 as D6318 03 (2008).DOI: 10.1520
15、/D6318-03R14.2For referenced ASTM standards, visit the ASTM website, www.astm.org, orcontact ASTM Customer Service at serviceastm.org. For Annual Book of ASTMStandards volume information, refer to the standards Document Summary page onthe ASTM website.Copyright ASTM International, 100 Barr Harbor Dr
16、ive, PO Box C700, West Conshohocken, PA 19428-2959. United States15. Apparatus5.1 The device used for bar-checking must be a sound-reflecting surface that can be lowered to a known depth belowthe transducer of the survey vessel. See Fig. 1. These sounding-refecting surfaces (or sounding targets) can
17、 be a bar made outof a section of metallic I-beam or T-beam, pipe, a rectangularsection of sheet metal, or a section of metal screen.5.2 Bars used in depths greater than 30 ft (10 m) should beat least 9 in. (23 cm) wide. The dimensions of the target dependon the type of survey vessel, location of th
18、e transducer, and thedepth range to be covered during the survey. Usually, thelength of the bar is equal to the beam or width for small surveyvessels. For larger vessels, a spherical metal ball or steel plateis lowered through a well in the hull.5.3 The weight of the bar will be dependent on the typ
19、e ofcurrents experienced, typical project depths, and beam of thevessel. Typical weights range from 40 to 100 lb (20 to 50 kg).In deep water areas with large currents, a heavy bar is essentialbecause subsurface currents will pull to light a bar from thetransducers vertical plane. On small, shallow i
- 1.请仔细阅读文档,确保文档完整性,对于不预览、不比对内容而直接下载带来的问题本站不予受理。
- 2.下载的文档,不会出现我们的网址水印。
- 3、该文档所得收入(下载+内容+预览)归上传者、原创作者;如果您是本文档原作者,请点此认领!既往收益都归您。
下载文档到电脑,查找使用更方便
5000 积分 0人已下载
下载 | 加入VIP,交流精品资源 |
- 配套讲稿:
如PPT文件的首页显示word图标,表示该PPT已包含配套word讲稿。双击word图标可打开word文档。
- 特殊限制:
部分文档作品中含有的国旗、国徽等图片,仅作为作品整体效果示例展示,禁止商用。设计者仅对作品中独创性部分享有著作权。
- 关 键 词:
- ASTMD6318200320143183STANDARDPRACTICEFORCALIBRATINGAFATHOMETERUSINGABARCHECKMETHOD 采用 检验 校准 回声 测深 标准

链接地址:http://www.mydoc123.com/p-522155.html