ASTM D5110-1998(2004) Standard Practice for Calibration of Ozone Monitors and Certification of Ozone Transfer Standards Using Ultraviolet Photometry《使用紫外光度法进行臭氧转变标准认证和臭氧监测器校准的标准实施规.pdf
《ASTM D5110-1998(2004) Standard Practice for Calibration of Ozone Monitors and Certification of Ozone Transfer Standards Using Ultraviolet Photometry《使用紫外光度法进行臭氧转变标准认证和臭氧监测器校准的标准实施规.pdf》由会员分享,可在线阅读,更多相关《ASTM D5110-1998(2004) Standard Practice for Calibration of Ozone Monitors and Certification of Ozone Transfer Standards Using Ultraviolet Photometry《使用紫外光度法进行臭氧转变标准认证和臭氧监测器校准的标准实施规.pdf(8页珍藏版)》请在麦多课文档分享上搜索。
1、Designation: D 5110 98 (Reapproved 2004)Standard Practice forCalibration of Ozone Monitors and Certification of OzoneTransfer Standards Using Ultraviolet Photometry1This standard is issued under the fixed designation D 5110; the number immediately following the designation indicates the year oforigi
2、nal adoption or, in the case of revision, the year of last revision. A number in parentheses indicates the year of last reapproval. Asuperscript epsilon (e) indicates an editorial change since the last revision or reapproval.1. Scope1.1 This practice covers a means for calibrating ambient,workplace,
3、 or indoor ozone monitors, and for certifyingtransfer standards to be used for that purpose.1.2 This practice describes means by which dynamicstreams of ozone in air can be designated as primary ozonestandards.1.3 This standard does not purport to address all of thesafety concerns, if any, associate
4、d with its use. It is theresponsibility of the user of this standard to establish appro-priate safety and health practices and determine the applica-bility of regulatory limitations prior to use. See Section 8 forspecific precautionary statements.2. Referenced Documents2.1 ASTM Standards:2D 1356 Ter
5、minology Relating to Sampling and Analysis ofAtmospheresD 3195 Practice for Rotameter CalibrationD 3249 Practice for General Ambient Air Analyzer Proce-duresD 3631 Test Methods for Measuring Surface AtmosphericPressureD 5011 Practices for Calibration of Ozone Monitors UsingTransfer StandardsE 220 Me
6、thod for Calibration of Thermocouples By Com-parison Techniques3E 591 Practice for Safety and Health Requirements Relatingto Occupational Exposure to Ozone3E 644 Test Methods for Testing Industrial Resistance Ther-mometers3. Terminology3.1 For definitions of terms used in this practice, refer toTerm
7、inology D 1356.3.2 Definitions of Terms Specific to This Standard:3.2.1 primary standarda standard directly defined andestablished by some authority, against which all secondarystandards are compared.3.2.2 secondary standarda standard used as a means ofcomparison, but checked against a primary stand
8、ard.3.2.3 standardan accepted reference sample or deviceused for establishing measurement of a physical quantity.3.2.4 transfer standarda type of secondary standard. It isa transportable device or apparatus that, together with opera-tional procedures, is capable of reproducing pollutant concen-trati
9、on or producing acceptable assays of pollutant concentra-tions.3.2.5 zero airpurified air that does not contain ozone, anddoes not contain any other component that may interfere withthe measurement (see 7.1).4. Summary of Practice4.1 This practice is based on the photometric assay of ozone(O3) conce
10、ntrations in a dynamic flow system. The concentra-tion of O3in an absorption cell is determined from a measure-ment of the amount of 253.7 nm light absorbed by the sample.This determination requires knowledge of (1)4the absorptioncoefficient of O3at 253.7 nm, (2) the optical path lengththrough the s
11、ample, (3) the transmittance of the sample at awavelength of 253.7 nm, and (4) the temperature and pressureof the sample. The transmittance is defined as the ratio:I/Iowhere:I = the intensity of light that passes through the cell and issensed by the detector when the cell contains an O3sample, and1T
12、his practice is under the jurisdiction of ASTM Committee D22 on Samplingand Analysis of Atmospheres and is the direct responsibility of SubcommitteeD22.03 on Ambient Atmospheres and Source Emissions.Current edition approved October 1, 2004. Published December 2004. Originallyapproved in 1990. Last p
13、revious edition approved in 1998 as D 5110 - 98.2For referenced ASTM standards, visit the ASTM website, www.astm.org, orcontact ASTM Customer Service at serviceastm.org. For Annual Book of ASTMStandards volume information, refer to the standards Document Summary page onthe ASTM website.3Withdrawn.4T
14、he boldface numbers in parentheses refer to the references listed at the end ofthis practice.1Copyright ASTM International, 100 Barr Harbor Drive, PO Box C700, West Conshohocken, PA 19428-2959, United States.Io= the intensity of light that passes through the cell and issensed by the detector when th
15、e cell contains zero air.It is assumed that all conditions of the system, except for thecontents of the absorption cell, are identical during measure-ments of I and Io. The quantities defined above are related bythe Beer-Lambert absorption law:Transmittance5I/Io5e2acd(1)where:a = absorption coeffici
16、ent of O3at 253.7 nm,(308 6 4) 3 106ppm1cm1at 0C and 101.3 kPa (1atm) (1, 2, 3, 4, 5, 6, 7, 8)c =O3concentration, ppm, andd = optical path length, cm.4.1.1 In practice, a stable O3generator (see 6.1.4) is used toproduce O3concentrations over the required range. Each O3concentration is determined fro
17、m the measurement of thetransmittance of the sample at 253.7 nm, and is calculated fromthe equation:c52lnIIoad!(2)The calculated O3concentrations must be corrected for O3losses, which may occur in the photometer, and for thetemperature and pressure of the sample.5. Significance and Use5.1 The reacti
18、vity and instability of O3preclude the storageof O3concentration standards for any practical length of time,and precludes direct certification of O3concentrations asStandard Reference Materials (SRMs). Moreover, there is noavailable SRM that can be readily and directly adapted to thegeneration of O3
19、standards analogous to permeation devicesand standard gas cylinders for sulfur dioxide and nitrogenoxides. Dynamic generation of O3concentrations is relativelyeasy with a source of ultraviolet (UV) radiation. However,accurately certifying an O3concentration as a primary standardrequires assay of the
20、 concentration by a comprehensivelyspecified analytical procedure, which must be performed everytime a standard is needed (9).5.2 This practice is not designed for the routine calibrationof O3monitors at remote locations (see Practices D 5011).6. Apparatus6.1 Atypical complete UV calibration system
21、consists of anO3generator, an output port or manifold, a photometer, asource of zero air, and other components as necessary. Theconfiguration must provide a stable O3concentration at thesystem output and allow the photometer to assay accurately theoutput concentration to the precision specified for
22、the photom-eter. Fig. 1 shows the system, and illustrates the calibrationsystem. Ozone is highly reactive and subject to losses uponcontact with surfaces. All components between the O3genera-tor and the photometer absorption cell shall be of inertmaterial, such as glass or TFE-fluorocarbon. Lines an
23、d inter-connections shall be as short as possible, and all surfaces shallbe chemically clean. For certification of transfer standards thatprovide their own source of O3, the generator and possiblyother components shown in Fig. 1 may not be required (seePractices D 5011).6.1.1 UV Photometer, consisti
24、ng of a low-pressure mercurydischarge lamp, collimation optics (optional), an absorptioncell, a detector, and signal-processing electronics, as shown inFig. 1. It shall be capable of measuring the transmittance, I/Io,at a wavelength of 253.7 nm with sufficient precision that thestandard deviation of
- 1.请仔细阅读文档,确保文档完整性,对于不预览、不比对内容而直接下载带来的问题本站不予受理。
- 2.下载的文档,不会出现我们的网址水印。
- 3、该文档所得收入(下载+内容+预览)归上传者、原创作者;如果您是本文档原作者,请点此认领!既往收益都归您。
下载文档到电脑,查找使用更方便
10000 积分 0人已下载
下载 | 加入VIP,交流精品资源 |
- 配套讲稿:
如PPT文件的首页显示word图标,表示该PPT已包含配套word讲稿。双击word图标可打开word文档。
- 特殊限制:
部分文档作品中含有的国旗、国徽等图片,仅作为作品整体效果示例展示,禁止商用。设计者仅对作品中独创性部分享有著作权。
- 关 键 词:
- ASTMD511019982004STANDARDPRACTICEFORCALIBRATIONOFOZONEMONITORSANDCERTIFICATIONOFOZONETRANSFERSTANDARDSUSINGULTRAVIOLETPHOTOMETRY

链接地址:http://www.mydoc123.com/p-518890.html