《ASTM C1477-2008 Standard Test Method for Isotopic Abundance Analysis of Uranium Hexafluoride and Uranyl Nitrate Solutions by Multi-Collector Inductively Coupled Plasma-Mass Spectro.pdf》由会员分享,可在线阅读,更多相关《ASTM C1477-2008 Standard Test Method for Isotopic Abundance Analysis of Uranium Hexafluoride and Uranyl Nitrate Solutions by Multi-Collector Inductively Coupled Plasma-Mass Spectro.pdf(6页珍藏版)》请在麦多课文档分享上搜索。
1、Designation: C 1477 08Standard Test Method forIsotopic Abundance Analysis of Uranium Hexafluoride andUranyl Nitrate Solutions by Multi-Collector, InductivelyCoupled Plasma-Mass Spectrometry1This standard is issued under the fixed designation C 1477; the number immediately following the designation i
2、ndicates the year oforiginal adoption or, in the case of revision, the year of last revision. A number in parentheses indicates the year of last reapproval. Asuperscript epsilon () indicates an editorial change since the last revision or reapproval.1. Scope1.1 This test method covers the isotopic ab
3、undance analysisof234U,235U,236U and238U in samples of hydrolysed uraniumhexafluoride (UF6) by inductively coupled plasma source,multicollector, mass spectrometry (ICP-MC-MS). The methodapplies to material with235U abundance in the range of 0.2 to6 % mass. This test method is also described in ASTMS
4、TP 1344.1.2 The values stated in SI units are to be regarded asstandard. No other units of measurement are included in thisstandard.1.3 This standard does not purport to address all of thesafety concerns, if any, associated with its use. It is theresponsibility of the user of this standard to establ
5、ish appro-priate safety and health practices and determine the applica-bility of regulatory limitations prior to use.2. Referenced Documents2.1 ASTM Standards:2C 761 Test Methods for Chemical, Mass Spectrometric,Spectrochemical, Nuclear, and Radiochemical Analysis ofUranium HexafluorideC 787 Specifi
6、cation for Uranium Hexafluoride for Enrich-mentC 996 Specification for Uranium Hexafluoride Enriched toLess Than 5 %235UD 1193 Specification for Reagent Water2.2 Other Document:STP 1344 Applications of Inductively Coupled Plasma-Mass Spectrometry (ICP-MS) to Radionuclide Determi-nations33. Terminolo
7、gy3.1 Acronyms:3.1.1 amuatomic mass unit3.1.2 ICP-MC-MSInductively Coupled Plasma Multi-Collector Mass Spectrometer3.1.3 ICP-MSInductively Coupled Plasma Mass Spec-trometer3.1.4 UIRMUranium Isotopic Reference Material4. Summary of Test Method4.1 Samples are received either in the form of uraniumhexa
8、fluoride (UF6) or aqueous uranic solution. The UF6samples are hydrolysed, diluted and acidified with nitric acid.Uranic solution samples are diluted and acidified with nitricacid. If required, an internal reference of thorium isotopes canbe subsequently added to each diluted sample. As detailed inSe
9、ction 8, isotope pairs of elements other than thorium couldbe used for an internal reference.4.2 The samples are contained in polypropylene tubes thatare inserted into the auto-sampler rack of the mass spectrom-eter. Sample details are input to the computer and the instru-ment is prepared for measur
10、ement. The automatic measuringsequence is initiated.4.3 Uranium Isotopic Reference Materials (UIRMs) areused to calibrate the instrument. Each UIRM is prepared inaqueous solution (acidified with nitric acid) and if requiredspiked with the same internal reference as the samples. Thiscalibration solut
11、ion is measured and a mass bias parameter iscalculated that is stored and subsequently imported into each ofthe sample measurements4to correct the measured uraniumisotopic ratios.4.4 Measurements of isotopic ratios in the calibration solu-tion and the subsequent samples are initiated by customisedso
12、ftware. The mass bias factor is computed from the measuredisotopic ratios in the calibration solution. This parameter isthen exported to correct the measured isotopic ratios of thesamples for mass bias. The corrected isotopic abundances are1This test method is under the jurisdiction of ASTM Committe
13、e C26 on NuclearFuel Cycle and is the direct responsibility of Subcommittee C26.05 on Methods ofTest.Current edition approved June 1, 2008. Published August 2008. Originallyapproved in 2000. Last previous edition approved in 2006 as C 1477 06.2For referenced ASTM standards, visit the ASTM website, w
14、ww.astm.org, orcontact ASTM Customer Service at serviceastm.org. For Annual Book of ASTMStandards volume information, refer to the standards Document Summary page onthe ASTM website.3Available from ASTM Headquarters.4The uranium isotopic precision of measurement, limit of detection anduncertainty of
15、 measurement are listed in Section 15 and Appendix X1.1Copyright ASTM International, 100 Barr Harbor Drive, PO Box C700, West Conshohocken, PA 19428-2959, United States.expressed as % atomic and are converted to % mass prior toreporting. Details of the mass bias correction are presented inAppendix X
16、1.5. Significance and Use5.1 The test method is capable of measuring uraniumisotopic abundances of234U,235U,236U and238U as requiredby Specifications C 787 and C 996.6. Interferences6.1 Mass BiasElectrostatic repulsion between uraniumions causes a so-called “mass bias” effect. Mass bias isobserved a
17、s an enhancement in the number of ions detected atthe collectors from the heavier uranium isotopes relative to thelighter uranium isotopes. A calibration procedure is used tocorrect the mass spectrometer for mass bias.6.2 Adjacent Isotopic PeaksThe abundance sensitivity ofthe ICP-MC-MS at mass 237 i
18、s specified to be less than 0.5parts per million of the238U ion beam. The method is limitedto the measurement of235U isotopic abundances below 6 %,consequently interference effects with the234U and236U ionbeams are negligible.6.3 Isobaric Molecular InterferencesA molecular inter-ference exists at ma
19、ss 236 between236U and a hydride of235U,which is formed in the plasma. This interference can becorrected by measuring the beam height of the238U hydride atmass 239, and applying the correction defined in Eq 1, to themeasured236U ion beam:5236Uc5236UmS235U 3238UH238UD(1)where:236Uc= the corrected236U
20、 ion beam,236Um= the measured236U ion beam,235U = the measured235U ion beam,238UH = the measured238U hydride ion beam, and238U = the measured238U ion beam.6.4 Memory Effects:6.4.1 Contamination of the sample introduction systemfrom previous samples produces memory interference effects.Such effects a
21、re accentuated when samples that are depletedin235U are measured after enriched samples. Memory effectscan be readily assessed by aspirating a 0.3 M nitric acidsolution and measuring the background238U ion beam. Thesample introduction system should be periodically disas-sembled and cleaned, to minim
22、ise the background238U ionbeam.6.4.2 A background correction is performed during themeasurement run by monitoring the analyte signals of the 0.3M nitric acid rinse solution. The background correction ismeasured prior to the mass calibration and is re-measuredbefore each subsequent sample.7. Apparatu
23、s7.1 Mass Spectrometer:7.1.1 The mass spectrometer has an inductively coupledplasma (ICP) source and a double focusing electrostatic/magnetic sector analyser equipped with twelve Faraday detec-tors and two ion counters.67.1.2 The mass spectrometer is fully computer controlledusing customised softwar
24、e and is equipped with an auto-sampler.7.2 Polypropylene Sample Tubes, Screw-Cap,50mL.7.3 Polypropylene Sample Tubes, Screw-Cap,10mL.7.4 Positive Displacement Pipette, and Tips to Suit, 0.01mL.7.5 Positive Displacement Pipette, and Tips to Suit, 1 mL.7.6 Variable-Volume Dispenser, 1 to 5 mL, fitted
- 1.请仔细阅读文档,确保文档完整性,对于不预览、不比对内容而直接下载带来的问题本站不予受理。
- 2.下载的文档,不会出现我们的网址水印。
- 3、该文档所得收入(下载+内容+预览)归上传者、原创作者;如果您是本文档原作者,请点此认领!既往收益都归您。
下载文档到电脑,查找使用更方便
10000 积分 0人已下载
下载 | 加入VIP,交流精品资源 |
- 配套讲稿:
如PPT文件的首页显示word图标,表示该PPT已包含配套word讲稿。双击word图标可打开word文档。
- 特殊限制:
部分文档作品中含有的国旗、国徽等图片,仅作为作品整体效果示例展示,禁止商用。设计者仅对作品中独创性部分享有著作权。
- 关 键 词:
- ASTMC14772008STANDARDTESTMETHODFORISOTOPICABUNDANCEANALYSISOFURANIUMHEXAFLUORIDEANDURANYLNITRATESOLUTIONSBYMULTICOLLECTORINDUCTIVELYCOUPLEDPLASMAMASSSPECTROPDF

链接地址:http://www.mydoc123.com/p-464784.html