1、 Beneficial comments, recommendations, additions, deletions, clarifications, etc. and any data that may improve this document should be sent to: WR-ALC/404 SCMS/GUEEA, Robins AFB GA 31098-1670. Since contact information can change, you may want to verify the currency of this address information usin
2、g the ASSIST Online database at https:/assist.dla.mil . AMSC N/A FSC 6685 DISTRIBUTION STATEMENT A. Approved for public release; distribution is unlimited. INCH-POUND A-A-59928A 18 JUL 2013 SUPERCEDING A-A-59928 26 JUNE 2013 COMMERCIAL ITEM DESCRIPTION METER, FUEL FLOW, PORTABLE The General Services
3、 Administration has authorized the use of this Commercial Item Description (CID) as a replacement for PD07WRGBZOENI27 Rev 3, for all federal agencies. 1. SCOPE. SCOPE: The test set covered by this specification is intended to be a portable fuel flow meter unit and eight test cables. The Fuel Flow Me
4、ter Test Set (here on referred to as test set or flow meter) is a support equipment test set which is used for depot and flight line test of fuel flow meter system components. The test set complies with Class 1 according to MIL-PRF-28800, is used to test flow meter transmitters, indicators of the se
5、cond harmonic type, total summation indicators, individual summation indicators and system power supplies. 2. SALIENT CHARACTERISTICS. 2.1 Components. The test set consists of a high brightness display, master transmitter assembly, linear transmitter, multi-meter, Test Selector Switch, Cables. Drawi
6、ngs of all cables will be provided. All cables and manuals shall fit within the case. COMPONENT Military Specification Item Type Indicator, Individual Rate MIL-I-25906 EFU-3A Indicator, Individual Summation MIL-I-25324 MIL-I-25907 MA-7A, MA-7 EFU4/A Indicator, Total Summation MIL-I-25325 MIL-I-25908
7、 MA-8 EFU-5/A Transmitter MIL-T-8249 MIL-T-25905 MIL-T-26298 MB-2 TRU-7/A TRU-6/A Performance Specification MIL-PRF-28800 Provided by IHSNot for ResaleNo reproduction or networking permitted without license from IHS-,-,-A-A-59928A 2 Cable Assembly Part Number Cage Code Length 5796297G700 See MIL-T-2
8、6298 TJ-49,TJ-51 9172332-10 98750 10 inches DJ-71 9172332-30 98750 10 inches DJ-72 9172332-50 98750 10 inches DJ-64 9172332-70 98750 10 inches TJ-50 9172332-110 98750 22 feet Main 9172332-130 98750 5 feet 5 inches TJ-51E 9172332-150 98750 10 inches GND-1 9172332-170 98750 10 feet DJ-71A 9172332-190
9、98750 10 inches TBD Used to drive all 8 FFIs at the same time B-52 requirement TBD Used to test indicators from engine connector B-52 requirement Function Selector/Switch Function Function Switch Allows test set to perform: Self Test; 4Hz Power Supply Check; Transmitter; Individual, Linear and total
10、 indicator. Mode Switch Allows displays in DEGREES, 12K PPH, 15K PPH, LINE FREQ DISP, STEP MODE Master XMTR Sets master transmitter to desired settings Linear XMTR PPH X 100 Checks scale error and operation of Type 8DJ72 total summation indicators. 2.2 Functional requirements. This test set shall be
11、, at minimum, capable of: a. Test the output signals of flow-meter transmitters of the second-harmonic type. b. Generate a calibration signal used to test flow-meter indicators of the second-harmonic type. c. Measure the output voltage of individual summation indicators. d. Generate a calibration vo
12、ltage used to test total summation indicators. e. Check output continuity of flow-meter system power supplies. f. Measure output frequency of flow-meter system power supplies. g. Isolate the malfunction of a defective system component through substitution of individual components. h. Identify the fa
13、iled Components. Per X-Technologies, there is a user desire to move the tester internal Circuit Breaker, exhibited by existing units, to the front panel to the location currently occupied by an unused 28 VDC input connector, and delete the associated 28 VDC connector, modules and circuitry. This wou
14、ld allow the CB to be easily reset. Per X-Technologies, none of the platforms use the 28 VDC function. If true, this 28 VDC function could be removed from applicable paragraphs below. Provided by IHSNot for ResaleNo reproduction or networking permitted without license from IHS-,-,-A-A-59928A 3 2.3 C
15、arrying Case. The carrying case shall be 24 X 14 X 22 inch or less. A handle for carrying the test set shall be attached to the case. A mechanism for locking the test set shall be provided. The weight of the test set including the test cables shall be 35 pounds or less. The case shall be airtight an
16、d a mechanism will be added for equalizing air pressure within the enclosure, when the enclosure is sealed. The mechanism provided shall not protrude beyond the enclosure. 2.4 Interface. An interface shall be provided. It shall display settings and test values in Degrees, flow rate in Pounds Per Hou
17、r(PPH), input voltage, the scale error of the flow meter indicators of the second-harmonic type, the scale error and operation of Total Summation Indicators, output voltage of individual summation indicators, frequency and output of the flow meter power supply, test selector positions. 2.5 Tested it
18、ems: The test set shall test the following transmitters and indicators which were built according to the mil-specs: COMPONENT Military Specification Indicator, Individual Rate MIL-I-25906 Indicator, Individual Summation MIL-I-25324 MIL-I-25907 Indicator, Total Summation MIL-I-25325 MIL-I-25908 Trans
19、mitter MIL-T-8249 MIL-T-25905 Performance Specification MIL-PRF-28800 2.5.1 Testing transmitters. The test set will be testing transmitters designated type MB-2, TRU-7/A, TRU-/A. Info on such part can be found in Mil-T-8249, MIL-T-25905, and MIL-T-26298 respectively. The test set will be powered fro
20、m the aircraft through a cable from Mil Spec MS33678-14S-7P (P/N 9172332-130, Cage Code 98750) or 120V and 60Hz or 28V DC power supply. MS33678-14S-2P, 9172332-110, 98750, will be used for connection to the flow rate indicator. For bench testing, the system will use wall voltage. As the flow through
21、 the transmitter is increased and decreased by operating the engine, or through a controlled environment, the test set is used to verify that the transmitter is functioning and that the voltage output varies when the fuel flow rate changes. The test set shall have a function, XMTR that is used to ch
22、eck the operation of flow meter transmitters when installed in a flow meter system or to check transmitter accuracy when mounted on a test stand. The harmonic modulated outputs of the flow meter transmitter are demodulated by the master indicator circuit on the test set. Data is displayed in degrees
23、 to two decimal places or in pounds-per-hour for 12,000 and 15,000 PPH systems as selected by the user from the mode switch. 2.5.2 Testing indicators of the second-harmonic type. The test set will be testing indicators designated type EFU-3IA. Info on such part can be found MIL-I-25906. Provided by
24、IHSNot for ResaleNo reproduction or networking permitted without license from IHS-,-,-A-A-59928A 4 The test set will be powered from the aircraft through a cable from Mil Spec MS33678-14S-7P (P/N 9172332-130, Cage Code 98750 or wall voltage, 120V and 60Hz or 28V DC power supply) and connected throug
- 1.请仔细阅读文档,确保文档完整性,对于不预览、不比对内容而直接下载带来的问题本站不予受理。
- 2.下载的文档,不会出现我们的网址水印。
- 3、该文档所得收入(下载+内容+预览)归上传者、原创作者;如果您是本文档原作者,请点此认领!既往收益都归您。
下载文档到电脑,查找使用更方便
5000 积分 0人已下载
下载 | 加入VIP,交流精品资源 |
- 配套讲稿:
如PPT文件的首页显示word图标,表示该PPT已包含配套word讲稿。双击word图标可打开word文档。
- 特殊限制:
部分文档作品中含有的国旗、国徽等图片,仅作为作品整体效果示例展示,禁止商用。设计者仅对作品中独创性部分享有著作权。
- 关 键 词:
- AIRFORCEAA59928A2013METERFUELFLOWPORTABLEPDF

链接地址:http://www.mydoc123.com/p-427205.html