1、 INCH-POUND A-A-59322B October 14, 2009 Superseding A-A-59322A April 27, 2006 COMMERCIAL ITEM DESCRIPTION OHMMETER The general Services Administration has authorized the use of this Commercial Item description, for all federal agencies. 1 SCOPE. This Commercial Item Description (CID) describes an Oh
2、mmeter herein referred to as equipment. This CID is meant as a minimum requirement in which only those manufactures meeting or surpassing the following requirements are to be supplied per this CID. 2 SALIENT CHARACTERISTICS. The equipment shall be capable of operation within the accuracies, limits,
3、and specifications herein. Equipment covered by this CID may be commercially available equipment modified to the extent necessary to meet the following description. The equipment assembly shall be Class 1 as described in the classification and class paragraphs of the latest version of MIL-PRF-28800
4、except as specified herein. 2.1 Size. The equipment dimensions shall permit efficient portable usage. The equipments weight, including any and all accessories (excluding written manuals) shall not exceed 20 pounds. 2.2 Safety and Environmental. The equipment shall meet all safety and environmental r
5、equirements as specified in MIL-PRF-28800 for the classification as stated herein. 2.2.1 Water Resistance. The equipment shall not be required to be watertight and splash proof. Drip proof requirements shall conform to MIL-PRF-28800, Class 2 while not operating and covers are closed. 2.2.2 Explosive
6、 Atmosphere. The equipment shall meet the explosive atmosphere requirements of MIL-PRF-28800 or the requirements of MIL-STD-810, METHOD 511. FSC 6625 Beneficial comments, recommendations, additions, deletions, clarifications, etc. and any other data that may improve this document should be sent to W
7、R-ALC/742 CBSG/GBEC, 450 3rd St. Bldg 323, Robins AFB, GA 31098-1611. Since contact information can change, you may want to verify the currency of this address information using the ASSIST Online database at http:/assist.daps.dla.mil . Provided by IHSNot for ResaleNo reproduction or networking permi
8、tted without license from IHS-,-,-A-A-59322B 2.2.3 Dust Resistance. Dust resistance testing shall not be required. 2.2.4 Salt Exposure. Salt exposure verification testing shall not be required. 2.2.5 Temperature Operating. The equipment shall meet its performance and accuracy requirements in an oper
9、ating environment of at least 0 C to +40 C. 2.2.6 Transit Drop. The transit drop requirements shall conform to MIL-PRF-28800, Class 2 2.2.7 Random Vibration. The random vibration requirements shall conform to MIL-PRF-28800, Class 2. 2.3 Reliability. Unless otherwise specified, the design of the equi
10、pment shall be such that under normal use and operation the equipment does not fail within 2000 hours of operation with a statistical certainty of 95%. 2.4 Calibration and Maintenance Adjustments. The design of the system shall provide for readily accessible calibration and maintenance adjustments.
11、These adjustments shall be provided by variable value components, which are adjustable, by the use of simple means. The calibration by substitution of selected components or parts is unacceptable unless specifically approved. The calibration adjustments, wherever possible, shall be accessible withou
12、t removal of the instrument case or modules. The calibration interval shall be a period of one year or greater, based on an operating time of 2000 hours per year. 2.5 Electrical Power Sources and Connections. The equipment shall operate from an internal power source. Batteries shall be provided with
13、 the equipment. Batteries shall be easily accessible for maintenance and weatherproof from the environment. The equipment shall operate in accordance with the steady state conditions, transient state conditions, and interruption of power source requirements outlined the latest version of MIL-PRF-288
14、00. 2.6 Internal DC Power. DC power shall be supplied by internal batteries. The equipment shall be capable of operating away from AC power source. The batteries shall have the capability of operating the unit with full performance for a period of not less than eight hours. 2.7 Accessories. Accessor
15、ies to be supplied by the manufacturer and listed in the manufacturers catalog or data sheet, a written tutorial/user reference manual, a traceable calibration certificate, and a complete set of test leads with provisions for storage. 2.8 Warm-up Performance. The warm-up requirement shall be met whe
- 1.请仔细阅读文档,确保文档完整性,对于不预览、不比对内容而直接下载带来的问题本站不予受理。
- 2.下载的文档,不会出现我们的网址水印。
- 3、该文档所得收入(下载+内容+预览)归上传者、原创作者;如果您是本文档原作者,请点此认领!既往收益都归您。
下载文档到电脑,查找使用更方便
2000 积分 0人已下载
下载 | 加入VIP,交流精品资源 |
- 配套讲稿:
如PPT文件的首页显示word图标,表示该PPT已包含配套word讲稿。双击word图标可打开word文档。
- 特殊限制:
部分文档作品中含有的国旗、国徽等图片,仅作为作品整体效果示例展示,禁止商用。设计者仅对作品中独创性部分享有著作权。
- 关 键 词:
- AIRFORCEAA59322B2009OHMMETER 欧姆 PDF