《ASTM E2063-12(2017) Standard Practice for Calibration and Functionality Checks Used in Forensic Psychophysiological Detection of Deception (Polygraph) Examinations.pdf》由会员分享,可在线阅读,更多相关《ASTM E2063-12(2017) Standard Practice for Calibration and Functionality Checks Used in Forensic Psychophysiological Detection of Deception (Polygraph) Examinations.pdf(2页珍藏版)》请在麦多课文档分享上搜索。
1、Designation: E2063 12 (Reapproved 2017)Standard Practice forCalibration and Functionality Checks Used in ForensicPsychophysiological Detection of Deception (Polygraph)Examinations1This standard is issued under the fixed designation E2063; the number immediately following the designation indicates th
2、e year oforiginal adoption or, in the case of revision, the year of last revision. A number in parentheses indicates the year of last reapproval. Asuperscript epsilon () indicates an editorial change since the last revision or reapproval.1. Scope1.1 This practice provides guidelines for calibration
3、andinstrumentation (both analog and computerized systems) ac-tively used in the psychophysiological detection of deception(verification of truth). As a minimum, such instrumentationshall simultaneously record an individuals respiratory, exoso-matic electrodermal, and cardiovascular activity.1.2 Anal
4、og polygraphs shall be calibrated by the psycho-physiological detection of deception (PDD) examiner,manufacturer, or factory-authorized individual.1.3 Computerized instrumentation shall be calibrated by themanufacturer or factory-authorized individual.1.4 The PDD examiner or factory-authorized indiv
5、idualshall perform functionality checks to ensure instrumentation isoperating properly.1.5 This practice does not prohibit additional componentswhich may be offered as supplemental measurements ofphysiological change. Additional recording components (suchas movement sensors) may be used in addition
6、to, but notreplace, the required minimum components and these addi-tional components shall meet the manufacturers specificationsfor calibration.1.6 This international standard was developed in accor-dance with internationally recognized principles on standard-ization established in the Decision on P
7、rinciples for theDevelopment of International Standards, Guides and Recom-mendations issued by the World Trade Organization TechnicalBarriers to Trade (TBT) Committee.2. Referenced Documents2.1 ASTM Standards:2E1954 Practice for Conduct of Research in Psychophysi-ological Detection of Deception (Pol
8、ygraph)E2000 Guide for Minimum Basic Education and Training ofIndividuals Involved in the Detection of Deception (PDD)(Withdrawn 2014)3E2035 Terminology Relating to Forensic Psychophysiology2.2 Other Document:Manufacturer Manual(s) for System(s) in Use3. Terminology3.1 Definitions of TermsSee Termin
9、ology E2035.3.2 Terminology may vary according to different manufac-turers.4. Significance and Use4.1 This practice sets forth the minimum requirements forcalibration and functionality checks when conducting PDDexaminations and related activities. For additional informationsee Practice E1954 and Gui
10、de E2000.5. Minimum Requirements for Calibration andFunctionality Checks of Polygraph Instrumentation5.1 A recorded chart shall be created demonstrating correctfunctioning of the instrument. This chart will be maintained fora minimum period of one year.5.1.1 This chart shall contain the following in
11、formation, asapplicable: name of person performing calibration or function-ality check, time and date, location, manufacturer, model, andinstrument identification.1This practice is under the jurisdiction of ASTM Committee E52 on ForensicPsychophysiology and is the direct responsibility of Subcommitt
12、ee E52.02 onInstrumentation.Current edition approved Oct. 1, 2017. Published October 2017. Originallyapproved in 2000. Last previous edition approved in 2012 as E2063 12. DOI:10.1520/E2063-12R17.2For referenced ASTM standards, visit the ASTM website, www.astm.org, orcontact ASTM Customer Service at
- 1.请仔细阅读文档,确保文档完整性,对于不预览、不比对内容而直接下载带来的问题本站不予受理。
- 2.下载的文档,不会出现我们的网址水印。
- 3、该文档所得收入(下载+内容+预览)归上传者、原创作者;如果您是本文档原作者,请点此认领!既往收益都归您。
下载文档到电脑,查找使用更方便
5000 积分 0人已下载
下载 | 加入VIP,交流精品资源 |
- 配套讲稿:
如PPT文件的首页显示word图标,表示该PPT已包含配套word讲稿。双击word图标可打开word文档。
- 特殊限制:
部分文档作品中含有的国旗、国徽等图片,仅作为作品整体效果示例展示,禁止商用。设计者仅对作品中独创性部分享有著作权。
- 关 键 词:
- ASTM E2063 12 2017 Standard Practice for Calibration and Functionality Checks Used in Forensic Psychophysiological Detection of Deception Polygraph Examinations

链接地址:http://www.mydoc123.com/p-288032.html