1、 IT IS THE USERS RESPONSIBILITY TO ESTABLISH APPROPRIATE PRECAUTIONARY PRACTICES AND TO DETERMINE THE APPLICABILITY OF REGULATORY LIMITATIONS PRIOR TO USE. EFFECTIVE HEALTH AND SAFETY PRACTICES ARE TO BE FOLLOWED WHEN UTILIZING THIS PROCEDURE. FAILURE TO UTILIZE THIS PROCEDURE IN THE MANNER PRESCRIB
2、ED HEREIN CAN BE HAZARDOUS. SAFETY DATA SHEETS (SDS) OR EXPERIMENTAL SAFETY DATA SHEETS (ESDS) FOR ALL OF THE MATERIALS USED IN THIS PROCEDURE SHOULD BE REVIEWED FOR SELECTION OF THE APPROPRIATE PERSONAL PROTECTION EQUIPMENT (PPE). COPYRIGHT 2014 UOP LLC. All rights reserved. Nonconfidential UOP Met
3、hods are available from ASTM International, 100 Barr Harbor Drive, PO Box C700, West Conshohocken, PA 19428-2959, United States. The UOP Methods may be obtained through the ASTM website, www.astm.org, or by contacting Customer Service at serviceastm.org, 610.832.9555 FAX, or 610.832.9585 PHONE. Trac
4、e Silicon in Petroleum Liquids by ICP-MS UOP Method 1006-14 Scope This method is for the determination of silicon (Si) in petroleum liquids using an Inductively Coupled Plasma - Mass Spectrometer (ICP-MS). Silicon is typically present in petroleum liquids as soluble organic silicones, but the method
5、 will also detect other forms of silicon such as the generally insoluble inorganic silicon salts and silica. The method is primarily used to analyze reformer charge stocks for silicones, but may also be used for refinery charge stocks as heavy as vacuum gas oil (VGO). The lower limit of quantitation
6、 is less than 0.1 mg/kg (mass-ppm) as silicon. Less sensitive tests for trace silicon are UOP Method 796, “Silicon in Petroleum Liquids by ICP-OES,” (to 1 mg/kg) and ASTM Method D7757, “Silicon in Gasoline and Related Products by Monochromatic Wavelength Dispersive X-ray Fluorescence Spectrometry,”
7、(to 3 mg/kg). References ASTM Specification D1193, “Reagent Water,” www.astm.org ASTM Practice D7455, “Sample Preparation of Petroleum and Lubricant Products for Elemental Analysis,” www.astm.org ASTM Method D7757, “Silicon in Gasoline and Related Products by Monochromatic Wavelength Dispersive X-ra
8、y Fluorescence Spectrometry,” www.astm.org ASTM Practice D7876, “Sample Decomposition Using Microwave Heating (With or Without Prior Ashing) for Atomic Spectroscopic Elemental Determination in Petroleum Products and Lubricants,” www.astm.org UOP Method 796, “Silicon in Petroleum Liquids by ICP-OES,”
9、 www.astm.org UOP Method 999, “Precision Statements in UOP Methods,” www.astm.org Outline of Method The sample is treated intially with water and nitric acid to reduce the volatility of silicon, followed by hydrofluoric acid, after which the samples are digested. The digestate is quantitatively tran
10、sferred to an autosampler tube using purified water. Indium is added as a monitoring standard and the mixture is diluted to the desired volume. The concentration of silicon in the resulting solution is determined by ICP-MS. 2 of 11 1006-14 Apparatus References to catalog numbers and suppliers are in
11、cluded as a convenience to the method user. Other suppliers may be used. Balance, readable to 0.0001 g Beaker, PTFE, 600-mL, Fisher Scientific, Cat. No. 02-586-1K Bottles, polypropylene, narrow mouth, screw cap, 125-mL, VWR, Cat. No. 16067-066 Bottle, wash, VWR, Cat. No. 2402-0500, for deionized wat
12、er Cover, PTFE, Milestone, Cat. No. MCL0435, with o-ring, Cat. No. OR0138, used to cover the stainless steel vessel in the microwave oven Cylinders, graduated, Class B, 5-, 10-, 25-, and 100-mL, VWR, Cat. Nos. 14201-610, -462, -364, and -718, respectively Flasks, volumetric, polypropylene, Class B,
13、25-, 50-, 100-, and 1000-mL, VWR, Cat. Nos. 83008-964, -974, -984, and 83009-008, respectively; several of each required Fume hood Hot plate, VWR, Cat. No. 11301-006 ICP-MS, PerkinElmer, DRC-e, equipped with an AS10 autosampler Microwave oven, 15-vessel capacity, with closed digestion vessels, quart
14、z tubes and lids, CryoLab chiller, and computer, Milestone, UltraWAVE Pipet, automatic, 10-100-L, VWR, Cat. No. 89079-968; calibrated and/or certified according to manufacturers recommendations Pipet, automatic, 100-1000-L, VWR, Cat. No. 89079-974; calibrated and/or certified according to manufactur
15、ers recommendations Pipet, electronic, 110-mL, EDP 3-Plus, Rainin, Cat. No. SE3-10mL; calibrated and/or certified according to manufacturers recommendations Regulator, argon or nitrogen, two-stage, high purity, delivery pressure range 30-700 kPa (4-100 psi), Matheson Tri-Gas, Model 3122-580 Timer, c
16、apable of timing two hours, VWR, Cat. No. 62344-440 Watch glass, PTFE, 125-mm, Fisher Scientific, Cat. No. 02-617-1F Reagents and Materials References to catalog numbers and suppliers are included as a convenience to the method user. Other suppliers may be used. Unless otherwise specified, reference
17、s to water mean deionized water. Argon, 99.995% minimum purity Cleaning compound, Alconox detergent, VWR, Cat. No. 21835-032 Dry ice Gloves, neoprene/natural rubber, VWR, Cat. No. 32917-212 (size 10) Gloves, Viton, VWR, Cat. No. 89236-768 (size 10) Hydrofluoric acid, concentrated, trace metals grade
18、, VWR, Cat. No. EM-HX0628-7 3 of 11 1006-14 Hydrogen peroxide, 29-32%, VWR, Cat. No. VW3690-5 Indium metal solution, aqueous, 1000-g/mL, SPEX CertiPrep, Cat. No. PLIN2-2X, for use as a monitoring solution Indium metal stock solution, aqueous, 100-g/mL. To prepare 50 mL, pipet 5 mL of the 1000-g/mL i
19、ndium metal solution into a 50-mL volumetric flask. Fill to the mark with deionized water. Cap and invert several times to mix. The solution should remain stable for two months. Indium monitoring standard solution, 10-g/mL. To prepare 100 mL, pipet 10.0 mL of the 100-g/mL indium metal stock solution
20、 into a 100-mL volumetric flask. Fill to the mark with deionized water. Cap and invert several times to mix. Store in a 125-mL polypropylene bottle. The solution should remain stable for two months. Lubricant, dry spray, PTFE, W. W. Grainger, Cat. No. 4JB33, or local supply Nitric acid, concentrated
21、, trace metals grade, VWR, Cat. No. EM-NX0408-2 Nitric acid, 2%. Pipet 20 mL of concentrated nitric acid into a 1000-mL volumetric flask containing 800 mL of deionized water; fill to mark with deionized water. Cap and invert several times to mix. Nitrogen, 99.995% minimum purity, optional, see Micro
22、wave Digestion, Step 6 O-rings, replacement, for microwave oven, Milestone, Cat. No. OR0159 O-ring, top of chamber, replacement, Milestone, Cat. No. OR0170 Pipet, glass, dropping, VWR, Cat. No. 52950-206 Pipet tips, 10-mL, Rainin, Cat. No. RC-10 mL Pipet tips, 1-200-L, VWR, Cat. No. 53508-810 Pipet
23、tips, 100-1250-L, VWR, Cat. No. 53508-922 Silicon standard solution, 1000-g/mL, SPEX CertiPrep, Cat. Nos. PLSi9-2X. Sulfuric acid, concentrated, trace metals grade, VWR, Cat. No. EM-SX1248-6 Towels, paper, disposable, local supply Tubes, autosampler, 50 mL, with caps, PerkinElmer, Cat. No. BO193234
24、Vials, PTFE, 15-mL, Milestone, Cat. No. HB00086, with caps, Cat. No. DD00134 Water, deionized, ASTM D1193 Type III reagent grade water, unless otherwise specified Water, hot, tap Wipers, Kimwipes Ex-L, VWR, Cat. No. 21905-026 The following materials are not required to perform the analyses described
- 1.请仔细阅读文档,确保文档完整性,对于不预览、不比对内容而直接下载带来的问题本站不予受理。
- 2.下载的文档,不会出现我们的网址水印。
- 3、该文档所得收入(下载+内容+预览)归上传者、原创作者;如果您是本文档原作者,请点此认领!既往收益都归您。
下载文档到电脑,查找使用更方便
10000 积分 0人已下载
下载 | 加入VIP,交流精品资源 |
- 配套讲稿:
如PPT文件的首页显示word图标,表示该PPT已包含配套word讲稿。双击word图标可打开word文档。
- 特殊限制:
部分文档作品中含有的国旗、国徽等图片,仅作为作品整体效果示例展示,禁止商用。设计者仅对作品中独创性部分享有著作权。
- 关 键 词:
- UOP10062014TRACESILICONINPETROLEUMLIQUIDSBYICPMSPDF

链接地址:http://www.mydoc123.com/p-1074461.html