1、 IT IS THE USERS RESPONSIBILITY TO ESTABLISH APPROPRIATE PRECAUTIONARY PRACTICES AND TO DETERMINE THE APPLICABILITY OF REGULATORY LIMITATIONS PRIOR TO USE. EFFECTIVE HEALTH AND SAFETY PRACTICES ARE TO BE FOLLOWED WHEN UTILIZING THIS PROCEDURE. FAILURE TO UTILIZE THIS PROCEDURE IN THE MANNER PRESCRIB
2、ED HEREIN CAN BE HAZARDOUS. SAFETY DATA SHEETS (SDS) OR EXPERIMENTAL SAFETY DATA SHEETS (ESDS) FOR ALL OF THE MATERIALS USED IN THIS PROCEDURE SHOULD BE REVIEWED FOR SELECTION OF THE APPROPRIATE PERSONAL PROTECTION EQUIPMENT (PPE). COPYRIGHT 2014 UOP LLC. All rights reserved. Nonconfidential UOP Met
3、hods are available from ASTM International, 100 Barr Harbor Drive, PO Box C700, West Conshohocken, PA 19428-2959, United States. The UOP Methods may be obtained through the ASTM website, www.astm.org, or by contacting Customer Service at serviceastm.org, 610.832.9555 FAX, or 610.832.9585 PHONE. Trac
4、e Metals in Organics by ICP-MS UOP Method 1005-14 Scope This method is for determining the concentrations of aluminum (Al), arsenic (As), calcium (Ca), cadmium (Cd), chromium (Cr), cobalt (Co), copper (Cu), iron (Fe), gallium (Ga), lead (Pb), lithium (Li), magnesium (Mg), manganese (Mn), molybdenum
5、(Mo), nickel (Ni), palladium (Pd), phosphorus (P), platinum (Pt), potassium (K), sodium (Na), strontium (Sr), tin (Sn), titanium (Ti), vanadium (V), zinc (Zn), and zirconium (Zr), in organic matrices such as crude petroleum, asphalts, vacuum tower bottoms, vacuum gas oils, atmospheric gas oils, dies
6、el and jet fuels and their blending components, pyrolysis oils, and fatty acid derivatives by Inductively Coupled Plasma Mass Spectrometry (ICP-MS). The lower limits of quantitation for the above elements, are listed in Table 1. Determination of additional elements is possible if they are compatible
7、 with other analytes during digestion. Table 1 Lower Limits of Quantitation, mg/kg (mass-ppm) Al 0.02 K 0.03 Pd 0.01 As 0.01 Li 0.01 Pt 0.01 Ca 0.03 Mg 0.02 Sn 0.01 Cd 0.01 Mn 0.01 Sr 0.01 Co 0.01 Mo 0.01 Ti 0.03 Cr 0.01 Na 0.06 V 0.01 Cu 0.01 Ni 0.01 Zn 0.01 Fe 0.02 P 0.50 Zr 0.01 Ga 0.01 Pb 0.01 A
8、lternatively, many of the elements listed above can be determined using UOP Method 389, “Trace Metals in Organics by ICP-OES.” Some of the elements listed above may be analyzed by Atomic Absorption Spectroscopy (AAS). See UOP Method 391, “Trace Metals in Petroleum Products or Organics by AAS,” for s
9、pecific metals and their range of quantitation. Metals known to be non-volatile may be analyzed by UOP Method 407, “Trace Metals in Organics by Dry Ashing - ICP-OES.” Many of these elements can be determined in kerosene using ASTM Method D7111, “Determination of Trace Elements in Middle Distillate F
10、uels by Inductively Coupled Plasma Atomic Emission Spectrometry (ICP-AES).” References ASTM Specification D1193, “Reagent Water,” www.astm.org 2 of 11 1005-14 ASTM Method D7111, “Determination of Trace Elements in Middle Distillate Fuels by Inductively Coupled Plasma Atomic Emission Spectrometry (IC
11、P-AES),” www.astm.org ASTM Practice D7455, “Sample Preparation of Petroleum and Lubricant Products for Elemental Analysis,” www.astm.org ASTM Practice D7876, “Sample Decomposition Using Microwave Heating (With or Without Prior Ashing) for Atomic Spectroscopic Elemental Determination in Petroleum Pro
12、ducts and Lubricants,” www.astm.org UOP Method 389, “Trace Metals in Organics by ICP-OES,” www.astm.org UOP Method 391, “Trace Metals in Petroleum Products or Organics by AAS,” www.astm.org UOP Method 407, “Trace Metals in Organics by Dry Ashing - ICP-OES,” www.astm.org UOP Method 999, “Precision St
13、atements in UOP Methods,” www.astm.org Outline of Method The sample is treated intially with water, then nitric and hydrofluoric acids, to reduce the volatility of some of the metals. Then the samples are digested. The digestate is quantitatively transferred to a polypropylene flask using purified w
14、ater. Scandium and bismuth are added as internal standards and the mixture is diluted to the desired volume. The concentrations of elements in the resulting solutions are determined by ICP-MS. Apparatus References to catalog numbers and suppliers are included as a convenience to the method user. Oth
15、er suppliers may be used. Balance, readable to 0.0001 g Beaker, PTFE, 600-mL, Fisher Scientific, Cat. No. 02-586-1K Bottles, polypropylene, narrow mouth, screw cap, 125-mL, VWR, Cat. No. 16067-066 Bottle, wash, VWR, Cat. No. 2402-0500, for deionized water Cover, PTFE, Milestone, Cat. No. MCL0435, wi
16、th o-ring, Cat. No. OR0138, used to cover the stainless steel vessel in the microwave oven Cylinders, graduated, Class B, 5-, 10-, 25-, and 100-mL, VWR, Cat. Nos. 14201-610, -462, -364, and -718, respectively Flasks, volumetric, polypropylene, Class B, 25-, 50-, 100-, and 1000 mL, VWR, Cat. Nos. 830
17、08-964, -974, -984, and 83009-008, respectively; several of each required Fume hood Hot plate, VWR, Cat. No. 11301-006 ICP-MS, PerkinElmer, DRC-e, equipped with an AS10 autosampler, or equivalent newer version Microwave oven, 15-vessel capacity, with closed digestion vessels, quartz tubes and lids,
18、CryoLab chiller, and computer, Milestone, UltraWAVE, or equivalent newer version Pipet, automatic, 10-100-L, VWR, Cat. No. 89079-968; calibrated and/or certified according to manufacturers recommendations 3 of 11 1005-14 Pipet, automatic, 100-1000-L, VWR, Cat. No. 89079-974; calibrated and/or certif
19、ied according to manufacturers recommendations Pipet, electronic, 1-10-mL, EDP 3-Plus, Rainin, Cat. No. SE3-10mL; calibrated and/or certified according to manufacturers recommendations Regulator, argon or nitrogen, two-stage, high purity, delivery pressure range 30-700 kPa (4-100 psi), Matheson Tri-
20、Gas, Model 3122-580 Timer, capable of timing two hours, VWR, Cat. No. 62344-440 Watch glass, PTFE, 125-mm, Fisher Scientific, Cat. No. 02-617-1F Reagents and Materials References to catalog numbers and suppliers are included as a convenience to the method user. Other suppliers may be used. Unless ot
21、herwise specified, references to water mean deionized water. Argon, 99.995% minimum purity Bismuth, indium, and scandium stock solution, aqueous, 100-g/mL. To prepare 50 mL, pipet 0.5 mL of the 10,000- g/mL scandium, 5 mL of the 1000- g/mL indium, and 5 mL of the 1000- g/mL bismuth metal solutions i
22、nto a 50-mL volumetric flask. Fill to the mark with deionized water. Cap and invert several times to mix. The solution should remain stable for two months. Bismuth, indium, and scandium internal standard solution, 10-g/mL. To prepare 100 mL, pipet 10.0 mL of the 100-g/mL scandium, indium, and bismut
23、h metal stock solution into a 100-mL volumetric flask. Fill to the volume mark with deionized water and mix. Store in a 125-mL polypropylene bottle. The solution should remain stable for two months. Bismuth metal solution, aqueous, 1000-g/mL, SPEX CertiPrep, Inc., Cat. No. PLBI-2, for use as an inte
24、rnal standard Cleaning compound, Alconox detergent, VWR, Cat. No. 21835-032 Gloves, neoprene/natural rubber, VWR, Cat. No. 32917-212 (size 10) Gloves, Viton, VWR, Cat. No. 89236-768 (size 10) Hydrofluoric acid, concentrated, trace metals grade, VWR, Cat. No. EM-HX0628-7 Hydrogen peroxide, 29-32%, VW
- 1.请仔细阅读文档,确保文档完整性,对于不预览、不比对内容而直接下载带来的问题本站不予受理。
- 2.下载的文档,不会出现我们的网址水印。
- 3、该文档所得收入(下载+内容+预览)归上传者、原创作者;如果您是本文档原作者,请点此认领!既往收益都归您。
下载文档到电脑,查找使用更方便
10000 积分 0人已下载
下载 | 加入VIP,交流精品资源 |
- 配套讲稿:
如PPT文件的首页显示word图标,表示该PPT已包含配套word讲稿。双击word图标可打开word文档。
- 特殊限制:
部分文档作品中含有的国旗、国徽等图片,仅作为作品整体效果示例展示,禁止商用。设计者仅对作品中独创性部分享有著作权。
- 关 键 词:
- UOP10052014TRACEMETALSINORGANICSBYICPMSPDF

链接地址:http://www.mydoc123.com/p-1074460.html