1、A-A-54717 b5 = 74 00875b 8bO W i dr I INCH-POUND 1 10 September A-A-54 7 17 COMMERCIAL ITEM DESCRIPTION 1992 COMPRESSOR AND OXYGEN BLENDER, VENTILATOR (Portable, Battery and Line Powered) The General Services Administration has authorized the use of this Commercial Item Description. This Commercial
2、Item Description covers a compressor with oxygen blender, AC power supply, and other required components. Salient Characteristics! The portable compressor system shall be suitable for use in a hospital, ambulance, military aeromedical aircraft, and the field where no external power is available. It
3、shall provide dry, 100 percent oil-free, compressed air and oxygen blended mixtures. volume ventilator specified elsewhere in the procurement document . The system shall be compatible with a specific A mechanical means shall be provided to attach this compressor system to the ventilator. Such mechan
4、ical means shall be compatible with the ventilator. The complete portable compressor system shall include the following: a self contained portable compressor, integral or separate air/oxygen blender, integrel or separate AC power supply unit, accumulator type compressed air storage reservoir, batter
5、y recharge circuitry, rechargeable battery, voltage selector switch, low battery alarm, power connection indicator, electrical connections for external power sources, gas filter, pneumatic connections, condensed operating instructions, dedicated compressor case with integral carrying handle(s), main
6、tenance and operating instructions, and the additional components required to form an operational Ventilation system when combined with the ventilator. condensed operating instructions shall be marked on this The #!SC N/A FSC 6530 pWERBUTION STATEMENT A: distribution is unlimited. Approved for publi
7、c release; Provided by IHSNot for ResaleNo reproduction or networking permitted without license from IHS-,-,-A-A-5Y717 65 939997lI 0087570 582 A-A-54 7 17 dedicated case or on a label attached to this case. Controls and indicators shall be functionally grouped to facilitate operator usage. The compl
8、ete compressor system shall be contained in a carry-all carrying case with handle. The compressor shall be capable of sustaining a positive pressure of approximately 7 psig and a peak airflow of at least 40 Lpm. It shall be designed with compressor unit components. that interconnect with the ventila
9、tor to allow the ventilator to operate using 7 psig air and oxygen blended input. All ventilator functions shall be operational and all ventilator controls shall be either directly operational or superseded by corresponding controls on the compressor system. The compressor shall be capable of allowi
10、ng the ventilator to deliver the following peak tidal volume: a. O to 300 ml in 0.5 seconds b. O to 700 ml in 1.0 seconds c. O to 900 ml in 1.5 seconds d. O to 1100 ml in 2.0 seconds The compressor shall have a minimum operating temperature range of minus 4OF to plus 115F from all power sources cite
11、d below, including the internal rechargeable battery. The minimum long term storage temperature range shall be from 5OoF to 80F. The compressor shall contain both audible and visual low battery alarms which activate when 10 minutes (at least) of battery power remains. The audible alarm shall be muta
12、ble. The compressor shall be provided with an indicator which denotes the presence and source (internal/external) of operating power. A mechanical means shall be provided to allow attachment of the compressor, with blender, to the ventilator to form a composite ventilator system. !- a e e Provided b
13、y IHSNot for ResaleNo reproduction or networking permitted without license from IHS-,-,-A-A-SV717 b5 9999974 0087573 4L9 H A-A-54 7 17 The compressor with the AC power supply unit shall be operational from any of the following sources: a. Internal battery. b. Nominal 12 volts DC, vehicle generating
14、system rated at 120 amperes at 14 volts DC with negati,ve ground polarity. c. External 120 amp-hour heavy duty 12 volt battery with negative ground polarity. d. Aircraft 28 volt DC geserating system with negative ground polarity. e. 115/230 volt, 50-400 Hz. The self-contained portable compressor sha
15、ll be housed in a semi-rigid plastic or corrosion-resistant metal case with a carrying handle. contained portable compressor (and integral blender and/or integral AC power supply unit if they are not adjunctive). It and all other components, including operating and maintenance manuals, shall be cont
16、ained in a carry-all carrying case with handle(s). The circuitry of the self-contained compressor shall contain electrical connections to allow both battery recharge and compressor operation from all the electrical sources listed herein. The internal, rechargeable battery shall be 12 volts (2 batter
17、ies wired together to provide 12 volts is acceptable). This battery shall have the capability to allow the compressor to deliver compressed gas to the ventilator, allowing the ventilator to be operational for at least one hour without diminished ventilator capability (peak flow 40 Lpm with inspirato
18、ry time set at one second). This internal battery shall not deplete more than approximately 1.5 percent per month, or 18 percent per year if stored at 70F. It shall be rechargeable, complete charge within 24 hours, from any of the following sources (with the AC power supply unit) : This case shall h
19、ouse only the self- a. 115/230 volt, 50-400 Hz. b. Nominal 12 volt DC, vehicle generating system r”ated at 120 amperes at 14 volts DC with negative ground polarity. 3 Provided by IHSNot for ResaleNo reproduction or networking permitted without license from IHS-,-,-A-A-54717 65 m 9999974 0087572 355
20、D A-A-5 4 7 17 d. Aircraft 28 volt DC generating system with negative ground polarity. The gas blender shall either be integral or adjunctive to the compressor. proportionately mixing medical grade oxygen with compressed air. Blender settings shall permit delivery of oxygen concentrations ranging be
21、tween 21 percent and 100 percent. The AC power supply shall be either integral or adjunctive to the portable compressor apparatus. This supply shall operate from nominal 115/230 V, 50-400 Hz, single phase power sources and shall contain a voltage selector switch. This AC power supply shall contain a
22、n appropriately strain-relieved line cord, at least 6 feet long, consisting of SJT-type wire with a minimum rating of 18/3 AWG, and terminated in a 3-conductor male hospital-grade plug. If the AC power supply is integral to the compressor, a means by which to wrap and secure the AC power cord shall
23、be provided. If the AC power supply is adjunctive, its output cable shall be at least 4 feet long and terminate in a suitable, non-shorting type plug to interface with the compressor. The power supply output voltage shall be within the range of acceptance of the compressor apparatus described herein
24、, for both battery recharge and operation. The blending mechanism shall be capable of Ground polarity shall be negative. Connecting hose(s) and pneumatic connections shall be provided with the portable compressor. If the blender is integral to the compressor, one connecting hose for attachment to th
- 1.请仔细阅读文档,确保文档完整性,对于不预览、不比对内容而直接下载带来的问题本站不予受理。
- 2.下载的文档,不会出现我们的网址水印。
- 3、该文档所得收入(下载+内容+预览)归上传者、原创作者;如果您是本文档原作者,请点此认领!既往收益都归您。
下载文档到电脑,查找使用更方便
10000 积分 0人已下载
下载 | 加入VIP,交流精品资源 |
- 配套讲稿:
如PPT文件的首页显示word图标,表示该PPT已包含配套word讲稿。双击word图标可打开word文档。
- 特殊限制:
部分文档作品中含有的国旗、国徽等图片,仅作为作品整体效果示例展示,禁止商用。设计者仅对作品中独创性部分享有著作权。
- 关 键 词:
- DODAA547171992COMPRESSORANDOXYGENBLENDERVENTILATORPORTABLEBATTERYANDLINEPOWERED 通风设备 压缩 搅拌机 PDF

链接地址:http://www.mydoc123.com/p-701765.html