DOD A-A-54015-1989 ELECTROCARDIOGRAPH AUTOMATIC SOLID STATE PORTABLE《便携式固态自动心电图仪》.pdf
《DOD A-A-54015-1989 ELECTROCARDIOGRAPH AUTOMATIC SOLID STATE PORTABLE《便携式固态自动心电图仪》.pdf》由会员分享,可在线阅读,更多相关《DOD A-A-54015-1989 ELECTROCARDIOGRAPH AUTOMATIC SOLID STATE PORTABLE《便携式固态自动心电图仪》.pdf(7页珍藏版)》请在麦多课文档分享上搜索。
1、g - 5 - 35 A-A-5 4 O 15 25 October 1989 COMMERCIAL ITEM DESCRIPTION ELECTROCARDIOGRAPH Automatic, Solid State, Portable This General Services Administration has authorized the use of this Commercial Item Description. This Commercial Item Description covers an all solid state, electrocardiograph with
2、 a minimum of three sensitivities. single channel, dual speed, direct writing, portable Salient characteristics: The device shall be all solid state, single channel, dual speed, direct writing, portable, with a minimum of three sensitivities. It shall have twelve lead recording capability for I, II,
3、 III, aVR, aVL, aVF, V1, V2, V3, V4, V5 and V6 leads. The strip chart recorder shall have two selectable speeds of 25 and 50 mm/sec, plus or minus 2 percent. The electrocardiograph shall have sensitivities of 5, 10, and 20 mm/mv, plus or minus 2 percent. The recorder shall have a thermal stylus or d
4、ot matrix printhead. If a stylus, it shall be easily replaceable,. producing sharp, clear, and uniform traces. There shall be adjustments for stylus heat, position, and pressure. Hysteresis-return to baseline and baseline drift shall be minimal. For heated stylus recorders, the recording chart paper
5、 shall be 63 mm wide, overall 150 feet long, with a 50 mm grid width ruled in 1 mm by 5 mm squares. (dot matrix) recorders, the recording chart paper shall be 59 mm wide, overall 150 feet long, with a 45 mm grid width ruled in 1 mm by 5 mm squares. Manual and automatic recording modes shall be selec
6、table for each lead. For thermal array Automatic lead marking shall function for both automatic and manual modes. AMSC/NA FSC 6515 DISTRIBUTION STATEMENT A. Approved for public release; distribution is unlimited. F . THIS DOCUMENT CONTAINS 7 PAGES. Provided by IHSNot for ResaleNo reproduction or net
7、working permitted without license from IHS-,-,-A-A-5 4 O 15 Calibration pulse shall function for both automatic and manual modes , The electrocardiograph shall be protected for standard Lown or equivalent differential pulse to 400 watt seconds terminated to patient. Isolation circuitry shall limit t
8、otal leakage current between patient leads (ECG) and ground to the value specified in ANSI/AAMI ES 1 - 1985, American National Standard, Safe Current Limitg fox Electromedical Apparatus. Electrical characteristics, The equipment item furnished by the contractor must be capable of operating on both 5
9、0 and 60 Hertz (Hz) power souxes. Voltage characteristics or this item shall peimit the item to be operated on both o the Lollowing single phase voltage sources: 110 volt, 60 Hz; and 110 volt, Li0 Hz. Items offered which normally operate on a nominal voltage range of 220 volt to 380 volt (single pha
10、se) rathei: than 110 volt will be accepted only if the item is capable of functioning on both 50 and 60 Hz, In order to meet this requirement, contractors may include a 60 Hz conversion kit with adequate instructions or the Government to perform the modification. The kit will be separately identifie
11、d and priced on the bid/pcoposal and include a statement as to the number o manhours required to perorm the modification. Complex modifications requiring more than one manhour to accompXish will not be accepted, a 60 Hz rated pioduct is offered which does not require modification to allow opration a
12、t other than its rated requency, the contractor will so certify in his bid/pioposal, as well as identify any special operating conditions/limitations/requirements and liability and considecations which might ensue because of operation on other than the items normal rated frequency, Accessories, stan
13、dard accessories: In instances in which a 50 or Each unit shall be supplied with the following 0 One patient cable shall be supplied, 0 One each adult Welsh 0 One power cord shall 0 Foui each aduJ-t: limb 0 Four each adult limb supplied suction chest electrode shall be be supplied, electrodes shall
14、be supplied. straps shall be supplied. 2 O 2- O 9 Y Provided by IHSNot for ResaleNo reproduction or networking permitted without license from IHS-,-,-A-A-54015 65 W 7777774 0032772 4 A-A-5 4 O 15 0 One each tube/bottle electrode crearn/gel shall be 0 Two each rolls ECG chart paper shall be supplied.
15、 supplied. 0 Chart paper mounts shall be supplied. A dust cover shall be supplied. The electrocardiograph shall conform to the standards of electrical leakage currents or operator safety as specified in UL544 - “Standard for Safety of Electrical Medical and Dental Equipment“, and NFPA 99-1987, Chapt
16、er 9, “Manufacturer Requirements.“ TWQ complete instruction, operation, maintenance, and repair manuals shall be furnished. The patient cable leads supplied with the electrocardiograph shall be color coded in conformance with AHA publication, “Recommendation for Standardization of Leads and of Speci
17、fication for Instruments in Electrocardiography and Vectorcardiography“. The patient cable connector shall be a 3 lead/ pin type configured in accordance with AAMI ECGC-5/83 “Standard for ECG Connectors“. The electrocardiograph performance standards shall be in accordance with ANSI/AAMI ECll-1982 “A
18、merican National Standard for Diagnostic Electrocardiograph Devices“. The electrocardiograph shall measure no more than 20 inches long by 15 inches wide by 6 inches high. The unit shall weigh no more than 10 pounds. Workmanship. The electrocardiograph shall be free from defects Unit, Each (EA). One
19、electrocardiograph, as specified, which detract from its appearance or impair its serviceability. constitutes one unit. Provided by IHSNot for ResaleNo reproduction or networking permitted without license from IHS-,-,-A-A-5403.5 PUALITY ASSURANCE PROVISIONS Responsibility Por inspection. Unless othe
20、rwise specified in the contxact or purchase order, the contractor is responsible for the performance of all inspection requirements (examinations and tests) as specified herein. Except as otherwise specified in the contract or purchase order, the contractor may use his own or any facilities suitable
- 1.请仔细阅读文档,确保文档完整性,对于不预览、不比对内容而直接下载带来的问题本站不予受理。
- 2.下载的文档,不会出现我们的网址水印。
- 3、该文档所得收入(下载+内容+预览)归上传者、原创作者;如果您是本文档原作者,请点此认领!既往收益都归您。
下载文档到电脑,查找使用更方便
10000 积分 0人已下载
下载 | 加入VIP,交流精品资源 |
- 配套讲稿:
如PPT文件的首页显示word图标,表示该PPT已包含配套word讲稿。双击word图标可打开word文档。
- 特殊限制:
部分文档作品中含有的国旗、国徽等图片,仅作为作品整体效果示例展示,禁止商用。设计者仅对作品中独创性部分享有著作权。
- 关 键 词:
- DODAA540151989ELECTROCARDIOGRAPHAUTOMATICSOLIDSTATEPORTABLE 便携式 固态 自动 心电图 PDF

链接地址:http://www.mydoc123.com/p-701701.html