1、BRITISH STANDARD BS 6859-1: 1987 Operating tables Part 1: Specification for mobile general operating tables UDC 615.478.62BS6859-1:1987 This British Standard, having been prepared under the directionof the Health Care Standards Committee, was published under the authorityofthe Board of BSI and comes
2、 into effect on 30 November1987 BSI 11-1999 The following BSI references relate to the work on this standard: Committee reference HCC/33 Draft for comment 80/62157 DC ISBN 0 580 15902 7 Committees responsible for this British Standard The preparation of this British Standard was entrusted by the Hea
3、lth Care Standards Committee (HCC/-) to Technical Committee HCC/33, upon which the following bodies were represented: Association of Anaesthetists of Great Britain and Ireland Association of Clinical Pathologists British Medical Association British Surgical Trades Association Incorporated Castor Man
4、ufacturers (UK) Association Department of Health and Social Security Department of the Environment (Property Services Agency) Malaysian Rubber Producers Research Association National Association of Theatre Nurses Royal College of Midwives Royal College of Nursing and National Council for Nurses of t
5、he UnitedKingdom Royal College of Surgeons of England Amendments issued since publication Amd. No. Date of issue CommentsBS6859-1:1987 BSI 11-1999 i Contents Page Committees responsible Inside front cover Foreword ii 1 Scope 1 2 Dimensions and height 1 3 General requirements 1 4 Table side bars 2 5
6、Dimensions of side bar accessory clamps 2 6 Radiographic requirements 2 7 Finishes of exposed surfaces of table 2 8 Mattress 2 9 Marking 3 10 Information to be supplied by the manufacturer 3 Appendix A Recommendations for design of mobile general operating tables 4 Appendix B Recommendations for rad
7、io-translucency of the table top 4 Figure 1 Typical mobile general operating table 5 Figure 2 Some table top positions 6 Table 1 Tests and requirements for paint and plastics coating finishes 3 Publications referred to Inside back coverBS6859-1:1987 ii BSI 11-1999 Foreword This Part of BS6859 has be
8、en prepared under the direction of the Health Care Standards Committee at the request of the Department of Health and Social Security. BS6859 is published in two Parts, Part2 specifying requirements for fixed-pedestal general operating tables. Attention is drawn to BS5724-2.22 which specifies requir
9、ements for safety of general operating tables. BS2891 which specifies requirements for operating table covers and pads is currently being revised. When the revision is published, clause8 of this Part of BS6859 will be amended to refer to it. Design recommendations for mobile general operating tables
10、 are given in Appendix A. A British Standard does not purport to include all the necessary provisions of a contract. Users of British Standards are responsible for their correct application. Compliance with a British Standard does not of itself confer immunity from legal obligations. Summary of page
11、s This document comprises a front cover, an inside front cover, pagesi andii, pages1 to6, an inside back cover and a back cover. This standard has been updated (see copyright date) and may have had amendments incorporated. This will be indicated in the amendment table on the inside front cover.BS685
12、9-1:1987 BSI 11-1999 1 1 Scope This Part of BS6859 specifies requirements for mobile general operating tables with non-transferable tops of four or more sections. Mobile operating tables intended for use with specialist procedures and accessories to tables are outside the scope of this Part of BS685
13、9, except that dimensions for accessory clamps have been specified to ensure interchangeability of accessories between tables. NOTEThe titles of the publications referred to in this standard are listed on the inside back cover. 2 Dimensions and height NOTEFigure 1 shows schematic representations of
14、a typical mobile general operating table. 2.1 Table top The length of the table top shall be not less than1850mm, and the width, including the side bars, shall be not more than620mm. 2.2 Trunk section(s) The length of the trunk section(s), i.e.excluding head and leg sections, shall be not less than8
15、40mm and the width of the trunk sections excluding side bars shall be not less than470mm. 2.3 Height With the table top horizontal, the height of the table, excluding the mattress, shall be not less than650mm. The height of the table shall be adjustable upwards by not less than200mm. 2.4 Floor clear
16、ance The minimum clearance between the floor and table base shall be12mm. 3 General requirements NOTE 1Except for movements of the leg section, all movements of the table should be controllable from or near the usual head end of the table. NOTE 2When assessing compliance with 3.1 and 3.2, the table
17、should be lowered by an operator experienced in the use of operating tables. 3.1 Trendelenburg position see Figure 2(a) It shall be possible to lower the table top to a position where the angle between the head end and the horizontal is at least35 but it shall not be possible to lower the table top
18、so that the angle exceeds45 . The time taken to lower the table top to an angle of35 from the horizontal shall not exceed8s from engagement of the lowering mechanism. 3.2 Reverse Trendelenburg position seeFigure 2(b) It shall be possible to lower the table top to a position where the angle between t
19、he foot end and the horizontal is at least35 but it shall not be possible to lower the table top so that the angle exceeds45 . The time taken to lower the table top to an angle of35 from the horizontal shall not exceed8s from engagement of the lowering mechanism. 3.3 Lateral tilt see Figure 2(c) It
20、shall be possible to tilt the table top laterally to either side to a position where the angle between the table top and the horizontal is at least15 but does not exceed20 . 3.4 Break-back point see Figure 2(d) and Figure 2(e) At the break-back point, the table top shall be capable of flexion to130
21、in any position where all sections are above the horizontal. The table top shall be capable of extension to an angle of not less than220 between the upper surfaces when all sections are below the horizontal. 3.5 Chair or sitting position It shall be possible to adjust the sections of the table top t
22、o form a chair or sitting position. 3.6 Rotation of table top see Figure 1(c) If the table top can be rotated in the horizontal plane with respect to the base, a mechanism shall be provided to lock it in any position. 3.7 Mobility The table shall move in a forward or backward direction without later
23、al drift, and it shall be possible to steer the table without lateral drift. It shall be possible to immobilize the table by means of a single control. 3.8 Head section The head section shall be capable of adjustment through any angle between45 1 downwards and10 upwards. A quick-release mechanism sh
24、all be fitted incorporating a self-locking device and capable of overriding the adjustment control. 3.9 Single leg section The leg section on tables with a single leg section shall be adjustable downwards through a minimum angle of100 .BS6859-1:1987 2 BSI 11-1999 3.10 Divided leg section The two leg
- 1.请仔细阅读文档,确保文档完整性,对于不预览、不比对内容而直接下载带来的问题本站不予受理。
- 2.下载的文档,不会出现我们的网址水印。
- 3、该文档所得收入(下载+内容+预览)归上传者、原创作者;如果您是本文档原作者,请点此认领!既往收益都归您。
下载文档到电脑,查找使用更方便
10000 积分 0人已下载
下载 | 加入VIP,交流精品资源 |
- 配套讲稿:
如PPT文件的首页显示word图标,表示该PPT已包含配套word讲稿。双击word图标可打开word文档。
- 特殊限制:
部分文档作品中含有的国旗、国徽等图片,仅作为作品整体效果示例展示,禁止商用。设计者仅对作品中独创性部分享有著作权。
- 关 键 词:
- BS685911987OPERATINGTABLESSPECIFICATIONFORMOBILEGENERALOPERATINGTABLES 手术台 部分 通用 活动 操作 工作台 规范 PDF

链接地址:http://www.mydoc123.com/p-546324.html