ASTM F2503-2013 Standard Practice for Marking Medical Devices and Other Items for Safety in the Magnetic Resonance Environment《用于保证磁共振环境安全的医疗器械及其他项标记的标准实施规程》.pdf
《ASTM F2503-2013 Standard Practice for Marking Medical Devices and Other Items for Safety in the Magnetic Resonance Environment《用于保证磁共振环境安全的医疗器械及其他项标记的标准实施规程》.pdf》由会员分享,可在线阅读,更多相关《ASTM F2503-2013 Standard Practice for Marking Medical Devices and Other Items for Safety in the Magnetic Resonance Environment《用于保证磁共振环境安全的医疗器械及其他项标记的标准实施规程》.pdf(9页珍藏版)》请在麦多课文档分享上搜索。
1、Designation: F2503 13Standard Practice forMarking Medical Devices and Other Items for Safety in theMagnetic Resonance Environment1This standard is issued under the fixed designation F2503; the number immediately following the designation indicates the year oforiginal adoption or, in the case of revi
2、sion, the year of last revision. A number in parentheses indicates the year of last reapproval. Asuperscript epsilon () indicates an editorial change since the last revision or reapproval.1. Scope1.1 This international standard applies to the practice ofmarking of items that might be used in the mag
3、netic resonance(MR) environment.1.2 The purpose of this practice is to mark items that mightbe brought into the MR environment and to recommendinformation that should be included in the marking.1.3 The standard specifies the permanent marking of items,which are used in an MR environment, by means of
4、 terms andicons.1.4 MR image artifacts are not considered to be a perfor-mance issue and so are not addressed in this internationalstandard practice (see X1.5).1.5 The values stated in SI units are to be regarded asstandard. No other units of measurement are included in thisstandard.1.6 This standar
5、d does not purport to address all of thesafety concerns, if any, associated with its use. It is theresponsibility of the user of this standard to establish appro-priate safety and health practices and determine the applica-bility of regulatory limitations prior to use.2. Referenced Documents2.1 The
6、following referenced documents are indispensablefor the application of this document. For dated references, onlythe edition cited applies. For undated references, the latestedition of the referenced document (including any amend-ments) applies.2.2 ASTM Standards:2F2052 Test Method for Measurement of
7、 Magnetically In-duced Displacement Force on Medical Devices in theMagnetic Resonance EnvironmentF2119 Test Method for Evaluation of MR Image Artifactsfrom Passive ImplantsF2182 Test Method for Measurement of Radio FrequencyInduced Heating On or Near Passive Implants DuringMagnetic Resonance Imaging
8、F2213 Test Method for Measurement of Magnetically In-duced Torque on Medical Devices in the Magnetic Reso-nance Environment2.3 Other Standards:IEC 60601-2-33 Medical Electrical EquipmentPart 2-33:Particular Requirements for the Safety of Magnetic Reso-nance Equipment for Medical Diagnosis3ISO 14971
9、Medical DevicesApplication of Risk Manage-ment to Medical Devices4ISO/IEC Guide 51 Safety AspectsGuidelines for theirInclusion in Standards4ISO TS 10974 Assessment of the Safety of Magnetic Reso-nance Imaging for Patients with an Active ImplantableMedical Device43. Terminology3.1 Definitions:3.1.1 h
10、armful interactionunintended direct or indirect in-teraction of items with MR equipment, especially with thestatic magnetic field, the gradient fields and the RF fields of theMR equipment, that can pose hazards to patients or otherpersons.NOTE 1In this context, the affected image quality or image ar
11、tifactsare not considered to be a harmful interaction.3.1.2 hazardpotential source of harm. ISO/IEC Guide 513.1.3 itemobject that might be brought into the MRenvironment.3.1.4 magnetically induced displacement forceforce pro-duced when a magnetic object is exposed to the spatial gradient1This practi
12、ce is under the jurisdiction ofASTM Committee F04 on Medical andSurgical Materials and Devices and is the direct responsibility of SubcommitteeF04.15 on Material Test Methods.Current edition approved June 1, 2013. Published July 2013. Originally approvedin 2005. Last previous edition approved in 200
13、8 as F250308. DOI: 10.1520/F2503-13.2For referenced ASTM standards, visit the ASTM website, www.astm.org, orcontact ASTM Customer Service at serviceastm.org. For Annual Book of ASTMStandards volume information, refer to the standards Document Summary page onthe ASTM website.3Available from Internati
14、onal Electrotechnical Commission (IEC), 3, rue deVaremb, P.O. Box 131, CH-1211 Geneva 20, Switzerland, http:/www.iec.ch.4Available from American National Standards Institute (ANSI), 25 W. 43rd St.,4th Floor, New York, NY 10036, http:/www.ansi.org.Copyright ASTM International, 100 Barr Harbor Drive,
15、PO Box C700, West Conshohocken, PA 19428-2959. United States1of a static magnetic field. This force will tend to cause theobject to translate in the spatial gradient of the static magneticfield.3.1.5 magnetically induced torquetorque produced whena magnetic object is exposed to a magnetic field. Thi
16、s torquewill tend to cause the object to align itself along the magneticfield in an equilibrium direction that induces no torque.3.1.6 magnetic induction or magnetic flux density (B inT)that magnetic vector quantity which at any point in amagnetic field is measured either by the mechanical forceexpe
17、rienced by an element of electric current at the point, or bythe electromotive force induced in an elementary loop duringany change in flux linkages with the loop at the point. Themagnetic induction is frequently referred to as the magneticfield. B0is the static field in an MR equipment and accessor
18、ies.Plain type indicates a scalar (for example, B) and bold typeindicates a vector (for example, B).3.1.7 magnetic resonance (MR)resonant absorption ofelectromagnetic energy by an ensemble of atomic nucleisituated in a magnetic field. IEC 60601-2-33, definition201.3.2173.1.8 magnetic resonance (MR)
19、equipmentmedical elec-trical equipment which is intended for in vivo magneticresonance examination of a patient comprising all parts inhardware and software from the supply mains to the displaymonitor.NOTE 2The MR equipment is a programmable electrical medicalsystem (PEMS). IEC 60601-2-33, definitio
20、n 201.3.2183.1.9 magnetic resonance (MR) examinationprocess ofacquiring data by magnetic resonance from a patient. IEC60601-2-33, definition 201.3.2193.1.10 magnetic resonance (MR) environmentthe threedimensional volume of space surrounding the MR magnet thatcontains both the Faraday shielded volume
21、 and the 0.50 mTfield contour (5 gauss (G) line). This volume is the region inwhich an item might pose a hazard from exposure to theelectromagnetic fields produced by the MR equipment andaccessories.3.1.11 MR Conditionalan item with demonstrated safetyin the MR environment within defined conditions.
22、 At aminimum, address the conditions of the static magnetic field,the switched gradient magnetic field and the radiofrequencyfields. Additional conditions, including specific configurationsof the item, may be required.3.1.12 Supplementary Markingadditional informationthat, in association with a mark
23、ing as “MR Conditional,” statesvia additional language the conditions in which an item can beused safely within the MR environment.3.1.13 MR Safean item that poses no known hazardsresulting from exposure to any MR environment. MR Safeitems are composed of materials that are electricallynonconductive
24、, nonmetallic, and nonmagnetic.NOTE 3An item composed entirely of electrically nonconductive,nonmetallic and nonmagnetic materials may be determined to be MR Safeby providing a scientifically based rationale rather than test data. Ex-amples of MR Safe items are a cotton blanket or a silicone cathete
- 1.请仔细阅读文档,确保文档完整性,对于不预览、不比对内容而直接下载带来的问题本站不予受理。
- 2.下载的文档,不会出现我们的网址水印。
- 3、该文档所得收入(下载+内容+预览)归上传者、原创作者;如果您是本文档原作者,请点此认领!既往收益都归您。
下载文档到电脑,查找使用更方便
5000 积分 0人已下载
下载 | 加入VIP,交流精品资源 |
- 配套讲稿:
如PPT文件的首页显示word图标,表示该PPT已包含配套word讲稿。双击word图标可打开word文档。
- 特殊限制:
部分文档作品中含有的国旗、国徽等图片,仅作为作品整体效果示例展示,禁止商用。设计者仅对作品中独创性部分享有著作权。
- 关 键 词:
- ASTMF25032013STANDARDPRACTICEFORMARKINGMEDICALDEVICESANDOTHERITEMSFORSAFETYINTHEMAGNETICRESONANCEENVIRONMENT

链接地址:http://www.mydoc123.com/p-538862.html