《ASTM F2228-2002(2007) Standard Test Method for Non-Destructive Detection of Leaks in Medical Packaging Which Incorporates Porous Barrier Material by CO2 Tracer Gas Method《CO2示踪气体法对.pdf》由会员分享,可在线阅读,更多相关《ASTM F2228-2002(2007) Standard Test Method for Non-Destructive Detection of Leaks in Medical Packaging Which Incorporates Porous Barrier Material by CO2 Tracer Gas Method《CO2示踪气体法对.pdf(5页珍藏版)》请在麦多课文档分享上搜索。
1、Designation: F 2228 02 (Reapproved 2007)Standard Test Method forNon-Destructive Detection of Leaks in Medical PackagingWhich Incorporates Porous Barrier Material by CO2TracerGas Method1This standard is issued under the fixed designation F 2228; the number immediately following the designation indica
2、tes the year oforiginal adoption or, in the case of revision, the year of last revision. A number in parentheses indicates the year of last reapproval. Asuperscript epsilon (e) indicates an editorial change since the last revision or reapproval.1. Scope1.1 This non-destructive test method detects le
3、aks in non-porous rigid thermoformed trays, as well as the seal betweenthe porous lid and the tray. The test method detects channelleaks in packages as small as 100 m (0.004 in.) diameter in theseal as well as 50 m (0.002 in.) diameter pinholes, orequivalently sized cracks in the tray, subject to tr
4、ace gasconcentration in the package, package design and manufactur-ing tolerances.NOTE 1This test method does not claim to challenge the porous(breathable) lidding material. Any defects that may exist in the porousportion of the package will not be detected by this test method.1.2 The values stated
5、in SI units are to be regarded asstandard units. Values in parentheses are for information only.1.3 This standard does not purport to address all of thesafety concerns, if any, associated with its use. It is theresponsibility of the user of this standard to establish appro-priate safety and health p
6、ractices and determine the applica-bility of regulatory limitations prior to use.2. Referenced Documents2.1 ASTM Standards:2D 996 Terminology of Packaging and Distribution Environ-mentsF17 Terminology Relating to Flexible Barrier PackagingF 1327 Terminology Relating to Barrier Materials for Medi-cal
7、 Packaging3. Terminology3.1 General Term DefinitionsFor definitions used in thisstandard, see Terminologies D 996, F 17, and F 1327.3.2 Definitions of Terms Specific to This Standard:3.2.1 basal flowtransverse transport of trace gas acrossthe seal due to gas flow within the plane of the porous barri
8、ermaterial as well as flow between the porous barrier and thetemporary gasketing. This is an expected property of theporous barrier material and does not represent a leak. Experi-mentally, this flow may be thought of as noise, which willalways be present, to some degree, during testing and must beac
9、counted for.3.2.2 trace gasa compound selected solely for use toidentify leakage flow.4. Summary of Test Method4.1 This test method utilizes CO2sensing techniques in thedetection of a CO2trace gas to quantify leaks in medicalpackaging, which incorporates porous barrier material. Thistest method prov
10、ides a qualitative (accept/reject) inspectionmethod to evaluate packages for pinhole, crack and channelleaks. Further information on the “Leak Test Theory” may befound in Annex A1.5. Significance and Use5.1 Harmful biological or particulate contaminants mayenter the package through incomplete seals
11、or imperfectionssuch as pinholes or cracks in the trays.5.2 After initial instrument set-up and calibration, the op-erations of individual tests and test results do not need operatorinterpretation. The non-destructive nature of the test may beimportant when testing high value added products.5.3 Leak
12、 test results that exceed the permissible thresholdsetting are indicated by audible or visual signal responses, orboth, or by other means.5.4 This non-destructive test method may be performed ineither laboratory or production environments. This testing maybe undertaken on either a 100 % or a statist
13、ical sampling basis.This test method, in single instrument use and current imple-mentation, may not be fast enough to work on a productionpackaging line, but is well suited for statistical testing as wellas package developmental design work.1This test method is under the jurisdiction of ASTM Committ
14、ee F02 on FlexibleBarrier Packaging and is the direct responsibility of Subcommittee F02.40 onPackage Integrity.Current edition approved April 1, 2007. Published May 2007. Originallyapproved in 2002. Last previous edition approved in 2002 as F 2228 02.2For referenced ASTM standards, visit the ASTM w
15、ebsite, www.astm.org, orcontact ASTM Customer Service at serviceastm.org. For Annual Book of ASTMStandards volume information, refer to the standards Document Summary page onthe ASTM website.1Copyright ASTM International, 100 Barr Harbor Drive, PO Box C700, West Conshohocken, PA 19428-2959, United S
16、tates.6. Apparatus6.1 Non-destructive Trace Gas Leak Detection ApparatusThe apparatus test fixture consists of three major elements andis shown in Fig. 1.6.2 Sealing MembraneThe purpose of the sealing mem-brane is to seal off the tracer gas transmission, normal to theporous lid surface. However, the
17、 membrane does not com-pletely control the transmission of tracer gas basal flow in thetransverse direction.6.3 Control PackagesPackages with calibrated capillarychannel leaks as well as packages with calibrated pinholes inthe tray constructed for instrument calibration as well as fortest procedure
18、verification.6.4 Test FixtureApparatus, which must be designed toensure detection of a calibrated leak.7. Preparation of Apparatus7.1 The test apparatus is to be started, warmed-up, and madeready according to the manufacturers specifications. Theinstrument must be operated in an environment as descr
19、ibed inthe instruments user manual.8. Reagents and Materials8.1 CO2Trace Gas Cylinder and RegulatorA cylinder of“Commercial” or “Bone Dry” grade carbon dioxide with aminimum of 206.84 kPa (30 psi) pressure is required forcalibration and testing.8.2 Sealing MembraneThe temporary sealing membranemust
20、exhibit the correct pliability and tackiness in order toform a gas-tight bond with the porous lidding materials duringthe testing process, and must release at the end of the testwithout damaging the porous lid or the edge seal.8.3 Sealing Membrane-induced DamageDuring the pro-cess of membrane select
21、ion for a specific package design andconfiguration, inspect the packages for the following indica-tions of membrane-induced damage after the membrane isremoved from the package:8.3.1 Sticky residue remaining on the porous barrier mate-rial at the end of the test cycle.8.3.2 Fibers from the porous ba
22、rrier material remaining onthe sealing membrane at the end of the test cycle.8.3.3 Visible changes to the texture or structure of theporous lidding material at the end of the test cycle, undermicroscope or other magnified examination.8.3.4 Damage to the printed information on the porousbarrier. The
23、adhesive of the sealing membrane may lift off theink from the barrier.8.3.5 Failure of the package to release from the sealingmembrane at the end of the test cycle.8.3.6 Damage to the seal incurred on removal of themembrane from the package.9. Hazards9.1 As the test fixture is closed, it may present
24、 pinch-pointhazards.9.2 CO2, although inert and non-toxic, can cause danger ofsuffocation if it is allowed to displace oxygen. Thus it isrecommended that the spent carbon dioxide be naturally ventedaway from the test area and that adequate ventilation beprovided.10. Calibration and Standardization10
- 1.请仔细阅读文档,确保文档完整性,对于不预览、不比对内容而直接下载带来的问题本站不予受理。
- 2.下载的文档,不会出现我们的网址水印。
- 3、该文档所得收入(下载+内容+预览)归上传者、原创作者;如果您是本文档原作者,请点此认领!既往收益都归您。
下载文档到电脑,查找使用更方便
5000 积分 0人已下载
下载 | 加入VIP,交流精品资源 |
- 配套讲稿:
如PPT文件的首页显示word图标,表示该PPT已包含配套word讲稿。双击word图标可打开word文档。
- 特殊限制:
部分文档作品中含有的国旗、国徽等图片,仅作为作品整体效果示例展示,禁止商用。设计者仅对作品中独创性部分享有著作权。
- 关 键 词:
- ASTMF222820022007STANDARDTESTMETHODFORNONDESTRUCTIVEDETECTIONOFLEAKSINMEDICALPACKAGINGWHICHINCORPORATESPOROUSBARRIERMATERIALBYCO2TRACERGASMETHODCO2

链接地址:http://www.mydoc123.com/p-538116.html