ASTM F2227-2002(2007) Standard Test Method for Non-Destructive Detection of Leaks in Non-sealed and Empty Medical Packaging Trays by CO2 Tracer Gas Method《CO2示踪气体法对不密封的及空的医用包装盘泄漏性进.pdf
《ASTM F2227-2002(2007) Standard Test Method for Non-Destructive Detection of Leaks in Non-sealed and Empty Medical Packaging Trays by CO2 Tracer Gas Method《CO2示踪气体法对不密封的及空的医用包装盘泄漏性进.pdf》由会员分享,可在线阅读,更多相关《ASTM F2227-2002(2007) Standard Test Method for Non-Destructive Detection of Leaks in Non-sealed and Empty Medical Packaging Trays by CO2 Tracer Gas Method《CO2示踪气体法对不密封的及空的医用包装盘泄漏性进.pdf(4页珍藏版)》请在麦多课文档分享上搜索。
1、Designation: F 2227 02 (Reapproved 2007)Standard Test Method forNon-Destructive Detection of Leaks in Non-sealed andEmpty Medical Packaging Trays by CO2Tracer Gas Method1This standard is issued under the fixed designation F 2227; the number immediately following the designation indicates the year of
2、original adoption or, in the case of revision, the year of last revision. A number in parentheses indicates the year of last reapproval. Asuperscript epsilon (e) indicates an editorial change since the last revision or reapproval.1. Scope1.1 This non-destructive test method detects pinhole leaksin t
3、rays, as small as 50 m (0.002 in.) in diameter, orequivalently sized cracks, subject to trace gas concentration inthe tray, tray design and manufacturing tolerances.1.2 The values stated in SI units are to be regarded asstandard units. Values in parentheses are for information only.1.3 This standard
4、 does not purport to address all of thesafety concerns, if any, associated with its use. It is theresponsibility of the user of this standard to establish appro-priate safety and health practices and determine the applica-bility of regulatory limitations prior to use.2. Referenced Documents2.1 ASTM
5、Standards:2D 996 Terminology of Packaging and Distribution Environ-mentsF 1327 Terminology Relating to Barrier Materials for Medi-cal Packaging3. Terminology3.1 General Term DefinitionsFor definitions used in thistest method, see Terminologies D 996 and F 1327, Sections 3.3.2 Definitions of Terms Sp
6、ecific to This Standard:3.2.1 trace gasa compound selected solely for use toidentify leakage flow.4. Summary of Test Method4.1 This test method utilizes CO2sensing techniques in thedetection of a CO2trace gas to quantify leaks in medicalpackaging trays. The test method provides a qualitative(accept/
7、reject) inspection method to evaluate trays for pinholesand cracks. Further information on the “Leak Test Theory”may be found in Annex A1.5. Significance and Use5.1 Harmful biological or particulate contaminants mayenter the package through imperfections such as pinholes orcracks in trays.5.2 After
8、initial instrument set-up and calibration, the op-erations of individual tests and test results do not need operatorinterpretation.5.3 Leak test results that exceed the permissible thresholdsetting are indicated by audible or visual signal responses, orboth, or by other means.5.4 This non-destructiv
9、e test method may be performed ineither laboratory or production environments and may beundertaken on either a 100 % or a statistical sampling basis.This test method, in single instrument use and current imple-mentation, may not be fast enough to work on a productionpackaging line, but is well suite
10、d for statistical testing as wellas package developmental design work.6. Apparatus6.1 Non-destructive Trace Gas Leak Detection ApparatusThe apparatus test fixture consists of three major elements andis shown in Fig. 1.6.2 Sealing MembraneThe purpose of the membrane is toseal off the tracer gas trans
11、mission out of the top of the opentray.6.3 Control TraysCalibrated pinholes, or leaks, con-structed in control trays for instrument calibration as well as fortest procedure verification.6.4 Test FixtureApparatus, which must be designed toensure detection of a calibrated leak.7. Preparation of Appara
12、tus7.1 The test apparatus is to be started, warmed-up, and madeready according to the manufacturers specifications. Theinstrument must be operated in an environment as described inthe instruments user manual.1This test method is under the jurisdiction of ASTM Committee F02 on FlexibleBarrier Packagi
13、ng and is the direct responsibility of Subcommittee F02.40 onPackage Integrity.Current edition approved April 1, 2007. Published May 2007. Originallyapproved in 2002. Last previous edition approved in 2002 as F 2227 02.2For referenced ASTM standards, visit the ASTM website, www.astm.org, orcontact A
14、STM Customer Service at serviceastm.org. For Annual Book of ASTMStandards volume information, refer to the standards Document Summary page onthe ASTM website.1Copyright ASTM International, 100 Barr Harbor Drive, PO Box C700, West Conshohocken, PA 19428-2959, United States.8. Reagents and Materials8.
15、1 CO2Trace Gas Cylinder and RegulatorA cylinder of“Commercial” or “Bone Dry” grade carbon dioxide with aminimum of 206.84-kPa (30-psi) pressure is required forcalibration and testing.8.2 Sealing MembraneThe sealing membrane must ex-hibit the correct pliability and tackiness in order to form agas-tig
16、ht bond without leaving a residue on the tray-sealingsurface after removal from the test fixture.9. Hazards9.1 As the test fixture is closed, it may present pinch-pointhazards.9.2 CO2, although inert and non-toxic, can cause danger ofsuffocation if it is allowed to displace oxygen. Thus it isrecomme
17、nded that the spent carbon dioxide be naturally ventedaway from the test area and that adequate ventilation beprovided.10. Calibration and Standardization10.1 Before any measurements are made, the apparatusmust be calibrated. The calibration procedure is used foroverall system checkout, as well as t
18、o establish an initialreference profile for simulated pinhole leaks, and to determinetest limits for each different tray geometry to be tested using aspecific test fixture. The calibration procedure is performed toestablish the sensitivity setting of the instrumentation. It isexpected that the calib
19、ration procedures be carried out fre-quently; typically, at least one or more times a day, preferablyat the beginning of every shift.10.2 Refer to the instrument manufacturers operating in-structions regarding preparation of Calibration Standards, Con-ditioning of Calibration Trays and Instrument Ca
20、libration usedin establishing baseline settings.11. Procedure11.1 Verify that sufficient CO2trace gas is available for thetests. Monitor the trace gas supply and functionality of the gasdelivery system.11.2 Select and implement the properly sized test fixture forthe trays to be tested. Verify that t
21、he instrument and associatedtest fixture have been calibrated for the trays to be tested. Thetest fixture is too large when the instrument is unable to detecta calibrated control pinhole leak.11.3 Adjust the instrument baseline settings determined incalibration.11.4 Place the tray to be tested into
22、the test fixture makingcertain that the tray is centered in the fixture and that goodsealing contact is made between the tray flange and the fixtureincorporated sealing.NOTE 1The sealing membrane needs to be clean in order to developa good seal with the sealing flange of the tray. Laboratory conditi
23、ons maycause dust or debris to be collected on the sealing membrane. Theseconditions thus will warrant frequent inspection and cleaning of thesealing membrane with a lint-free cloth soaked with a solvent recom-mended by the manufacturer of the equipment.11.5 Close the top cover of the test fixture.1
24、1.6 Start the test.11.7 Note the pass or fail indicator and record results. Setaside any “failed/defective” trays for further evaluation. Fur-ther evaluation should include re-testing of the tray.11.8 Select another tray and repeat the testing process.12. Report12.1 The report shall include the foll
- 1.请仔细阅读文档,确保文档完整性,对于不预览、不比对内容而直接下载带来的问题本站不予受理。
- 2.下载的文档,不会出现我们的网址水印。
- 3、该文档所得收入(下载+内容+预览)归上传者、原创作者;如果您是本文档原作者,请点此认领!既往收益都归您。
下载文档到电脑,查找使用更方便
5000 积分 0人已下载
下载 | 加入VIP,交流精品资源 |
- 配套讲稿:
如PPT文件的首页显示word图标,表示该PPT已包含配套word讲稿。双击word图标可打开word文档。
- 特殊限制:
部分文档作品中含有的国旗、国徽等图片,仅作为作品整体效果示例展示,禁止商用。设计者仅对作品中独创性部分享有著作权。
- 关 键 词:
- ASTMF222720022007STANDARDTESTMETHODFORNONDESTRUCTIVEDETECTIONOFLEAKSINNONSEALEDANDEMPTYMEDICALPACKAGINGTRAYSBYCO2TRACERGASMETHODCO2

链接地址:http://www.mydoc123.com/p-538114.html