ASTM F2102-2017 Standard Guide for Evaluating the Extent of Oxidation in Polyethylene Fabricated Forms Intended for Surgical Implants《评估预期供外科植入物使用的聚乙烯制品氧化程度的标准指南》.pdf
《ASTM F2102-2017 Standard Guide for Evaluating the Extent of Oxidation in Polyethylene Fabricated Forms Intended for Surgical Implants《评估预期供外科植入物使用的聚乙烯制品氧化程度的标准指南》.pdf》由会员分享,可在线阅读,更多相关《ASTM F2102-2017 Standard Guide for Evaluating the Extent of Oxidation in Polyethylene Fabricated Forms Intended for Surgical Implants《评估预期供外科植入物使用的聚乙烯制品氧化程度的标准指南》.pdf(7页珍藏版)》请在麦多课文档分享上搜索。
1、Designation: F2102 13F2102 17Standard Guide forEvaluating the Extent of Oxidation in PolyethyleneFabricated Forms Intended for Surgical Implants1This standard is issued under the fixed designation F2102; the number immediately following the designation indicates the year oforiginal adoption or, in t
2、he case of revision, the year of last revision. A number in parentheses indicates the year of last reapproval. Asuperscript epsilon () indicates an editorial change since the last revision or reapproval.1. Scope1.1 This guide covers a method for the measurement of the relative extent of oxidation pr
3、esent in HDPE homopolymers andultra-high-molecular-weight polyethylene (UHMWPE) intended for use in medical implants. The material is analyzed by infraredspectroscopy. The intensity (area) of the carbonyl absorptions (C=O) centered near 1720 cm-1 is related to the amount ofchemically bound oxygen pr
4、esent in the material. Other forms of chemically bound oxygen (C-O-C, C-O-O-C, C-O-H, and soforth) are not captured by this guide.1.2 Although this guide may give the investigator a means to compare the relative extent of carbonyl oxidation present invarious UHMWPE samples, it is recognized that oth
5、er forms of chemically bound oxygen may be important contributors to thesematerials characteristics.1.3 The applicability of the infrared method has been demonstrated by many literature reports. This particular method, usingthe intensity (area) of the C-H absorption centered near 1370 cm-1 to normal
6、ize for the samples thickness, has been validated byan Interlaboratory Study (ILS) conducted according to Practice E691.1.4 The following precautionary caveat pertains only to the test method portion, Section 5, of this specification: This standardmay involve hazardous materials, operations, and equ
7、ipment. This standard does not purport to address all of the safety concerns,if any, associated with its use. It is the responsibility of the user of this standard to establish appropriate safety and health practicesand determine the applicability of regulatory requirements prior to use.1.5 This int
8、ernational standard was developed in accordance with internationally recognized principles on standardizationestablished in the Decision on Principles for the Development of International Standards, Guides and Recommendations issuedby the World Trade Organization Technical Barriers to Trade (TBT) Co
9、mmittee.2. Referenced Documents2.1 ASTM Standards:2E691 Practice for Conducting an Interlaboratory Study to Determine the Precision of a Test MethodE2857 Guide for Validating Analytical Methods3. Terminology3.1 Definitions:3.1.1 bulk oxidation index (BOI)a samples bulk oxidation index (BOI) is the a
10、verage of the oxidation indices collected overa 500-m section at the center of the sample.3.1.1.1 DiscussionTypically, this is a plateau region with the smallest oxidation indices.3.1.1.2 Discussion1 This guide is under the jurisdiction of ASTM Committee F04 on Medical and Surgical Materials and Dev
11、ices and is the direct responsibility of Subcommittee F04.15on Material Test Methods.Current edition approved Nov. 1, 2013Sept. 1, 2017. Published December 2013September 2017. Originally approved in 2001. Last previous edition approved in 20062013as F2102 06F2102 13.1. DOI: 10.1520/F2102-13.10.1520/
12、F2102-17.2 For referencedASTM standards, visit theASTM website, www.astm.org, or contactASTM Customer Service at serviceastm.org. For Annual Book of ASTM Standardsvolume information, refer to the standards Document Summary page on the ASTM website.This document is not an ASTM standard and is intende
13、d only to provide the user of an ASTM standard an indication of what changes have been made to the previous version. Becauseit may not be technically possible to adequately depict all changes accurately, ASTM recommends that users consult prior editions as appropriate. In all cases only the current
14、versionof the standard as published by ASTM is to be considered the official document.Copyright ASTM International, 100 Barr Harbor Drive, PO Box C700, West Conshohocken, PA 19428-2959. United States1For samples less than about 8 to 10 mm thick, this central region may display the samples highest ox
15、idation indices, dependingon its state of oxidation.3.1.2 depth locator (DL)a measurement of the distance from the articular surface, or surface of interest, that a spectrum wascollected and a corresponding OI calculated.3.1.3 oxidation index (OI)an oxidation index (OI) is defined as the ratio of th
16、e area of the carbonyl absorption peak(s) centerednear 1720 cm-1 to the area of the absorption peak(s) centered near 1370 cm-1, as shown in Fig. 1. Note that the peak areas arecomputed after subtracting out the appropriate baseline, as further discussed in Section 6.3.1.4 oxidation index profilean o
17、xidation index profile is the graphical representation of variation of the samples oxidationindex with distance from its articular surface or the surface of interest. This is a plot of an OI versus DL. Typically, the graph willshow the profile through the entire thickness of the sample.3.1.5 surface
18、 oxidation index (SOI)a samples surface oxidation index (SOI) is the average of the oxidation indices from thesamples articular surface, or the surface of interest, to a depth of 3-mm subsurface.4. Apparatus4.1 Infrared Spectrometer:4.1.1 A calibrated infrared spectrometer capable of recording a tra
19、nsmission absorption spectrum over the range of about 1200to about 2000 cm-1 using about 200-sm-thick films at a resolution of 4 cm-1 and an aperture of about 200 by 200 m.FIG. 1 Typical FTIR Spectra of Oxidized UHMWPE, Showing the Definition of an Area-Based Oxidation Index Based on NormalizationUs
20、ing the 1370-cm-1 PeakFIG. 2 FTIR Spectra Showing the Carbonyl Absorption BandsNOTE 1Note that both reagents effectively extracted the lipids (the lipid absorption peak is centered at approximately 1740 cm-1). The tibial insertwas fabricated from highly crosslinked and remelted UHMWPE followed by te
21、rminal sterilization in EtO gas (Ref. 1).F2102 1724.1.1.1 Other modes of collection (that is, percent reflection, attenuated total reflection (ATR), and so forth) and aperture andincrement sizes may be used to generate the samples absorption spectrum provided they can be demonstrated to produceequiv
22、alent results. Too large an aperture can result in a loss of profile accuracy.4.1.1.2 When a Fourier Transform Infrared (FTIR) spectrometer is used, a minimum of 32 scans shall be collected perspectrum.spectrum as a default. A fewer number of scans may be performed if a user can verify that their FT
23、IR spectrometer canreproducibly measure the OI with less than 0.03 difference in OI compared with the result obtained using the default number ofscans. In no case shall an OI be reported based on fewer than 8 scans per spectrum. Guide E2857 provides guidance on how tovalidate analytical methods, whi
24、ch the user may use to determine the appropriate number of scans for their laboratory. Dependingon the level of oxidation in the tested samples, the user may elect to establish a lower threshold value of OI (nominally 0.01),particularly for UHMWPE samples that exhibit little to no detectable oxidati
- 1.请仔细阅读文档,确保文档完整性,对于不预览、不比对内容而直接下载带来的问题本站不予受理。
- 2.下载的文档,不会出现我们的网址水印。
- 3、该文档所得收入(下载+内容+预览)归上传者、原创作者;如果您是本文档原作者,请点此认领!既往收益都归您。
下载文档到电脑,查找使用更方便
5000 积分 0人已下载
下载 | 加入VIP,交流精品资源 |
- 配套讲稿:
如PPT文件的首页显示word图标,表示该PPT已包含配套word讲稿。双击word图标可打开word文档。
- 特殊限制:
部分文档作品中含有的国旗、国徽等图片,仅作为作品整体效果示例展示,禁止商用。设计者仅对作品中独创性部分享有著作权。
- 关 键 词:
- ASTMF21022017STANDARDGUIDEFOREVALUATINGTHEEXTENTOFOXIDATIONINPOLYETHYLENEFABRICATEDFORMSINTENDEDFORSURGICALIMPLANTS

链接地址:http://www.mydoc123.com/p-537754.html