ASTM F2102-2013 Standard Guide for Evaluating the Extent of Oxidation in Polyethylene Fabricated Forms Intended for Surgical Implants《评估预期供外科植入物使用的聚乙烯制品氧化程度的标准指南》.pdf
《ASTM F2102-2013 Standard Guide for Evaluating the Extent of Oxidation in Polyethylene Fabricated Forms Intended for Surgical Implants《评估预期供外科植入物使用的聚乙烯制品氧化程度的标准指南》.pdf》由会员分享,可在线阅读,更多相关《ASTM F2102-2013 Standard Guide for Evaluating the Extent of Oxidation in Polyethylene Fabricated Forms Intended for Surgical Implants《评估预期供外科植入物使用的聚乙烯制品氧化程度的标准指南》.pdf(5页珍藏版)》请在麦多课文档分享上搜索。
1、Designation: F2102 13Standard Guide forEvaluating the Extent of Oxidation in PolyethyleneFabricated Forms Intended for Surgical Implants1This standard is issued under the fixed designation F2102; the number immediately following the designation indicates the year oforiginal adoption or, in the case
2、of revision, the year of last revision. A number in parentheses indicates the year of last reapproval. Asuperscript epsilon () indicates an editorial change since the last revision or reapproval.1. Scope1.1 This guide covers a method for the measurement of therelative extent of oxidation present in
3、HDPE homopolymersand ultra-high-molecular-weight polyethylene (UHMWPE) in-tended for use in medical implants. The material is analyzed byinfrared spectroscopy. The intensity (area) of the carbonylabsorptions (C=O) centered near 1720 cm-1is related to theamount of chemically bound oxygen present in t
4、he material.Other forms of chemically bound oxygen (C-O-C, C-O-O-C,C-O-H, and so forth) are not captured by this guide.1.2 Although this guide may give the investigator a meansto compare the relative extent of carbonyl oxidation present invarious UHMWPE samples, it is recognized that other forms ofc
5、hemically bound oxygen may be important contributors tothese materials characteristics.1.3 The applicability of the infrared method has beendemonstrated by many literature reports. This particularmethod, using the intensity (area) of the C-H absorptioncentered near 1370 cm-1to normalize for the samp
6、lesthickness, has been validated by an Interlaboratory Study (ILS)conducted according to Practice E691.1.4 The following precautionary caveat pertains only to thetest method portion, Section 5, of this specification: Thisstandard may involve hazardous materials, operations, andequipment. This standa
7、rd does not purport to address all of thesafety concerns, if any, associated with its use. It is theresponsibility of the user of this standard to establish appro-priate safety and health practices and determine the applica-bility of regulatory requirements prior to use.2. Referenced Documents2.1 AS
8、TM Standards:2E691 Practice for Conducting an Interlaboratory Study toDetermine the Precision of a Test Method3. Terminology3.1 Definitions:3.1.1 bulk oxidation index (BOI)a samples bulk oxida-tion index (BOI) is the average of the oxidation indicescollected over a 500-m section at the center of the
9、 sample.3.1.1.1 DiscussionTypically, this is a plateau region withthe smallest oxidation indices.3.1.1.2 DiscussionFor samples less than about 8 to 10 mmthick, this central region may display the samples highestoxidation indices, depending on its state of oxidation.3.1.2 depth locator (DL)a measurem
10、ent of the distancefrom the articular surface, or surface of interest, that a spectrumwas collected and a corresponding OI calculated.3.1.3 oxidation index (OI)an oxidation index (OI) isdefined as the ratio of the area of the carbonyl absorptionpeak(s) centered near 1720 cm-1to the area of the absor
11、ptionpeak(s) centered near 1370 cm-1, as shown in Fig. 1. Note thatthe peak areas are computed after subtracting out the appro-priate baseline, as further discussed in Section 6.3.1.4 oxidation index profilean oxidation index profile isthe graphical representation of variation of the samplesoxidatio
12、n index with distance from its articular surface or thesurface of interest. This is a plot of an OI versus DL. Typically,the graph will show the profile through the entire thickness ofthe sample.3.1.5 surface oxidation index (SOI)a samples surfaceoxidation index (SOI) is the average of the oxidation
13、 indicesfrom the samples articular surface, or the surface of interest, toa depth of 3-mm subsurface.4. Apparatus4.1 Infrared Spectrometer:4.1.1 A calibrated infrared spectrometer capable of record-ing a transmission absorption spectrum over the range of about1200 to about 2000 cm-1using about 200-s
14、m-thick films at aresolution of 4 cm-1and an aperture of about 200 by 200 m.4.1.1.1 Other modes of collection (that is, percent reflection,attenuated total reflection (ATR), and so forth) and apertureand increment sizes may be used to generate the samples1This guide is under the jurisdiction of ASTM
15、 Committee F04 on Medical andSurgical Materials and Devicesand is the direct responsibility of SubcommitteeF04.15 on Material Test Methods.Current edition approved Nov. 1, 2013. Published December 2013. Originallyapproved in 2001. Last previous edition approved in 2006 as F2102 061. DOI:10.1520/F210
16、2-13.2For referenced ASTM standards, visit the ASTM website, www.astm.org, orcontact ASTM Customer Service at serviceastm.org. For Annual Book of ASTMStandards volume information, refer to the standards Document Summary page onthe ASTM website.Copyright ASTM International, 100 Barr Harbor Drive, PO
17、Box C700, West Conshohocken, PA 19428-2959. United States1absorption spectrum provided they can be demonstrated toproduce equivalent results. Too large an aperture can result ina loss of profile accuracy.4.1.1.2 When a Fourier Transform Infrared (FTIR) spec-trometer is used, a minimum of 32 scans sh
18、all be collected perspectrum.4.1.1.3 The FTIR instrument and sample compartment maybe purged with a moisture-free inert gas (for example, nitrogen,helium, or argon) to minimize spectral interference from thesecomponents.4.2 Specimen HolderEquipment capable of accuratelypositioning the sample under t
19、he orifice in increments at thescale of the aperture dimensions.4.3 MicrotomeEquipment capable of producing about200-m-thick slices (films) of a sample perpendicular to thearticular surface or the surface of interest.5. Procedure5.1 Preparation of the Infrared Spectrometer:5.1.1 Prepare the infrared
20、 spectrometer for collection of atransmission absorption spectrum from a thin film of theUHMWPE sample according to the manufacturers recommen-dations and the conditions described in Section 4 above.5.1.2 Collect the sequence of spectra per 5.2 and 5.3.5.2 Preparation of the Test Specimen:5.2.1 Usin
21、g a microtome, or other appropriate device, pre-pare a thin slice of the sample about 200 m thick.5.2.2 The slice shall be taken near the center of the samplesarticular surface or the surface of interest.5.2.3 The orientation of the slice shall typically be perpen-dicular to the articular surface or
22、 the surface of interest.5.2.4 For explanted components retrieved after in vivo useor in vitro samples that have been exposed to lipids (forexample, simulator specimens exposed to lubricants containingserum), the film should be submerged in a reagent (heptane orhexane) to extract lipids from the pol
23、ymer that interfere withthe carbonyl peak absorptions. The extraction technique shouldbe verified to confirm that the oxidation level has stabilized.5.3 Configuration of the Test Specimen in the Spectrometer:5.3.1 The test film (slice) shall be first configured in thespectrometer (after an appropria
24、te background spectrum hasbeen collected) such that the aperture is positioned over thefirst 200 m of the film starting at the surface of interest.5.3.2 Subsequent spectra shall be collected sequentially atincrements matching the aperture size (that is, about 200 m)from the articular surface, or sur
- 1.请仔细阅读文档,确保文档完整性,对于不预览、不比对内容而直接下载带来的问题本站不予受理。
- 2.下载的文档,不会出现我们的网址水印。
- 3、该文档所得收入(下载+内容+预览)归上传者、原创作者;如果您是本文档原作者,请点此认领!既往收益都归您。
下载文档到电脑,查找使用更方便
5000 积分 0人已下载
下载 | 加入VIP,交流精品资源 |
- 配套讲稿:
如PPT文件的首页显示word图标,表示该PPT已包含配套word讲稿。双击word图标可打开word文档。
- 特殊限制:
部分文档作品中含有的国旗、国徽等图片,仅作为作品整体效果示例展示,禁止商用。设计者仅对作品中独创性部分享有著作权。
- 关 键 词:
- ASTMF21022013STANDARDGUIDEFOREVALUATINGTHEEXTENTOFOXIDATIONINPOLYETHYLENEFABRICATEDFORMSINTENDEDFORSURGICALIMPLANTS

链接地址:http://www.mydoc123.com/p-537753.html