ASTM F2081-2006(2013) Standard Guide for Characterization and Presentation of the Dimensional Attributes of Vascular Stents《脉管展幅的尺寸特征的描述与介绍的标准指南》.pdf
《ASTM F2081-2006(2013) Standard Guide for Characterization and Presentation of the Dimensional Attributes of Vascular Stents《脉管展幅的尺寸特征的描述与介绍的标准指南》.pdf》由会员分享,可在线阅读,更多相关《ASTM F2081-2006(2013) Standard Guide for Characterization and Presentation of the Dimensional Attributes of Vascular Stents《脉管展幅的尺寸特征的描述与介绍的标准指南》.pdf(5页珍藏版)》请在麦多课文档分享上搜索。
1、Designation: F2081 06 (Reapproved 2013)Standard Guide forCharacterization and Presentation of the DimensionalAttributes of Vascular Stents1This standard is issued under the fixed designation F2081; the number immediately following the designation indicates the year oforiginal adoption or, in the cas
2、e of revision, the year of last revision. A number in parentheses indicates the year of last reapproval. Asuperscript epsilon () indicates an editorial change since the last revision or reapproval.1. Scope1.1 This guide covers the identification of and recom-mended measurement methods for those dime
3、nsional attributesof vascular stents that are deemed relevant to successfulclinical performance. The delivery system packaged with andlabeled specifically for use during the placement of the stent isalso included within the scope of this guide.1.2 This guide addresses only the dimensional characteri
4、s-tics of stents. Material property and stent functional character-istics are not addressed herein. All dimensional characteristicsdescribed in this guide refer to in vitro (“bench-top”) charac-terization. Because of variable patient factors, for example,vessel compliance, the actual in vivo charact
5、eristics may beslightly different.1.3 This guide includes recommendations generally appli-cable to balloon-expandable and self-expanding stents fabri-cated from metals and metal alloys. It does not specificallyaddress any attributes unique to coated stents or polymeric orbiodegradable stents, althou
6、gh the application of this guide tothose products is not precluded.1.4 While they are not specifically included within the scopeof this guide, stents indicated for placement in nonvascularlocations, such as the esophagus or bile duct, also might becharacterized by the methods contained herein. Likew
7、ise, thisguide does not include recommendations for endovasculargrafts (“stent-grafts”) or other conduit devices commonly usedto treat aneurysmal disease or peripheral vessel trauma or toprovide vascular access, although some information includedherein may be applicable to those devices.1.5 This gui
8、de does not include recommendations for bal-loon catheters sold as stand-alone angioplasty catheters, eventhough some of those catheters may be used for the delivery ofunmounted stents supplied without a delivery system. Require-ments for angioplasty catheters are contained in standards ISO10555-1 a
9、nd ISO 10555-4.1.6 The values stated in either SI units or inch-pound unitsare to be regarded separately as standard. The values stated ineach system may not be exact equivalents; therefore, eachsystem shall be used independently of the other. Combiningvalues from the two systems may result in non-c
10、onformancewith the standard.1.6.1 The units of measurements used throughout this guidereflect the hybrid system in common clinical use in the UnitedStates as of the time of the original approval of this guide.Since a primary purpose of this guide is to promote uniformityof labeling to facilitate the
11、 selection of devices by clinicalusers, the units most preferred by users were selected for thisguide. Where those units are not SI units, or derivativesthereof, SI units are provided in parentheses.2. Referenced Documents2.1 ISO Standards:ISO 10555-1, Sterile, Single-Use Intravascular Catheters,Gen
12、eral Requirements2ISO 10555-4, Sterile, Single-Use Intravascular CathetersBalloon Dilation Catheters23. Terminology3.1 Definitions of Terms Specific to This Standard:3.1.1 balloon-expandable stent, na stent that is expandedat the treatment site by a balloon catheter. The stent is alteredpermanently
13、by the balloon expansion such that the stentremains expanded after deflation of the balloon.3.1.2 bridge, na connecting element between the radialsupport aspects of a stent. A bridge may have unique designfeatures, as compared to a strut, to enhance longitudinalflexibility and minimize shortening.3.
14、1.3 crimp, vto secure the stent on the delivery system bycompressing the stent onto the balloon. Stents sold unmountedmust be crimped manually by the clinical staff before use.3.1.4 crossing profile, na linear measure of the maximumbreadth of the stent/delivery system over the distal-most regionof t
15、he delivery system.1This guide is under the jurisdiction of ASTM Committee F04 on Medical andSurgical Materials and Devicesand is the direct responsibility of SubcommitteeF04.30 on Cardiovascular Standards.Current edition approved March 1, 2013. Published March 2013. Originallyapproved in 2001. Last
16、 previous edition approved in 2006 as F2081 06. DOI:10.1520/F2081-06R13.2Available from American National Standards Institute (ANSI), 25 W. 43rd St.,4th Floor, New York, NY 10036, http:/www.ansi.org.Copyright ASTM International, 100 Barr Harbor Drive, PO Box C700, West Conshohocken, PA 19428-2959. U
17、nited States13.1.5 delivery system, na system that is used to deliver anddeploy a stent at the target site. A delivery system may besimilar to a balloon dilatation catheter; a delivery system for aself-expanding stent might not have a balloon.3.1.6 diameter, nrefers to the inside diameter (ID) of th
18、estent unless otherwise noted.3.1.7 diameter, labeled, nsingle value representation ofintended-use diameters, typically rounded to nearest 0.25 or0.5 mm. The labeled diameter is within the range recom-mended in the Instructions for Use (IFU) for that particularstent and delivery system.3.1.8 percent
19、 solid area, nthat percentage of the projectedcylindrical side surface area ( times outside diameter timeslength at labeled diameter) that is covered by stent material,when the stent is expanded to its labeled diameter.3.1.9 premounted stent, na stent supplied by a manufac-turer already mounted on a
20、 delivery system.3.1.10 self-expanding stent, na stent that expands withoutextrinsic force or pressure, to a size and shape close to thedesired final size and shape, when released from the deliverysystem. The self-expanding nature of some stents is a designfeature resulting from the materials of con
21、struction or thestructural geometry, or both.3.1.11 sheath, na movable cover that constrains a self-expanding stent on the delivery system until its desired releaseor protects a balloon-expandable stent during delivery beforedeployment.3.1.12 shortening/lengthening, nthe percentage change inlength b
22、etween the undeployed mounted condition and theexpanded labeled-diameter condition.3.1.13 stent, vascular, na synthetic tubular structure that ispermanently implanted in the native or grafted vasculature andthat is intended to provide mechanical radial support toenhance vessel patency. For the purpo
23、ses of this guide, a stentis metallic and not covered by synthetic textile or tissue graftmaterial.3.1.14 strut, nthe smallest individual element of the radialsupport aspect of a stent that has a solid cross section in boththe radial and circumferential directions.3.1.15 unmounted stent, na stent th
24、at is not crimped on adelivery system as supplied by the manufacturer. Before use,the clinical staff must manually crimp unmounted stents onto adelivery device.4. Significance and Use4.1 Vascular stents are intended for permanent implant inthe human vasculature (native or graft) for the purposes ofm
- 1.请仔细阅读文档,确保文档完整性,对于不预览、不比对内容而直接下载带来的问题本站不予受理。
- 2.下载的文档,不会出现我们的网址水印。
- 3、该文档所得收入(下载+内容+预览)归上传者、原创作者;如果您是本文档原作者,请点此认领!既往收益都归您。
下载文档到电脑,查找使用更方便
5000 积分 0人已下载
下载 | 加入VIP,交流精品资源 |
- 配套讲稿:
如PPT文件的首页显示word图标,表示该PPT已包含配套word讲稿。双击word图标可打开word文档。
- 特殊限制:
部分文档作品中含有的国旗、国徽等图片,仅作为作品整体效果示例展示,禁止商用。设计者仅对作品中独创性部分享有著作权。
- 关 键 词:
- ASTMF208120062013STANDARDGUIDEFORCHARACTERIZATIONANDPRESENTATIONOFTHEDIMENSIONALATTRIBUTESOFVASCULARSTENTS

链接地址:http://www.mydoc123.com/p-536505.html