《ASTM F1440-1992(2002) Standard Practice for Cyclic Fatigue Testing of Metallic Stemmed Hip Arthroplasty Femoral Components Without Torsion《无扭转装有金属杆的髋关节成型术股骨部件的循环疲劳检验》.pdf》由会员分享,可在线阅读,更多相关《ASTM F1440-1992(2002) Standard Practice for Cyclic Fatigue Testing of Metallic Stemmed Hip Arthroplasty Femoral Components Without Torsion《无扭转装有金属杆的髋关节成型术股骨部件的循环疲劳检验》.pdf(6页珍藏版)》请在麦多课文档分享上搜索。
1、Designation: F 1440 92 (Reapproved 2002)Standard Practice forCyclic Fatigue Testing of Metallic Stemmed Hip ArthroplastyFemoral Components Without Torsion1This standard is issued under the fixed designation F 1440; the number immediately following the designation indicates the year oforiginal adopti
2、on or, in the case of revision, the year of last revision. A number in parentheses indicates the year of last reapproval. Asuperscript epsilon (e) indicates an editorial change since the last revision or reapproval.1. Scope1.1 This practice describes a method for the fatigue testingof metallic stemm
3、ed femoral components used in hip arthro-plasty. The described method is intended to be used to evaluatethe comparison of various designs and materials used forstemmed femoral components used in the arthroplasty. Thispractice covers procedures for the performance of fatigue testsusing (as a forcing
4、function) a periodic constant amplitudeforce.1.2 This practice applies primarily to one-piece prosthesesand modular components, with head in place such that pros-theses should not have an anterior/posterior bow, and shouldhave a nearly straight section on the distal 50 mm of the stem.This practice m
5、ay require modifications to accommodate otherfemoral stem designs.1.3 The values stated in SI units are to be regarded as thestandard.1.4 For additional information see Refs. (1-5) .2. Referenced Documents2.1 ASTM Standards:E 4 Practices for Force Verification of Testing Machines2E 466 Practice for
6、Conducting Force Controlled ConstantAmplitude Axial Fatigue Tests of Metallic Materials2E 467 Practice for Verification of Constant Amplitude Dy-namic Forces in an Axial Fatigue Testing System2E 468 Practice for Presentation of Constant Amplitude Fa-tigue Test Results for Metallic Materials23. Termi
7、nology3.1 Definitions and Symbols (see Fig. 1(a) and 1(b):3.1.1 cantilever planea plane perpendicular to the line ofload application at the level on the stem where the stembecomes unsupported.3.1.2 distal stem axisthe centerline in the anterior/posterior projection of the most distal 50 mm of the st
8、em.3.1.3 estimated maximum bending momentthe maximumload times the unloaded moment arm.3.1.4 geometric centroid (cantilever plane) the point in across-sectional area of the cantilever plane whose coordinatesare the mean values of the coordinates of all the points in thearea.3.1.5 line of load applic
9、ationthe loading axis of the testmachine.3.1.6 Reference Line L1, distal stem axisthe medial-lateral (M-L) centerline of the most distal 50 mm of stem in theA-P projection.3.1.7 Reference Line L2:3.1.7.1 collared device the plane of the distal side of thecollar in the A-P projection.3.1.7.2 collarle
10、ss devicethe resection plane recommendedfor the device in the A-P projection.3.1.8 Reference Point P1the spherical center of the pros-thesis head.3.1.9 Reference Point P3:1This practice is under the jurisdiction of ASTM Committee F04 on Medical andSurgical Materials and Devices and is the direct res
11、ponsibility of SubcommitteeF04.22 on Arthroplasty.Current edition approved Sept. 15, 1992. Published November 1992.2Annual Book of ASTM Standards, Vol 03.01.FIG. 1 Collared Device1Copyright ASTM International, 100 Barr Harbor Drive, PO Box C700, West Conshohocken, PA 19428-2959, United States.3.1.9.
12、1 collared device the intersection of the principalaxis of the collar (L2) with the medial surface of the stem in theA-P projection.3.1.9.2 collarless devicethe intersection of the resectionplane (L2) with the medial surface of the stem in the A-Pprojection.3.1.10 Reference Point P4the distal tip of
13、 the stem.3.1.11 Reference Point P63 the intersection of the canti-lever plane with the medial surface of the stem in the A-Pprojection.3.1.12 R valuethe R value is the ratio of the minimumforce to the maximum force.R 5minimum forcemaximum force3.1.13 Stem Reference Angle Xthe angle between the stem
14、reference line and the line of load application.3.1.14 stem reference linea line passing through Refer-ence Point P6 and the center of the prosthesis head (P1).3.1.15 supported stem lengththe vertical distance be-tween the distal tip of the stem (P4) and the cantilever plane.3.1.16 unloaded moment a
15、rmthe perpendicular distancebetween the line of load application and the geometric centroidof the stem cross section at the cantilever plane.3.1.17 unsupported stem lengththe vertical distance be-tween Point P3 and the cantilever plane.3.2 See Figs. 1 and 2.4. Significance and Use4.1 This practice c
16、an be used to describe the effects ofmaterials, manufacturing, and design variables on the fatigueresistance of metallic stemmed femoral components subjectedto cyclic loading for relatively large numbers of cycles. Therecommended test assumes a “worst case situation whereproximal support for the ste
17、m has been lost. It is alsorecognized that for some materials the environment may havean effect on the response to cyclic loading. The test environ-ment used and the rationale for the choice of that environmentshould be described in the report.4.2 It is recognized that actual in vivo loading conditi
18、ons arenot constant amplitude. However, there is not sufficient infor-mation available to crate standard load spectrums for metallicstemmed femoral components. Accordingly, a simple periodicconstant amplitude force is recommended.4.3 In order for fatigue data on femoral stems to be usefulfor compari
19、son, it must be reproducible among differentlaboratories. Consequently, it is essential that uniform proce-dures be established.5. Specimen Selection5.1 The specimen selection should have the same geometryas the final finished product, and the stem should be in the finalfinished condition.6. Apparat
20、us6.1 The specimen shall be constrained by a suitable groutingagent within a rigid cavity. A common grouting agent used ispoly methyl methacrylate (PMMAbone cement) that ispolymerized in place. The minimum thickness of the groutingagent should be 1 cm. Although bone cement is the recom-mended grouti
21、ng agent, other material may be used provided itdoes not chemically or mechanically interact with the testspecimen.6.2 The test fixtures shall be constructed so that the line ofload application is in the implant anterior/posterior symmetryplane of the supported portion of the stem.6.3 The test fixtu
22、res shall be constructed so that the line ofload application passes through the ball center.6.4 A ball- or roller-bearing low-friction mechanism shallbe included in the loading apparatus to minimize loads notperpendicular to the cantilever plane. An example of such amechanism is included in Appendix
23、 X1.7. Equipment Characteristics7.1 The action of the machine should be analyzed to ensurethat the desired form and periodic force amplitude is main-tained for the duration of the test. (See Practice E 467.)7.2 The test machine should have a load monitoring systemsuch as the transducer mounted in li
24、ne with the specimen. Thetest loads should be monitored continuously in the early stagesof the test and periodically thereafter to ensure the desired loadcycle is maintained. The varying load as determined bysuitable dynamic verification should be maintained at all timesto within 62 % of the maximum
- 1.请仔细阅读文档,确保文档完整性,对于不预览、不比对内容而直接下载带来的问题本站不予受理。
- 2.下载的文档,不会出现我们的网址水印。
- 3、该文档所得收入(下载+内容+预览)归上传者、原创作者;如果您是本文档原作者,请点此认领!既往收益都归您。
下载文档到电脑,查找使用更方便
5000 积分 0人已下载
下载 | 加入VIP,交流精品资源 |
- 配套讲稿:
如PPT文件的首页显示word图标,表示该PPT已包含配套word讲稿。双击word图标可打开word文档。
- 特殊限制:
部分文档作品中含有的国旗、国徽等图片,仅作为作品整体效果示例展示,禁止商用。设计者仅对作品中独创性部分享有著作权。
- 关 键 词:
- ASTMF144019922002STANDARDPRACTICEFORCYCLICFATIGUETESTINGOFMETALLICSTEMMEDHIPARTHROPLASTYFEMORALCOMPONENTSWITHOUTTORSION

链接地址:http://www.mydoc123.com/p-534888.html