《ASTM F1350-2015 Standard Specification for Wrought 18Chromium-14Nickel-2 5Molybdenum Stainless Steel Surgical Fixation Wire (UNS S31673)《锻造18铬-14镍-2 5钼不锈钢外科固定用线的标准规格 (UNS S31673)》.pdf》由会员分享,可在线阅读,更多相关《ASTM F1350-2015 Standard Specification for Wrought 18Chromium-14Nickel-2 5Molybdenum Stainless Steel Surgical Fixation Wire (UNS S31673)《锻造18铬-14镍-2 5钼不锈钢外科固定用线的标准规格 (UNS S31673)》.pdf(5页珍藏版)》请在麦多课文档分享上搜索。
1、Designation: F1350 08F1350 15Standard Specification forWrought 18Chromium-14Nickel-2.5Molybdenum StainlessSteel Surgical Fixation Wire (UNS S31673)1This standard is issued under the fixed designation F1350; the number immediately following the designation indicates the year oforiginal adoption or, i
2、n the case of revision, the year of last revision. A number in parentheses indicates the year of last reapproval. Asuperscript epsilon () indicates an editorial change since the last revision or reapproval.1. Scope*1.1 This specification covers the chemical, mechanical, and metallurgical requirement
3、s for the manufacture of wrought 18chromium-14 nickel-2.5 molybdenum stainless steel in the form of surgical fixation wire.1.2 UnitsThe values stated in either SI units or inch-pound units are to be regarded separately as the standard. The inch-poundvalues in parentheses are for information only.sta
4、ted in each system may not be exact equivalents; therefore, each system shall beused independently of the other. Combining values from the two systems may result in non-conformance with this specification.2. Referenced Documents2.1 ASTM Standards:2A555/A555M Specification for General Requirements fo
5、r Stainless Steel Wire and Wire RodsE8E8/E8M Test Methods for Tension Testing of Metallic MaterialsE29 Practice for Using Significant Digits in Test Data to Determine Conformance with SpecificationsF86 Practice for Surface Preparation and Marking of Metallic Surgical ImplantsF138 Specification for W
6、rought 18Chromium-14Nickel-2.5Molybdenum Stainless Steel Bar and Wire for Surgical Implants(UNS S31673)F981 Practice for Assessment of Compatibility of Biomaterials for Surgical Implants with Respect to Effect of Materials onMuscle and BoneIEEE/ASTM SI 10 American National Standard for Metric Practi
7、ce2.2 USP Standard:3Nonabsorbable Surgical Suture, U.S. Pharmacopeia2.3 ISO Standard:4ISO 9001 Quality Management SystemRequirements2.4 American Society for Quality Standard:5ASQ C1 Specification of General Requirements for a Quality Program3. General Requirements for Delivery3.1 In addition to the
8、requirements of this specification, all requirements of the current editions of Specifications A555/A555Mand F138 apply.3.2 In cases where a conflict exists between this specification and the standards listed in Section 2, this specification shall takeprecedence.4. Terminology4.1 Definitions of Term
9、s Specific to This Standard:1 This specification is under the jurisdiction of ASTM Committee F04 on Medical and Surgical Materials and Devices and is the direct responsibility of SubcommitteeF04.12 on Metallurgical Materials.Current edition approved May 1, 2008May 1, 2015. Published June 2008 June 2
10、015. Originally approved in 1991. Last previous edition approved in 20022008 asF1350 02.F1350 08. DOI: 10.1520/F1350-08.10.1520/F1350-15.2 For referencedASTM standards, visit theASTM website, www.astm.org, or contactASTM Customer Service at serviceastm.org. For Annual Book of ASTM Standardsvolume in
11、formation, refer to the standards Document Summary page on the ASTM website.3 Available from U.S. Pharmacopeia (USP), 12601 Twinbrook Pkwy., Rockville, MD 20852-1790, http:/www.usp.org.4 Available from American National Standards Institute (ANSI), 25 W. 43rd St., 4th Floor, New York, NY 10036, http:
12、/www.ansi.org.This document is not an ASTM standard and is intended only to provide the user of an ASTM standard an indication of what changes have been made to the previous version. Becauseit may not be technically possible to adequately depict all changes accurately, ASTM recommends that users con
13、sult prior editions as appropriate. In all cases only the current versionof the standard as published by ASTM is to be considered the official document.*A Summary of Changes section appears at the end of this standardCopyright ASTM International, 100 Barr Harbor Drive, PO Box C700, West Conshohocken
14、, PA 19428-2959. United States14.1.1 lot, nthe total number of mill products produced from the same melt heat under the same conditions at essentially thesame time.5. Ordering Information5.1 Inquiries and orders for material under this specification shall include the following information:5.1.1 Quan
15、tity,5.1.2 ASTM designation and date of issue,5.1.3 Material requirements,5.1.4 Mechanical properties,5.1.5 Form,5.1.6 Dimensional requirements, including diameter and diameter tolerance,5.1.7 Surface condition and handling,5.1.8 Special tests (if applicable), and5.1.9 Other requirements.6. Material
16、 Requirements6.1 The starting material used to make fixation wire must meet the requirements of Specification F138.6.2 Surgical fixation wire shall conform to the specified chemical requirements of Specification F138. Conformance with thisstandard shall be so identified by suitable packaging, labeli
17、ng, or both.7. Mechanical Requirements7.1 Surgical fixation wire shall conform to the appropriate mechanical properties specified in Table 1.7.2 Perform tension tests in accordance with Test Methods E8E8/E8M using a 254 mm (10 in.)10 in. gage length and across-head speed of 254 mm/min (10 in./min).1
18、0 in./min. Should any of the test specimens not meet the specified requirements,test two additional test pieces representative of the same lot, in the same manner, for each failed test piece. The lot shall beconsidered in compliance only if all additional test pieces meet the specified requirements.
19、7.3 Tensile test results for which any specimen fractures outside the gage length shall be considered acceptable if the elongationmeets the minimum requirements specified in Table 1. Refer to subsections 7.11.4 and 7.11.5 of Test Methods E8E8/E8M. If theelongation is less than the minimum requiremen
20、t, discard the test and retest. Retest one specimen for each specimen that did notmeet the minimum requirement.7.4 The wire shall meet the requirements of the latest version of USP for Nonabsorbable Surgical Sutures, when tested inaccordance with 7.2.8. Dimensional Requirements8.1 Surgical fixation
21、wire shall be fabricated in accordance with the dimensions and tolerances specified in Table 1.8.2 Unless otherwise specified, size tolerances are plus and minus as shown in Table 1. When required by the purchaser, roundwire tolerances may be specified all plus and nothing minus, or all minus and no
22、thing plus, or any combination of plus and minusif the total spread in size tolerance is not less than the total spread shown in Table 1.8.3 The maximum out-of-round tolerance for round wire is one-half of the size tolerance given in Table 1.8.4 Units of Measure8.4.1 SelectionThis specification requ
23、ires that the purchaser select the units (SI or inch-pound) to be used for productclarification. In the absence of a stated selection of units on the purchase order, this selection may be expressed by the purchaserin several alternate forms listed in order of precedence.8.4.1.1 If the purchaser and
24、supplier have a history of using specific units, these units shall continue to be certified untilexpressly changed by the purchaser.8.4.1.2 In the absence of historic precedence, if the units used to define the product on the purchasers purchase order (PO) areconsistent with the specification and en
- 1.请仔细阅读文档,确保文档完整性,对于不预览、不比对内容而直接下载带来的问题本站不予受理。
- 2.下载的文档,不会出现我们的网址水印。
- 3、该文档所得收入(下载+内容+预览)归上传者、原创作者;如果您是本文档原作者,请点此认领!既往收益都归您。
下载文档到电脑,查找使用更方便
5000 积分 0人已下载
下载 | 加入VIP,交流精品资源 |
- 配套讲稿:
如PPT文件的首页显示word图标,表示该PPT已包含配套word讲稿。双击word图标可打开word文档。
- 特殊限制:
部分文档作品中含有的国旗、国徽等图片,仅作为作品整体效果示例展示,禁止商用。设计者仅对作品中独创性部分享有著作权。
- 关 键 词:
- ASTMF13502015STANDARDSPECIFICATIONFORWROUGHT18CHROMIUM14NICKEL25MOLYBDENUMSTAINLESSSTEELSURGICALFIXATIONWIREUNSS31673

链接地址:http://www.mydoc123.com/p-534717.html