ASTM E967-2008 923 Standard Practice for Temperature Calibration of Differential Scanning Calorimeters and Differential Thermal Analyzers《差示扫描量热仪与差热分析仪温度校准的标准实施规范》.pdf
《ASTM E967-2008 923 Standard Practice for Temperature Calibration of Differential Scanning Calorimeters and Differential Thermal Analyzers《差示扫描量热仪与差热分析仪温度校准的标准实施规范》.pdf》由会员分享,可在线阅读,更多相关《ASTM E967-2008 923 Standard Practice for Temperature Calibration of Differential Scanning Calorimeters and Differential Thermal Analyzers《差示扫描量热仪与差热分析仪温度校准的标准实施规范》.pdf(4页珍藏版)》请在麦多课文档分享上搜索。
1、Designation: E 967 08Standard Test Method forTemperature Calibration of Differential ScanningCalorimeters and Differential Thermal Analyzers1This standard is issued under the fixed designation E 967; the number immediately following the designation indicates the year oforiginal adoption or, in the c
2、ase of revision, the year of last revision. A number in parentheses indicates the year of last reapproval. Asuperscript epsilon () indicates an editorial change since the last revision or reapproval.1. Scope1.1 This test method describes the temperature calibrationof differential thermal analyzers a
3、nd differential scanningcalorimeters over the temperature range from 40 to +2500 C.1.2 The values stated in SI units are to be regarded asstandard. No other units of measurement are included in thisstandard.1.3 This test method is similar to ISO standard 113571.1.4 This standard does not purport to
4、address all of thesafety concerns, if any, associated with its use. It is theresponsibility of the user of this standard to establish appro-priate safety and health practices and determine the applica-bility of regulatory limitations prior to use. Specific precau-tionary statements are given in Sect
5、ion 7.2. Referenced Documents2.1 ASTM Standards:2E 473 Terminology Relating to Thermal Analysis and Rhe-ologyE 1142 Terminology Relating to Thermophysical Properties2.2 ISO Standard:113571 Plastics-Differential Scanning Calorimetry (DSC)-Part 1: General Principles3. Terminology3.1 Specific technical
6、 terms used in this test method aredefined in Terminologies E 473 and E 1142.4. Summary of Test Method4.1 This test method consists of heating the calibrationmaterials at a controlled rate in a controlled atmospherethrough a region of known thermal transition. The heat flowinto the calibration mater
7、ial or the difference of temperaturebetween the calibration material and a reference sample and areference material is monitored and continuously recorded. Atransition is marked by the absorption of energy by thespecimen resulting in a corresponding endothermic peak in theheating curve.NOTE 1Heat fl
8、ow calibrations are sometimes determined in conjunc-tion with temperature calibration. Some differential scanning calorimeterspermit both heat flow and temperature calibrations to be obtained from thesame experimental procedure.5. Significance and Use5.1 Differential scanning calorimeters and differ
9、ential ther-mal analyzers are used to determine the transition temperaturesof materials. For this information to be meaningful in anabsolute sense, temperature calibration of the apparatus orcomparison of the resulting data to that of known standardmaterials is required.5.2 This test method is usefu
10、l in calibrating the temperatureaxis of differential scanning calorimeters and differential ther-mal analyzers.6. Apparatus6.1 Apparatus shall be of either type listed below:6.1.1 Differential Scanning Calorimeter (DSC), capable ofheating a test specimen and a reference material at a controlledrate
11、and of automatically recording the differential heat flowbetween the sample and the reference material to the requiredsensitivity and precision.6.1.1.1 A Furnace(s), to provide uniform controlled heatingor cooling of a specimen and reference to a constant tempera-ture or at a constant rate within th
12、e applicable temperaturerange of this test method.6.1.1.2 A Temperature Sensor, to provide an indication ofthe specimen temperature.6.1.1.3 Differential sensors, to detect a heat flow (power)difference between the specimen and reference.6.1.1.4 Test Chamber Environment, a means of sustaining atest c
13、hamber environment of nitrogen or other inert purge gasat a purge rate of 10 to 50 mL/min.1This test method is under the jurisdiction ofASTM Committee E37 on ThermalMeasurements and is the direct responsibility of Subcommittee E37.01 on ThermalTest Methods and Practices.Current edition approved Sept
14、. 1, 2008. Published October 2008. Originallyapproved in 1983. Last previous edition approved in 2003 as E 967 03.2For referenced ASTM standards, visit the ASTM website, www.astm.org, orcontact ASTM Customer Service at serviceastm.org. For Annual Book of ASTMStandards volume information, refer to th
15、e standards Document Summary page onthe ASTM website.1Copyright ASTM International, 100 Barr Harbor Drive, PO Box C700, West Conshohocken, PA 19428-2959, United States.6.1.1.5 A Temperature Controller, capable of executing aspecific temperature program by operating the furnace(s)between selected tem
16、perature limits at a rate of temperaturechange of 10K/min.6.1.1.6 Data Collection Device, to provide a means ofacquiring, storing, and displaying measured or calculatedsignals, or both. The minimum output signals required for DSCare heat flow, temperature, and time.6.1.2 Differential Thermal Analyze
17、r (DTA), capable of heat-ing a test specimen and reference material at a controlled rateand of automatically recording the differential temperaturebetween sample and reference material both to the requiredsensitivity and precision.6.2 Containers (pans, crucibles, vials, lids, closures, seals,etc.),
18、that are inert to the specimen and reference materials andthat are of suitable structural shape and integrity to contain thespecimen and reference in accordance with the specific require-ments of this test method.6.3 Nitrogen, or other inert purge gas supply.6.4 A Balance, to weigh specimens or cont
19、ainers (pans,crucibles, vials, etc.), or both to 6 0.1 mg. The balance shouldhave a capacity greater than 20 mg.7. Precautions7.1 Toxic or corrosive effluents, or both, may be releasedwhen heating some material and could be harmful to personneland to apparatus.7.2 This test method assumes linear tem
20、perature indication.Care must be taken in the application of this test method toensure that calibration points are taken sufficiently closetogether so that linear temperature indication may be approxi-mated. Linear temperature indications means that there exists alinear, or first order, dependence o
21、n the temperature determinedby the instruments temperature sensor on the true temperatureof the sample material in its container and that this relation isadequately expressed by Eq 1.8. Calibration Materials8.1 For the temperature range covered by many applica-tions, the melting transition of 99.99
22、% pure materials inTable 1 may be used for calibration.9. Procedure9.1 Two Point Calibration:9.1.1 Select two calibration materials from Table 1, withmelting temperatures one above and one below the temperaturerange of interest. The calibration materials should be as closeto the temperature range of
23、 interest as practical.9.1.2 Determine the apparent transition temperature foreach calibration material.9.1.2.1 Into a clean specimen holder, placea5to15-mgweighed amount of calibration material. Other specimenmasses may be used but must be indicated in the report.9.1.2.2 Load the specimen into the
24、instrument chamber,purge the chamber with dry nitrogen (or other inert gas) at aflow rate of 10 to 50 cm3/min throughout the experiment.9.1.2.3 Heat (or cool) the calibration material rapidly to 30C below the calibration temperature and allow to stabilize.9.1.2.4 Heat the calibration material at 10
- 1.请仔细阅读文档,确保文档完整性,对于不预览、不比对内容而直接下载带来的问题本站不予受理。
- 2.下载的文档,不会出现我们的网址水印。
- 3、该文档所得收入(下载+内容+预览)归上传者、原创作者;如果您是本文档原作者,请点此认领!既往收益都归您。
下载文档到电脑,查找使用更方便
5000 积分 0人已下载
下载 | 加入VIP,交流精品资源 |
- 配套讲稿:
如PPT文件的首页显示word图标,表示该PPT已包含配套word讲稿。双击word图标可打开word文档。
- 特殊限制:
部分文档作品中含有的国旗、国徽等图片,仅作为作品整体效果示例展示,禁止商用。设计者仅对作品中独创性部分享有著作权。
- 关 键 词:
- ASTME9672008923STANDARDPRACTICEFORTEMPERATURECALIBRATIONOFDIFFERENTIALSCANNINGCALORIMETERSANDDIFFERENTIALTHERMALANALYZERS

链接地址:http://www.mydoc123.com/p-533863.html