ASTM E2593-2012 Standard Guide for Accuracy Verification of Industrial Platinum Resistance Thermometers《工业用铂电阻温度计的精度验证的标准指南》.pdf
《ASTM E2593-2012 Standard Guide for Accuracy Verification of Industrial Platinum Resistance Thermometers《工业用铂电阻温度计的精度验证的标准指南》.pdf》由会员分享,可在线阅读,更多相关《ASTM E2593-2012 Standard Guide for Accuracy Verification of Industrial Platinum Resistance Thermometers《工业用铂电阻温度计的精度验证的标准指南》.pdf(17页珍藏版)》请在麦多课文档分享上搜索。
1、Designation: E2593 111 E2593 12Standard Guide forAccuracy Verification of Industrial Platinum ResistanceThermometers1This standard is issued under the fixed designation E2593; the number immediately following the designation indicates the year oforiginal adoption or, in the case of revision, the yea
2、r of last revision. A number in parentheses indicates the year of last reapproval. Asuperscript epsilon () indicates an editorial change since the last revision or reapproval.1 NOTEThe previous epsilon note was removed in July 2011.1. Scope1.1 This guide describes the techniques and apparatus requir
3、ed for the accuracy verification of industrial platinum resistancethermometers constructed in accordance with Specification E1137/E1137M and the evaluation of calibration uncertainties. Theprocedures described apply over the range of 200C to 650C.1.2 This guide does not intend to describe procedures
4、 necessary for the calibration of platinum resistance thermometers usedas calibration standards or Standard Platinum Resistance Thermometers. Consequently, calibration of these types of instrumentsis outside the scope of this guide.1.3 Industrial platinum resistance thermometers are available in man
5、y styles and configurations. This guide does not purport todetermine the suitability of any particular design, style, or configuration for calibration over a desired temperature range.1.4 The evaluation of uncertainties is based upon current international practices as described in ISO/TAG 4/WG 3 “Gu
6、ide tothe Evaluation of Uncertainty in Measurement” and ANSI/NCSL Z540-2-1997 “U.S. Guide to the Expression of Uncertainty inMeasurement.”1.5 This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibilityof the user of this standard to
7、 establish appropriate safety and health practices and determine the applicability of regulatorylimitations prior to use.2. Referenced Documents2.1 ASTM Standards:2E344 Terminology Relating to Thermometry and HydrometryE563 Practice for Preparation and Use of an Ice-Point Bath as a Reference Tempera
8、tureE644 Test Methods for Testing Industrial Resistance ThermometersE1137/E1137M Specification for Industrial Platinum Resistance ThermometersE1502 Guide for Use of Fixed-Point Cells for Reference TemperaturesE1750 Guide for Use of Water Triple Point Cells2.2 ANSI Publication:ANSI/NCSL Z540-2-1997 U
9、.S. Guide to the Expression of Uncertainty in Measurement32.3 Other Publication:ISO/TAG 4/WG 3 Guide to the Evaluation of Uncertainty in Measurement3. Terminology3.1 DefinitionsThe definitions given in Terminology E344 shall be considered as applying to the terms used in this guide.3.2 Definitions o
10、f Terms Specific to This Standard:3.2.1 annealing, va heat treating process intended to stabilize resistance thermometers prior to calibration and use.1 This guide is under the jurisdiction of ASTM Committee E20 on Temperature Measurement and is the direct responsibility of Subcommittee E20.03 on Re
11、sistanceThermometers.Current edition approved May 1, 2011Nov. 1, 2012. Published June 2011December 2012. Originally approved in 2007. Last previous edition approved in 20072011 asD545607E01.11E01. DOI: 10.1520/E2593-11E01.10.1520/E2593-12.2 For referenced ASTM standards, visit the ASTM website, www.
12、astm.org, or contact ASTM Customer Service at serviceastm.org. For Annual Book of ASTM Standardsvolume information, refer to the standards Document Summary page on the ASTM website.3 Available from American National Standards Institute (ANSI), 25 W. 43rd St., 4th Floor, New York, NY 10036, http:/www
13、.ansi.org.This document is not an ASTM standard and is intended only to provide the user of an ASTM standard an indication of what changes have been made to the previous version. Becauseit may not be technically possible to adequately depict all changes accurately, ASTM recommends that users consult
14、 prior editions as appropriate. In all cases only the current versionof the standard as published by ASTM is to be considered the official document.Copyright ASTM International, 100 Barr Harbor Drive, PO Box C700, West Conshohocken, PA 19428-2959. United States13.2.2 check standard, na thermometer s
15、imilar in design to the unit under test, but of superior stability, which is included inthe calibration process for the purpose of quantifying the process variability.3.2.3 coverage factor, nnumerical factor used as a multiplier of the combined standard uncertainty in order to obtain anexpanded unce
16、rtainty.3.2.4 dielectric absorption, nan effect in an insulator caused by the polarization of positive and negative charges within theinsulator which manifests itself as an in-phase current when the voltage is removed and the charges recombine.3.2.5 expanded uncertainty, U, nquantity defining an int
17、erval about the result of a measurement that may be expected toencompass a large fraction of the distribution of values that could reasonably be attributed to the measurand.3.2.5.1 DiscussionNormally, U is given at a coverage factor of 2, approximating to a 95 % confidence interval.3.2.6 hysteresis,
18、 nproperty associated with the resistance of a thermometer whereby the value of resistance at a temperatureis dependant upon previous exposure to different temperatures.3.2.7 normal distribution, na frequency distribution characterized by a bell shaped curve and defined by two parameters: meanand st
19、andard deviation.3.2.8 platinum resistance thermometer (PRT), na resistance thermometer with the resistance element constructed fromplatinum or platinum alloy.3.2.9 rectangular distribution, na frequency distribution characterized by a rectangular shaped curve and defined by twoparameters: mean and
20、magnitude (semi-range).3.2.10 standard deviation of the mean, nan estimate of the standard deviation of the sampling distribution of means, basedon the data from one or more random samples.3.2.10.1 DiscussionNumerically, it is equal to the standard deviation obtained (s) when divided by the square r
21、oot of the size of the sample (n).Standard Deviation of the Mean5 s=n(1)3.2.11 standard platinum resistance thermometer (SPRT), na specialized platinum resistance thermometer constructed in sucha way that it fulfills the requirements of the ITS-90.43.2.12 standard uncertainty, nuncertainty of the re
22、sult of a measurement expressed as a standard deviation, designated as S.3.2.13 Type A evaluation (of uncertainty), nmethod of evaluation of uncertainty by the statistical analysis of a series ofobservations.3.2.14 Type B evaluation (of uncertainty), nmethod of evaluation of uncertainty by means oth
23、er than statistical analysis of aseries of observations.3.2.15 test uncertainty ratio (TUR), nthe ratio of the tolerance of the unit under test to the expanded calibration uncertainty.3.2.16 uncertainty budget, nan analysis tool used for assembling and combining component uncertainties expected in a
24、measurement process into an overall expected uncertainty.3.2.17 unit under test (UUT), nthe platinum resistance thermometer to be calibrated.4. Summary of Guide4.1 The UUT is calibrated by determining the electrical resistance of its sensing element at one or more known temperaturescovering the temp
- 1.请仔细阅读文档,确保文档完整性,对于不预览、不比对内容而直接下载带来的问题本站不予受理。
- 2.下载的文档,不会出现我们的网址水印。
- 3、该文档所得收入(下载+内容+预览)归上传者、原创作者;如果您是本文档原作者,请点此认领!既往收益都归您。
下载文档到电脑,查找使用更方便
5000 积分 0人已下载
下载 | 加入VIP,交流精品资源 |
- 配套讲稿:
如PPT文件的首页显示word图标,表示该PPT已包含配套word讲稿。双击word图标可打开word文档。
- 特殊限制:
部分文档作品中含有的国旗、国徽等图片,仅作为作品整体效果示例展示,禁止商用。设计者仅对作品中独创性部分享有著作权。
- 关 键 词:
- ASTME25932012STANDARDGUIDEFORACCURACYVERIFICATIONOFINDUSTRIALPLATINUMRESISTANCETHERMOMETERS 工业 铂电阻 温度计

链接地址:http://www.mydoc123.com/p-531504.html