《ASTM E2491-2008 Standard Guide for Evaluating Performance Characteristics of Phased-Array Ultrasonic Testing Instruments and Systems《评估性能特征的相控接收阵列超声波测试仪器和系统的标准指南》.pdf》由会员分享,可在线阅读,更多相关《ASTM E2491-2008 Standard Guide for Evaluating Performance Characteristics of Phased-Array Ultrasonic Testing Instruments and Systems《评估性能特征的相控接收阵列超声波测试仪器和系统的标准指南》.pdf(18页珍藏版)》请在麦多课文档分享上搜索。
1、Designation: E 2491 08Standard Guide forEvaluating Performance Characteristics of Phased-ArrayUltrasonic Testing Instruments and Systems1This standard is issued under the fixed designation E 2491; the number immediately following the designation indicates the year oforiginal adoption or, in the case
2、 of revision, the year of last revision. A number in parentheses indicates the year of last reapproval. Asuperscript epsilon () indicates an editorial change since the last revision or reapproval.1. Scope1.1 This guide describes procedures for evaluating someperformance characteristics of phased-arr
3、ay ultrasonic exami-nation instruments and systems.1.2 Evaluation of these characteristics is intended to be usedfor comparing instruments and systems or, by periodic repeti-tion, for detecting long-term changes in the characteristics of agiven instrument or system that may be indicative of impend-i
4、ng failure, and which, if beyond certain limits, will requirecorrective maintenance. Instrument characteristics measured inaccordance with this guide are expressed in terms that relate totheir potential usefulness for ultrasonic examinations. Otherelectronic instrument characteristics in phased-arra
5、y units aresimilar to non-phased-array units and may be measured asdescribed in E 1065 or E 1324.1.3 Ultrasonic examination systems using pulsed-wavetrains and A-scan presentation (rf or video) may be evaluated.1.4 This guide establishes no performance limits for exami-nation systems; if such accept
6、ance criteria are required, thesemust be specified by the using parties. Where acceptancecriteria are implied herein they are for example only and aresubject to more or less restrictive limits imposed by customersand end users controlling documents.1.5 The specific parameters to be evaluated, condit
7、ions andfrequency of test, and report data required, must also bedetermined by the user.1.6 This guide may be used for the evaluation of a completeexamination system, including search unit, instrument, inter-connections, scanner fixtures and connected alarm and auxil-iary devices, primarily in cases
8、 where such a system is usedrepetitively without change or substitution. This guide is notintended to be used as a substitute for calibration or standard-ization of an instrument or system to inspect any givenmaterial.1.7 Required test apparatus includes selected test blocks andposition encoders in
9、addition to the instrument or system to beevaluated.1.8 Precautions relating to the applicability of the proce-dures and interpretation of the results are included.1.9 Alternate procedures, such as examples described in thisdocument, or others, may only be used with customer approval.1.10 The values
10、 stated in SI units are to be regarded asstandard. No other units of measurement are included in thisstandard.1.11 This standard does not purport to address all of thesafety concerns, if any, associated with its use. It is theresponsibility of the user of this standard to establish appro-priate safe
11、ty and health practices and determine the applica-bility of regulatory limitations prior to use.2. Referenced Documents2.1 ASTM Standards:2E 317 Practice for Evaluating Performance Characteristicsof Ultrasonic Pulse-Echo Testing Instruments and Systemswithout the Use of Electronic Measurement Instru
12、mentsE 494 Practice for Measuring Ultrasonic Velocity in Mate-rialsE 1065 Guide for Evaluating Characteristics of UltrasonicSearch UnitsE 1316 Terminology for Nondestructive ExaminationsE 1324 Guide for Measuring Some Electronic Characteris-tics of Ultrasonic Examination Instruments3. Terminology3.1
13、 Refer to Terminology E 1316 for definitions of terms inthis guide3.2 Definitions:3.2.1 angle corrected gainalso called ACG. Is compensa-tion applied to focal laws in an S-scan to correct for the effectsof echo-transmittance variation at different angles. This may beaccomplished by equalizing the am
14、plitude response in the farfield from a uniform reflector at a constant sound path throughthe range of angles used in the S-scan.An “infinite radius” suchas that provided by the 100 mm radius of the IIW block is aconvenient target for this function.Aseries of side drilled holes1This guide is under t
15、he jurisdiction of ASTM Committee E07 on Nondestruc-tive Testing and is the direct responsibility of Subcommittee E07.06 on UltrasonicMethod.Current edition approved July 1, 2008. Published August 2008. Originallyapproved in 2006. Last previous edition approved in 2006 as E 2491 - 06.2For referenced
16、 ASTM standards, visit the ASTM website, www.astm.org, orcontact ASTM Customer Service at serviceastm.org. For Annual Book of ASTMStandards volume information, refer to the standards Document Summary page onthe ASTM website.1Copyright ASTM International, 100 Barr Harbor Drive, PO Box C700, West Cons
17、hohocken, PA 19428-2959, United States.arranged in a radiused pattern may also present uniformreflectors at a constant sound path but the corrections are thenin angular increments. Note that there are technical limits toACG, that is, beyond a certain angular range, compensation isnot possible.3.2.2
18、annular array probesphased-array probes that havethe transducers configured as a set of concentric rings. Theyallow the beam to be focused to different depths along an axis.The surface area of the rings is in most cases constant, whichimplies a different width for each ring.3.2.3 array (phased)a pat
19、terned arrangement of elements.Typical arrangements include linear, annular, two dimensionalmatrix, and “rho-theta”.3.2.4 electronic scanalso termed an E-scan. The samefocal law is multiplexed across a group of active elements;electronic raster scanning is performed at a constant angle andalong the
20、phased-array probe length. This is equivalent to aconventional ultrasonic probe performing a raster scan. Alsocalled electronic scanning.3.2.5 focal lawthe entire set of hardware and softwareparameters affecting the acoustic sensitivity field of a phasedarray search unit, whether a pulse-echo or a p
21、itch-catchconfiguration. Within focal laws, there are included delay lawsin transmitter and delay laws in receiver, as well as apodizationlaws, and element activation laws.3.2.6 linear array probesprobes made using a set ofelements juxtaposed and aligned along a linear axis. Theyenable a beam to be
22、moved, focused, and deflected along asingle azimuthal plane.3.2.7 matrix array probesthese probes have an active areadivided in two dimensions in different elements. This divisioncan, for example, be in the form of a checkerboard, or sectoredrings. These probes allow the ultrasonic beam steering in
23、morethan one plane.3.2.8 sectorial scanalso termed an S-scan or azimuthalscan. This may refer to either the beam movement or the datadisplay. As a data display it is a 2D view of all A-scans froma specific set of elements corrected for delay and refractedangle. When used to refer to the beam movemen
24、t it refers to theset of focal laws that sweeps a defined range of angles using thesame set of elements.3.2.9 S-scan (q.v. sectorial scan)4. Summary of Guide4.1 Phased-array instruments and systems have similar in-dividual components as are found in traditional ultrasonicsystems that are based on si
- 1.请仔细阅读文档,确保文档完整性,对于不预览、不比对内容而直接下载带来的问题本站不予受理。
- 2.下载的文档,不会出现我们的网址水印。
- 3、该文档所得收入(下载+内容+预览)归上传者、原创作者;如果您是本文档原作者,请点此认领!既往收益都归您。
下载文档到电脑,查找使用更方便
5000 积分 0人已下载
下载 | 加入VIP,交流精品资源 |
- 配套讲稿:
如PPT文件的首页显示word图标,表示该PPT已包含配套word讲稿。双击word图标可打开word文档。
- 特殊限制:
部分文档作品中含有的国旗、国徽等图片,仅作为作品整体效果示例展示,禁止商用。设计者仅对作品中独创性部分享有著作权。
- 关 键 词:
- ASTME24912008STANDARDGUIDEFOREVALUATINGPERFORMANCECHARACTERISTICSOFPHASEDARRAYULTRASONICTESTINGINSTRUMENTSANDSYSTEMS

链接地址:http://www.mydoc123.com/p-531247.html