1、Designation: E 2033 99 (Reapproved 2006)Standard Practice forComputed Radiology (Photostimulable LuminescenceMethod)1This standard is issued under the fixed designation E 2033; the number immediately following the designation indicates the year oforiginal adoption or, in the case of revision, the ye
2、ar of last revision. A number in parentheses indicates the year of last reapproval. Asuperscript epsilon (e) indicates an editorial change since the last revision or reapproval.1. Scope1.1 This practice2covers application details for computedradiology (CR) examination using a process in which photo-
3、stimulable luminescence is emitted by the penetrating radiationdetector, a storage phosphor imaging plate (SPIP). Because thetechniques involved and the applications for CR examinationare diverse, this practice is not intended to be limiting orrestrictive, but rather to address the general applicati
4、ons of thetechnology and thereby facilitate its use. Refer to Guides E94and E 2007, Terminology E 1316, and Practices E 747 andE 1025, and 21 CFR 1020.40 and 29 CFR 1910.96 foradditional information and guidance.1.2 The general principles discussed in this practice applybroadly to penetrating radiat
5、ion CR systems. However, thisdocument is written specifically for use with X-ray andgamma-ray systems. Other CR systems, such as those employ-ing neutrons, will involve equipment and application detailsunique to such systems.1.3 This standard does not purport to address all of thesafety concerns, if
6、 any, associated with its use. It is theresponsibility of the user of this standard to establish appro-priate safety and health practices and determine the applica-bility of regulatory limitations prior to use. For specific safetystatements, see Section 10 and 21 CFR 1020.40 and 29 CFR1910.96.2. Ref
7、erenced Documents2.1 ASTM Standards:3E94 Guide for Radiographic ExaminationE 747 Practice for Design, Manufacture and MaterialGrouping Classification of Wire Image Quality Indicators(IQI) Used for RadiologyE 1025 Practice for Design, Manufacture, and MaterialGrouping Classification of Hole-Type Imag
8、e Quality Indi-cators (IQI) Used for RadiologyE 1316 Terminology for Nondestructive ExaminationsE 1453 Guide for Storage of Media that Contains Analog orDigital Radioscopic DataE 1475 Guide for Data Fields for Computerized Transfer ofDigital Radiological Examination DataE 1817 Practice for Controlli
9、ng Quality of RadiologicalExamination by Using Representative Quality Indicators(RQIs)E 2007 Guide for Computed Radiology (PhotostimulableLuminescence (PSL) Method)2.2 ASNT Standards:SNT-TC-1A Recommended Practice for Personnel Qualifi-cation and Certification in Nondestructive Testing4ANSI/ASNT-CP-
10、189 Standard for Qualification and Certi-fication of Nondestructive Testing Personnel42.3 Federal Standards:Title 21, CFR 1020.40 Safety Requirements of CabinetX-Ray Systems5Title 29, CFR 1910.96 Ionizing Radiation52.4 AIA Standard:NAS-410 Certification and Qualification of NondestructiveTesting Per
11、sonnel63. Summary of Practice3.1 ACR examination system can be used for a wide varietyof applications. A typical CR examination system consists of aradiation source, a storage phosphor imaging plate detector, aplate reader, an electronic imaging system, a digital image1This test method is under the
12、jurisdiction of ASTM Committee E07 onNondestructive Testing and is the direct responsibility of Subcommittee E07.01 onRadiology (X and Gamma Method).Current edition approved May 1, 2006. Published June 2006. Originallyapproved in 1999. Last previous edition approved in 1999 as E 2033 - 99.2ForASME B
13、oiler and Pressure Code applications, see related Practice SE-2033in Section II of that code.3For referenced ASTM standards, visit the ASTM website, www.astm.org, orcontact ASTM Customer Service at serviceastm.org. For Annual Book of ASTMStandards volume information, refer to the standards Document
14、Summary page onthe ASTM website.4Available from American Society for Nondestructive Testing, 1711 ArlingatePlaza, P.O. Box 28518, Columbus, OH 43228-0518.5Available from Standardization Documents Order Desk, Bldg. 4 Section D, 700Robbins Ave., Philadelphia, PA 19111-5094, Attn: NPODS.6Available from
15、Aerospace IndustriesAssociation ofAmerica, Inc., 1250 Eye St.NW, Washington, D.C. 20005.1Copyright ASTM International, 100 Barr Harbor Drive, PO Box C700, West Conshohocken, PA 19428-2959, United States.processor, a monitor display, a digital image archiving system,and, if desired, equipment for pro
16、ducing hard copy analogimages. This practice establishes the basic parameters for theapplication and control of the CR method.4. Significance and Use4.1 The X-, gamma-ray detector discussed in this practice isa storage phosphor imaging plate, hereafter referred to as SPIP.The SPIP, which is the key
17、component in the CR process,differentiates CR from other radiologic methods. This practiceis written so that it can be specified on the engineeringdrawing, specification, or contract and must be supplementedby a detailed procedure (see Section 6 and Annex A1 andAnnex A2).5. Equipment5.1 System Confi
18、gurationDifferent examination systemsconfigurations are possible, and it is important to understandthe advantages and limitations of each. It is important that theoptimum system be selected for each examination requirementthrough a careful analysis of the benefits and limitations of theavailable sys
19、tem components and the chosen system configu-ration. The provider as well as the user of the examinationservices should be fully aware of the capabilities and limita-tions of the examination system that is proposed for examina-tion of the part. The provider and the user of examinationservices shall
20、agree upon the system configuration to be usedfor each application under consideration and how its perfor-mance is to be evaluated.5.1.1 The minimum system configuration will include anappropriate source of penetrating radiation, a phosphor platedetector, a plate reader, and an electronic imaging sy
21、stem witha CRT display.5.1.2 A more complex system might include a microfocusX-ray system, a digital image processing evaluation system,and an image recording and printing system.6. General Procedure Considerations6.1 The purchaser and supplier shall mutually agree upon awritten procedure using the
22、applicable annex of supplementalrequirements and also consider the following general require-ments.6.1.1 Equipment QualificationsA listing of the systemfeatures that must be qualified to ensure that the system iscapable of performing the desired examination.6.1.2 Source ParameterA listing of all the
23、 radiationsource-related variables that can affect the examination resultsfor the selected system configuration such as: source energy,intensity, focal spot size, range of source to object distances,range of object to image plane distances, and source to imageplane distances.6.1.3 Image Processing P
24、arametersA listing of the imageprocessing variables, if any, necessary to enhance fine detaildetectability in the part and to achieve the required imagequality. These would include, but are not limited to, techniquessuch as noise reduction, contrast enhancement, and spatialfiltering. Great care shou
- 1.请仔细阅读文档,确保文档完整性,对于不预览、不比对内容而直接下载带来的问题本站不予受理。
- 2.下载的文档,不会出现我们的网址水印。
- 3、该文档所得收入(下载+内容+预览)归上传者、原创作者;如果您是本文档原作者,请点此认领!既往收益都归您。
下载文档到电脑,查找使用更方便
5000 积分 0人已下载
下载 | 加入VIP,交流精品资源 |
- 配套讲稿:
如PPT文件的首页显示word图标,表示该PPT已包含配套word讲稿。双击word图标可打开word文档。
- 特殊限制:
部分文档作品中含有的国旗、国徽等图片,仅作为作品整体效果示例展示,禁止商用。设计者仅对作品中独创性部分享有著作权。
- 关 键 词:
- ASTME203319992006STANDARDPRACTICEFORCOMPUTEDRADIOLOGYPHOTOSTIMULABLELUMINESCENCEMETHOD 计算机 辐射 刺激 照明 标准

链接地址:http://www.mydoc123.com/p-530108.html