ASTM E1683-2002(2014)e1 Standard Practice for Testing the Performance of Scanning Raman Spectrometers《扫描拉曼分光仪性能试验的标准实施规程》.pdf
《ASTM E1683-2002(2014)e1 Standard Practice for Testing the Performance of Scanning Raman Spectrometers《扫描拉曼分光仪性能试验的标准实施规程》.pdf》由会员分享,可在线阅读,更多相关《ASTM E1683-2002(2014)e1 Standard Practice for Testing the Performance of Scanning Raman Spectrometers《扫描拉曼分光仪性能试验的标准实施规程》.pdf(6页珍藏版)》请在麦多课文档分享上搜索。
1、Designation: E1683 02 (Reapproved 2014)1Standard Practice forTesting the Performance of Scanning RamanSpectrometers1This standard is issued under the fixed designation E1683; the number immediately following the designation indicates the year oforiginal adoption or, in the case of revision, the year
2、 of last revision. A number in parentheses indicates the year of last reapproval. Asuperscript epsilon () indicates an editorial change since the last revision or reapproval.1NOTEUnits statement was inserted in Section 1.2 editorially in June 2014.1. Scope1.1 This practice covers routine testing of
3、scanning Ramanspectrometer performance and to assist in locating problemswhen performance has degraded. It is also intended as a guidefor obtaining and reporting Raman spectra.1.2 The values stated in SI units are to be regarded asstandard. No other units of measurement are included in thisstandard.
4、1.3 This standard does not purport to address all of thesafety concerns, if any, associated with its use. It is theresponsibility of the user of this standard to establish appro-priate safety and health practices and determine the applica-bility of regulatory limitations prior to use. For specificpr
5、ecautions, see 7.2.1.1.4 Because of the significant dangers associated with theuse of lasers, ANSI Z136.1 should be followed in conjunctionwith this practice.2. Referenced Documents2.1 ASTM Standards:2E131 Terminology Relating to Molecular SpectroscopyE1840 Guide for Raman Shift Standards for Spectr
6、ometerCalibration2.2 ANSI Standard:3Z136.1 Safe Use of Lasers3. Terminology3.1 Terminology used in this practice conforms to thedefinitions in Terminology E131.4. Significance and Use4.1 A scanning Raman spectrometer should be checkedregularly to determine if its condition is adequate for routinemea
7、surements or if it has changed. This practice is designed tofacilitate that determination and, if performance isunsatisfactory, to identify the part of the system that needsattention. These tests apply for single-, double-, or triplemono-chromator scanning Raman instruments commercially avail-able.
8、They do not apply for multichannel or Fourier transforminstruments, or for gated integrator systems requiring a pulsedlaser source. Use of this practice is intended only for trainedoptical spectroscopists and should be used in conjunction withstandard texts.5. Apparatus5.1 LaserAmonochromatic, conti
9、nuous laser source, suchas an argon, krypton, or helium-neon laser, is normally used forRaman measurements. The laser intensity should be measuredat the sample with a power meter because optical componentsbetween the laser and sample reduce laser intensity. A filteringdevice should also be used to r
10、emove non-lasting plasmaemission lines from the laser beam before they reach thesample. Plasma lines can seriously interfere with Ramanmeasurements. Filtering devices include dispersive monochro-mators and interference filters.5.2 Sampling OpticsCommercial instruments can be pur-chased with sampling
11、 optics to focus the laser beam onto asample and to image the Raman scattering onto the monochro-mator entrance slit. Sample chamber adjustments are used tocenter the sample properly and align the Raman scattered light.A schematic view of a conventional 90 Raman scatteringgeometry is shown in Fig. 1
12、. The laser beam propagates at aright angle to the direction in which scattered light is collected.It is focused on the sample at the same position as themonochromator entrance slit image. Other geometries such as180 backscattering are also used. With singlemonochromators, a filter is normally place
13、d in the opticalcollection path to block light at the laser frequency fromentering the monochromator.1This practice is under the jurisdiction of ASTM Committee E13 on MolecularSpectroscopy and Separation Science and is the direct responsibility of Subcom-mittee E13.08 on Raman Spectroscopy.Current e
14、dition approved May 1, 2014. Published June 2014. Originallyapproved in 1995. Last previous edition approved in 2007 as E1683 02(2007).DOI: 10.1520/E1683-02R14E01.2For referenced ASTM standards, visit the ASTM website, www.astm.org, orcontact ASTM Customer Service at serviceastm.org. For Annual Book
15、 of ASTMStandards volume information, refer to the standards Document Summary page onthe ASTM website.3Available from American National Standards Institute (ANSI), 25 W. 43rd St.,4th Floor, New York, NY 10036, http:/www.ansi.org.Copyright ASTM International, 100 Barr Harbor Drive, PO Box C700, West
16、Conshohocken, PA 19428-2959. United States15.3 PolarizationFor routine measurements the polariza-tion of the laser at the sample is oriented normal to the planeof the page in Fig. 1. However, measurements using differentpolarizations are sometimes used to determine vibrationalsymmetries as part of m
17、olecular structure determinations. Avariety of optical configurations can be used to make polariza-tion measurements; a detailed discussion of these is beyond thescope of this practice. Briefly, for polarization simple measure-ments of randomly-oriented samples (most of the clearliquids), an analyzi
18、ng element such as a polaroid filter oranalyzing prism is added to the optical system and Ramanspectra are collected for light scattered in (1) the same directionas the source (parallel), (2) perpendicular to the source.Depolarization ratios are calculated using Raman band inten-sities from the two
19、spectra as follows:Depolarization ratio 5Intensity parallelIntensity perpendicular(1)5.3.1 A polarization scrambler is shown in Fig. 1. Thiselement is used to avoid making corrections for polarization-dependent grating effects. The scrambler is also frequentlyused during routine measurements and sho
20、uld be placedbetween the sample and entrance slit, close to the collectionlens. A polaroid filter placed between the scrambler andcollection lens provides a simple polarization measurementsystem.5.4 MonochromatorA scanning monochromator used forRaman spectroscopy will exhibit high performance requir
21、e-ments. Double and triple monochromators have particularlystringent performance standards. During the original instru-ment design, features are usually introduced to minimizeoptical aberrations. However, proper maintenance of opticalalignment is essential. A focused image on the entrance slitshould
22、 be optically transferred to and matched with the otherslits. If the monochromator is not functioning properly contactthe manufacturer for assistance.5.5 Photomultiplier TubeA photomultiplier can be usedfor detecting Raman scattered radiation. A tube with goodresponse characteristics at and above th
23、e laser wavelengthshould be selected. Dark signal can be reduced with thermo-electric cooling for improved detection of weak signals.Current and voltage amplification or photon counting arecommercially available options with photomultiplier tubes.6. Guidelines for Obtaining and Reporting RamanSpectr
24、a6.1 Alignment of Optical ElementsRefer to the manufac-turer for detailed sample chamber alignment instructions. Uponinstallation, each optical component should be aligned indi-vidually. For optimal alignment the sample image should becentered on the entrance slit of the monochromator (oftenviewed t
- 1.请仔细阅读文档,确保文档完整性,对于不预览、不比对内容而直接下载带来的问题本站不予受理。
- 2.下载的文档,不会出现我们的网址水印。
- 3、该文档所得收入(下载+内容+预览)归上传者、原创作者;如果您是本文档原作者,请点此认领!既往收益都归您。
下载文档到电脑,查找使用更方便
5000 积分 0人已下载
下载 | 加入VIP,交流精品资源 |
- 配套讲稿:
如PPT文件的首页显示word图标,表示该PPT已包含配套word讲稿。双击word图标可打开word文档。
- 特殊限制:
部分文档作品中含有的国旗、国徽等图片,仅作为作品整体效果示例展示,禁止商用。设计者仅对作品中独创性部分享有著作权。
- 关 键 词:
- ASTME168320022014E1STANDARDPRACTICEFORTESTINGTHEPERFORMANCEOFSCANNINGRAMANSPECTROMETERS 扫描 分光仪 性能 试验

链接地址:http://www.mydoc123.com/p-529228.html