ASTM E1654-1994(2013) Standard Guide for Measuring Ionizing Radiation-Induced Spectral Changes in Optical Fibers and Cables for Use in Remote Raman FiberOptic Spectroscopy《远程拉曼光纤分光.pdf
《ASTM E1654-1994(2013) Standard Guide for Measuring Ionizing Radiation-Induced Spectral Changes in Optical Fibers and Cables for Use in Remote Raman FiberOptic Spectroscopy《远程拉曼光纤分光.pdf》由会员分享,可在线阅读,更多相关《ASTM E1654-1994(2013) Standard Guide for Measuring Ionizing Radiation-Induced Spectral Changes in Optical Fibers and Cables for Use in Remote Raman FiberOptic Spectroscopy《远程拉曼光纤分光.pdf(5页珍藏版)》请在麦多课文档分享上搜索。
1、Designation: E1654 94 (Reapproved 2013)Standard Guide forMeasuring Ionizing Radiation-Induced Spectral Changes inOptical Fibers and Cables for Use in Remote RamanFiberOptic Spectroscopy1This standard is issued under the fixed designation E1654; the number immediately following the designation indica
2、tes the year oforiginal adoption or, in the case of revision, the year of last revision. A number in parentheses indicates the year of last reapproval. Asuperscript epsilon () indicates an editorial change since the last revision or reapproval.1. Scope1.1 This guide covers the method for measuring t
3、he realtime, in situ radiation-induced alterations to the Raman spectralsignal transmitted by a multimode, step index, silica opticalfiber. This guide specifically addresses steady-state ionizingradiation (that is, alpha, beta, gamma, protons, etc.) withappropriate changes in dosimetry, and shieldin
4、g considerations,depending upon the irradiation source.1.2 The test procedure given in this guide is not intended totest the other optical and non-optical components of an opticalfiber-based Raman sensor system, but may be modified to testother components in a continuous irradiation environment.1.3
5、The values stated in SI units are to be regarded asstandard. No other units of measurement are included in thisstandard.1.4 This standard does not purport to address all of thesafety concerns, if any, associated with its use. It is theresponsibility of the user of this standard to establish appro-pr
6、iate safety and health practices and determine the applica-bility of regulatory limitations prior to use.2. Referenced Documents2.1 ASTM Standards:2E1614 Guide for Procedure for Measuring IonizingRadiation-Induced Attenuation in Silica-Based OpticalFibers and Cables for Use in Remote Fiber-Optic Spe
7、c-troscopy andBroadband Systems2.2 EIA Standards:32.2.1 Test or inspection requirements include the followingreferences:EIA-455-57 Optical Fiber End Preparation and ExaminationEIA-455-64 Procedure for Measuring Radiation-InducedAttenuation in Optical Fibers and Cables2.3 Military Standards:4MIL-STD-
8、2196-(SH) Glossary of Fiber Optic Terms3. Terminology3.1 DefinitionsRefer to the following documents for thedefinition of terms used in this guide: MIL-STD-2196-(SH)and Guide E1614.4. Significance and Use4.1 Ionizing environments will affect the performance ofoptical fibers/cables being used to tran
9、smit spectroscopicinformation from a remote location. Determination of the typeand magnitude of the spectral variations or interferencesproduced by the ionizing radiation in the fiber, or both, isnecessary for evaluating the performance of an optical fibersensor system.4.2 The results of the test ca
10、n be utilized as a selectioncriteria for optical fibers used in optical fiber Raman spectro-scopic sensor systems.NOTE 1The attenuation of optical fibers generally increases whenthey are exposed to ionizing radiation.This is due primarily to the trappingof radiolytic electrons and holes at defect si
11、tes in the optical materials, thatis, the formation of color centers. The depopulation of these color centersby thermal or optical (photobleaching) processes, or both, causesrecovery, usually resulting in a decrease in radiationinduced attenuation.Recovery of the attenuation after irradiation depend
12、s on many variables,including the temperature of the test sample, the composition of thesample, the spectrum and type of radiation employed, the total doseapplied to the test sample, the light level used to measure the attenuation,and the operating spectrum. Under some continuous conditions, recover
13、yis never complete.5. Apparatus5.1 The test schematic is shown in Fig. 1. The following listidentifies the equipment necessary to accomplish this testprocedure.1This guide is under the jurisdiction of ASTM Committee E13 on MolecularSpectroscopy and Separation Science and is the direct responsibility
14、 of Subcom-mittee E13.09 on Fiber Optics, Waveguides, and Optical Sensors.Current edition approved Jan. 1, 2013. Published January 2013. Originallyapproved in 1994. Last previous version approved in 2004 as E1654 94 (2004).DOI: 10.1520/E1654-94R13.2For referenced ASTM standards, visit the ASTM websi
15、te, www.astm.org, orcontact ASTM Customer Service at serviceastm.org. For Annual Book of ASTMStandards volume information, refer to the standards Document Summary page onthe ASTM website.3Available from Electronic Industries Alliance (EIA), 2500 Wilson Blvd.,Arlington, VA 22201, http:/www.ecaus.org/
16、eia.4Available from Standardization Documents Order Desk, DODSSP, Bldg. 4,Section D, 700 Robbins Ave., Philadelphia, PA 19111-5098, http:/dodssp.daps.dla.mil.Copyright ASTM International, 100 Barr Harbor Drive, PO Box C700, West Conshohocken, PA 19428-2959. United States15.2 Light SourceA laser sour
17、ce shall be used for theRaman analysis, and the wavelength must be chosen so that thefluorescent signals from the optical components (especially thespectral activator sample and optical fibers) are minimized, andso that the wavelength corresponds to the spectral sensitivity ofthe detection scheme. T
18、ypically, the wavelength range ex-ploited spans from 0.4 to 1.06 m. The laser source must havesufficient power to obtain the desired minimum signal-to-noiseratio (S/N) (see 10.3).5.3 Focusing/Collection OpticsA number of optical ele-ments are needed for the launch and collection of lightradiation in
19、to and from the optical fibers (interfacing, sampleand reference), and other instrumentation (light source,spectrograph, detector). The minimal requirement for theseelements shall be that the numerical aperture of the compo-nents are matched for efficient coupling. Optics may also benecessary to enh
20、ance the interaction of the input light with thespectral activator.FIG. 1 Test ConfigurationE1654 94 (2013)25.4 Interfacing Optical FiberThe primary requirement ofthe interfacing optical fiber is to provide the minimum powerto the activator sample at the proper wavelength(s). The fiberlength may be
21、adjusted so that the power requirements are met.5.5 Light Radiation FilteringIt is important that all neigh-boring laser lines are removed from the source beam prior tointeraction with the spectral activator. This can be accom-plished before or after the interfacing optical fiber. Placementof the fi
22、lter before the interfacing fiber will eliminate theneighboring laser lines, but any fluorescence and Ramanscattering due to the fiber or associated optics will be allowedto interact with the sample. Placement of the laser pass filterafter the interfacing fiber is preferable because it will eliminat
23、eany signals created within the fiber. If it is necessary to placethe filter before the interfacing fiber, then the fiber should bekept as short as possible (several metres).5.6 Spectral Activator SampleThe spectral activator usedmust demonstrate a strong, well-characterized Raman spectralsignal. Th
24、e sample may be either liquid, gas, or solid, depend-ing on the requirements of the optical fiber arrangement. It isrecommended that a liquid be used, since the Raman scatteringin the proposed configuration will launch similarly into thesample and reference fibers. Standard recommended samplesare: a
- 1.请仔细阅读文档,确保文档完整性,对于不预览、不比对内容而直接下载带来的问题本站不予受理。
- 2.下载的文档,不会出现我们的网址水印。
- 3、该文档所得收入(下载+内容+预览)归上传者、原创作者;如果您是本文档原作者,请点此认领!既往收益都归您。
下载文档到电脑,查找使用更方便
5000 积分 0人已下载
下载 | 加入VIP,交流精品资源 |
- 配套讲稿:
如PPT文件的首页显示word图标,表示该PPT已包含配套word讲稿。双击word图标可打开word文档。
- 特殊限制:
部分文档作品中含有的国旗、国徽等图片,仅作为作品整体效果示例展示,禁止商用。设计者仅对作品中独创性部分享有著作权。
- 关 键 词:
- ASTME165419942013STANDARDGUIDEFORMEASURINGIONIZINGRADIATIONINDUCEDSPECTRALCHANGESINOPTICALFIBERSANDCABLESFORUSEINREMOTERAMANFIBEROPTICSPECTROSCOPY

链接地址:http://www.mydoc123.com/p-529148.html