ASTM E1588-2016 Standard Guide for Gunshot Residue Analysis by Scanning Electron Microscopy Energy Dispersive X-Ray Spectrometry《采用扫描电子显微术 能量散射X射线光谱法进行射击残留物分析的标准指南》.pdf
《ASTM E1588-2016 Standard Guide for Gunshot Residue Analysis by Scanning Electron Microscopy Energy Dispersive X-Ray Spectrometry《采用扫描电子显微术 能量散射X射线光谱法进行射击残留物分析的标准指南》.pdf》由会员分享,可在线阅读,更多相关《ASTM E1588-2016 Standard Guide for Gunshot Residue Analysis by Scanning Electron Microscopy Energy Dispersive X-Ray Spectrometry《采用扫描电子显微术 能量散射X射线光谱法进行射击残留物分析的标准指南》.pdf(5页珍藏版)》请在麦多课文档分享上搜索。
1、Designation: E1588 16Standard Guide forGunshot Residue Analysis by Scanning ElectronMicroscopy/Energy Dispersive X-Ray Spectrometry1This standard is issued under the fixed designation E1588; the number immediately following the designation indicates the year oforiginal adoption or, in the case of re
2、vision, the year of last revision. A number in parentheses indicates the year of last reapproval. Asuperscript epsilon () indicates an editorial change since the last revision or reapproval.1. Scope1.1 This guide covers the analysis of gunshot residue (GSR)by scanning electron microscopy/energy-disp
3、ersive X-rayspectrometry (SEM/EDS) by manual and automated methods.The analysis may be performed manually, with the operatormanipulating the microscope controls and the EDS systemsoftware, or in an automated fashion, where some amount ofthe analysis is controlled by pre-set software functions.1.2 Si
4、nce software and hardware formats vary among com-mercial systems, guidelines will be offered in the most generalterms possible. For proper terminology and operation, consultthe SEM/EDS system manuals for each system.1.3 The values stated in SI units are to be regarded asstandard. No other units of m
5、easurement are included in thisstandard.1.4 This standard does not purport to address all of thesafety concerns, if any, associated with its use. It is theresponsibility of the user of this standard to establish appro-priate safety and health practices and determine the applica-bility of regulatory
6、limitations prior to use.2. Referenced Documents2.1 ASTM Standards:2E1492 Practice for Receiving, Documenting, Storing, andRetrieving Evidence in a Forensic Science Laboratory3. Terminology3.1 Definitions of Terms Specific to This Standard:3.1.1 major, adjelement whose main peak height isgreater tha
7、n13 of the peak height of the strongest peak in thespectrum. Wallace 1984 (15)33.1.2 minor, adjelement whose main peak height is be-tween110 and13 of the peak height of the strongest peak in thespectrum. Wallace 1984 (15)3.1.3 trace, adjelement whose main peak height is lessthan110 of the peak heigh
8、t of the strongest peak in thespectrum. Wallace 1984 (15)4. Summary of Practice4.1 From the total population of particles collected, thosethat are detected by SEM to be within the limits of certainparameters (for example, atomic number, size, or shape) areanalyzed by EDS (1-3). Typically, particles
9、composed of highmean atomic number elements are detected by their SEMbackscattered electron signals and an EDS spectrum is ob-tained from each. The EDS spectrum is evaluated for constitu-ent elements that may identify the particle as being consistentwith or characteristic of GSR, or both.5. Signific
10、ance and Use5.1 This document will be of use to forensic laboratorypersonnel who are involved in the analysis of GSR samples bySEM/EDS (4).5.2 SEM/EDS analysis of GSR is a non-destructive methodthat provides (5, 6) both morphological information and theelemental profiles of individual particles.5.3
11、Particle analysis contrasts with bulk sample methods,such as atomic absorption spectrophotometry (AAS) (7), neu-tron activation analysis (NAA) (8), inductively coupled plasmaatomic emission spectrometry (ICP-AES), and inductivelycoupled plasma mass spectrometry (ICP-MS), where thesampled material is
12、 dissolved or extracted prior to the deter-mination of total element concentrations, thereby sacrificingmorphological information and individual particle identifica-tion.5.4 X-ray fluorescence spectrometry (XRF) is a techniquethat has been used to map the placement and distribution ofGSR particles s
13、urrounding bullet holes in order to establishshooting distances (9). Unlike the solution-based bulk methodsof analysis, XRF is non-destructive; however, XRF still doesnot provide morphological information and is incapable ofindividual GSR particle identification.1This guide is under the jurisdiction
14、 of ASTM Committee E30 on ForensicSciences and is the direct responsibility of Subcommittee E30.01 on Criminalistics.Current edition approved March 1, 2016. Published March 20116. Originallyapproved in 1994. Last previous version approved in 2010 as E1588 101. DOI:10.1520/E1588-16.2For referenced AS
15、TM standards, visit the ASTM website, www.astm.org, orcontact ASTM Customer Service at serviceastm.org. For Annual Book of ASTMStandards volume information, refer to the standards Document Summary page onthe ASTM website.3The boldface numbers in parentheses refer to a list of references at the end o
16、fthis standard.Copyright ASTM International, 100 Barr Harbor Drive, PO Box C700, West Conshohocken, PA 19428-2959. United States16. Sample Preparation6.1 Once the evidence seal is broken, care should be takenso that no object touches the surface of the adhesive SEM/EDSsample collection stub and that
17、 the stub is not left uncoveredany longer than is reasonable for transfer, mounting, orlabeling.6.2 Label the sample collection stub in such a manner that itis distinguishable from other sample collection stubs withoutcompromising the sample; for example, label the bottom orside of the stub.6.3 If a
18、 non-conductive adhesive was used in the samplecollection stub, the sample will need to be coated to increase itselectrical conductivity, unless an environmental SEM orvariable-pressure/low-vacuum SEM is used for the analysis.Carbon is a common choice of coating material, since it willnot interfere
19、with X-ray lines of interest. For high-vacuumSEM, coat the sample sufficiently to eliminate charging of thesample.6.4 Observe the appropriate procedures for handling anddocumentation of all submitted samples as described in Prac-tice E1492.7. Sample Area7.1 Sample collection stubs for SEMs typically
20、 come in oneof two diameters: 12.7 mm or 25.4 mm, which yield surfaceareas of 126.7 mm2and 506.7 mm2respectively.7.2 Manual analysis of the total surface area of the stub isprohibitively time-consuming. Because the particles are col-lected onto an adhesive surface in a random manner and theparticles
21、 do not tend to cluster, it is reasonable to analyze aportion of the stub surface by employing an appropriatesampling and analytical protocol (6, 10).7.3 Automated SEM/EDS analysis can enable data collec-tion from nearly the entire surface area of the sample collectionstub. Due to the disparity betw
22、een the shape of the samplecollection stub (round) and the SEM field of view search area(square or rectangular), analysis of 100 % of the samplecollection area may not be possible in some systems.7.3.1 Analysis of the maximum allowable surface area ofthe sample is recommended, however, many automate
23、d sys-tems can be programmed to terminate the analysis of a stub orseries of stubs once a pre-established number of particles withspecified classification(s) have been detected. The decision asto how many particles satisfy the requirements of a particularcase is a matter for the analyst to decide bu
24、t should be subjectto guidelines set out in the laboratorys standard operatingprocedures.8. Instrument Requirements and Operation8.1 General:8.1.1 Most commercial-grade SEM/EDS systems should beadequate for GSR analysis.8.1.2 Automated data collection of GSR involves someportion of the data collecti
- 1.请仔细阅读文档,确保文档完整性,对于不预览、不比对内容而直接下载带来的问题本站不予受理。
- 2.下载的文档,不会出现我们的网址水印。
- 3、该文档所得收入(下载+内容+预览)归上传者、原创作者;如果您是本文档原作者,请点此认领!既往收益都归您。
下载文档到电脑,查找使用更方便
5000 积分 0人已下载
下载 | 加入VIP,交流精品资源 |
- 配套讲稿:
如PPT文件的首页显示word图标,表示该PPT已包含配套word讲稿。双击word图标可打开word文档。
- 特殊限制:
部分文档作品中含有的国旗、国徽等图片,仅作为作品整体效果示例展示,禁止商用。设计者仅对作品中独创性部分享有著作权。
- 关 键 词:
- ASTME15882016STANDARDGUIDEFORGUNSHOTRESIDUEANALYSISBYSCANNINGELECTRONMICROSCOPYENERGYDISPERSIVEXRAYSPECTROMETRY

链接地址:http://www.mydoc123.com/p-528996.html