1、Designation: E4 10American Association StateHighway and Transportation Officials StandardsAASHTO No: T67Standard Practices forForce Verification of Testing Machines1This standard is issued under the fixed designation E4; the number immediately following the designation indicates the year of original
2、adoption or, in the case of revision, the year of last revision. A number in parentheses indicates the year of last reapproval. A superscriptepsilon () indicates an editorial change since the last revision or reapproval.This standard has been approved for use by agencies of the Department of Defense
3、.1. Scope1.1 These practices cover procedures for the force verifica-tion, by means of standard calibration devices, of tension orcompression, or both, static or quasi-static testing machines(which may, or may not, have force-indicating systems). Thesepractices are not intended to be complete purcha
4、se specifica-tions for testing machines. Testing machines may be verified byone of the three following methods or combination thereof:1.1.1 Use of standard weights,1.1.2 Use of equal-arm balances and standard weights, or1.1.3 Use of elastic calibration devices.NOTE 1These practices do not cover the
5、verification of all types oftesting machines designed to measure forces, for example, the constant-rate-of-loading type which operates on the inclined-plane principle. Thistype of machine may be verified as directed in the applicable appendix ofSpecification D76.1.2 The procedures of 1.1.1-1.1.3 app
6、ly to the verification ofthe force-indicating systems associated with the testing ma-chine, such as a scale, dial, marked or unmarked recorder chart,digital display, etc. In all cases the buyer/owner/user mustdesignate the force-indicating system(s) to be verified andincluded in the report.1.3 The v
7、alues stated in inch-pound units are to be regardedas standard. The values given in parentheses are mathematicalconversions to SI units that are provided for information onlyand are not considered standard.1.3.1 Since conversion factors are not required in thispractice, either inch-pound units, SI u
8、nits, or metric values canbe used as the standard.1.4 Forces indicated on displays/printouts of testing ma-chine data systemsbe they instantaneous, delayed, stored, orretransmittedwhich are verified with provisions of 1.1.1,1.1.2,or1.1.3, and are within the 61 % accuracy requirement,comply with Prac
9、tices E4.1.5 This standard does not purport to address all of thesafety concerns, if any, associated with its use. It is theresponsibility of the user of this standard to establish appro-priate safety and health practices and determine the applica-bility of regulatory limitations prior to use.2. Ref
10、erenced Documents2.1 ASTM Standards:2D76 Specification for Tensile Testing Machines for TextilesE74 Practice of Calibration of Force-Measuring Instru-ments for Verifying the Force Indication of Testing Ma-chinesE467 Practice for Verification of Constant Amplitude Dy-namic Forces in an Axial Fatigue
11、Testing System3. Terminology3.1 Definitions:3.1.1 testing machine (force-measuring type)a mechani-cal device for applying a force to a specimen.3.1.1.1 portable testing machine (force-measuring type)adevice specifically designed to be moved from place to placeand for applying a force (load) to a spe
12、cimen.3.1.2 forcein the case of testing machines, a force mea-sured in units such as pound-force, newton, or kilogram-force.3.1.2.1 DiscussionThe pound-force is that force whichacting on a 1-lb mass will give to it an acceleration of 32.1740ft/s2(9.80665 m/s2). The newton is that force which acting
13、ona 1-kg mass will give to it an acceleration of 1 m/s2.3.1.3 accuracythe specified permissible variation fromthe reference value.3.1.3.1 DiscussionAtesting machine is said to be accurateif the indicated force is within the specified permissiblevariation from the actual force. In these methods the w
14、ord“accurate” applied to a testing machine is used withoutnumerical values, for example, “An accurate testing machinewas used for the investigation.” The accuracy of a testingmachine should not be confused with sensitivity. For example,a testing machine might be very sensitive; that is, it mightindi
15、cate quickly and definitely small changes in force, butnevertheless, be very inaccurate. On the other hand, theaccuracy of the results is in general limited by the sensitivity.1These practices are under the jurisdiction of ASTM Committee E28 onMechanical Testing and is the direct responsibility of S
16、ubcommittee E28.01 onCalibration of Mechanical Testing Machines and Apparatus.Current edition approved June 1, 2010. Published July 2010. Originally approvedin 1923. Last previous edition approved in 2009 as E4 09a. DOI: 10.1520/E0004-10.2For referenced ASTM standards, visit the ASTM website, www.as
17、tm.org, orcontact ASTM Customer Service at serviceastm.org. For Annual Book of ASTMStandards volume information, refer to the standards Document Summary page onthe ASTM website.1Copyright ASTM International, 100 Barr Harbor Drive, PO Box C700, West Conshohocken, PA 19428-2959, United States.3.1.4 er
18、ror (or the deviation from the correct value)in thecase of a testing machine, the difference obtained by subtract-ing the force indicated by the calibration device from the forceindicated by the testing machine.3.1.4.1 DiscussionThe word “error” shall be used withnumerical values, for example, “At a
19、 force of 30 000 lbf (133kN), the error of the testing machine was + 15 lbf (67 N).”3.1.5 percent errorin the case of a testing machine, theratio, expressed as a percent, of the error to the correct value ofthe applied force.3.1.5.1 DiscussionThe test force, as indicated by thetesting machine, and t
20、he applied force, as computed from thereadings of the verification device, shall be recorded at eachtest point. The error, E, and the percent error, Ep, shall becalculated from these data as follows:E 5 A 2 B (1)Ep5 A 2 B!/B 3 100where:A = force indicated by machine being verified, lbf (or N),andB =
21、 correct value of the applied force, lbf (or N), asdetermined by the calibration device.3.1.6 correctionin the case of a testing machine, thedifference obtained by subtracting the indicated force from thecorrect value of the applied force.3.1.7 permissible variation (or tolerance)in the case oftesti
22、ng machines, the maximum allowable error in the value ofthe quantity indicated.3.1.7.1 DiscussionIt is convenient to express permissiblevariation in terms of percentage of error. The numerical valueof the permissible variation for a testing machine is so statedhereafter in these practices.3.1.8 capa
23、city rangein the case of testing machines, therange of forces for which it is designed. Some testing machineshave more than one capacity range, that is, multiple ranges.3.1.9 verified range of forcesin the case of testing ma-chines, the range of indicated forces for which the testingmachine gives re
24、sults within the permissible variations speci-fied.3.1.9.1 calibration, nin the case of force testing machines,the process of comparing the force indication of the machineunder test to that of a standard, making adjustments as neededto meet error requirements.3.1.9.2 verification, nin the case of fo
- 1.请仔细阅读文档,确保文档完整性,对于不预览、不比对内容而直接下载带来的问题本站不予受理。
- 2.下载的文档,不会出现我们的网址水印。
- 3、该文档所得收入(下载+内容+预览)归上传者、原创作者;如果您是本文档原作者,请点此认领!既往收益都归您。
下载文档到电脑,查找使用更方便
5000 积分 0人已下载
下载 | 加入VIP,交流精品资源 |
- 配套讲稿:
如PPT文件的首页显示word图标,表示该PPT已包含配套word讲稿。双击word图标可打开word文档。
- 特殊限制:
部分文档作品中含有的国旗、国徽等图片,仅作为作品整体效果示例展示,禁止商用。设计者仅对作品中独创性部分享有著作权。
- 关 键 词:
- ASTME420100000STANDARDPRACTICESFORFORCEVERIFICATIONOFTESTINGMACHINES 试验 验证 标准 实施 规程 PDF

链接地址:http://www.mydoc123.com/p-527563.html