ASTM E317-2006a Standard Practice for Evaluating Performance Characteristics of Ultrasonic Pulse-Echo Testing Instruments and Systems without the Use of Electronic Measurement Inst.pdf
《ASTM E317-2006a Standard Practice for Evaluating Performance Characteristics of Ultrasonic Pulse-Echo Testing Instruments and Systems without the Use of Electronic Measurement Inst.pdf》由会员分享,可在线阅读,更多相关《ASTM E317-2006a Standard Practice for Evaluating Performance Characteristics of Ultrasonic Pulse-Echo Testing Instruments and Systems without the Use of Electronic Measurement Inst.pdf(13页珍藏版)》请在麦多课文档分享上搜索。
1、Designation: E 317 06aStandard Practice forEvaluating Performance Characteristics of Ultrasonic Pulse-Echo Testing Instruments and Systems without the Use ofElectronic Measurement Instruments1This standard is issued under the fixed designation E 317; the number immediately following the designation
2、indicates the year oforiginal adoption or, in the case of revision, the year of last revision. A number in parentheses indicates the year of last reapproval. Asuperscript epsilon (e) indicates an editorial change since the last revision or reapproval.This standard has been approved for use by agenci
3、es of the Department of Defense.1. Scope1.1 This practice describes procedures for evaluating thefollowing performance characteristics of ultrasonic pulse-echoexamination instruments and systems: Horizontal Limit andLinearity; Vertical Limit and Linearity; Resolution - EntrySurface and Far Surface;
4、Sensitivity and Noise; Accuracy ofCalibrated Gain Controls. Evaluation of these characteristics isintended to be used for comparing instruments and systems or,by periodic repetition, for detecting long-term changes in thecharacteristics of a given instrument or system that may beindicative of impend
5、ing failure, and which, if beyond certainlimits, will require corrective maintenance. Instrument charac-teristics measured in accordance with this practice are ex-pressed in terms that relate to their potential usefulness forultrasonic examinations. Instrument characteristics expressedin purely elec
6、tronic terms may be measured as described inE 1324.1.2 Ultrasonic examination systems using pulsed-wavetrains and A-scan presentation (rf or video) may be evaluated.1.3 The procedures are applicable to shop or field condi-tions; additional electronic measurement instrumentation is notrequired.1.4 Th
7、is practice establishes no performance limits forexamination systems; if such acceptance criteria are required,these must be specified by the using parties. Where acceptancecriteria are implied herein they are for example only and aresubject to more or less restrictive limits imposed by customersand
8、 end users controlling documents.1.5 The specific parameters to be evaluated, conditions andfrequency of test, and report data required, must also bedetermined by the user.1.6 This practice may be used for the evaluation of acomplete examination system, including search unit, instru-ment, interconne
9、ctions, fixtures and connected alarm andauxiliary devices, primarily in cases where such a system isused repetitively without change or substitution. This practiceis not intended to be used as a substitute for calibration orstandardization of an instrument or system to inspect any givenmaterial. The
10、re are limitations to the use of standard referenceblocks for that purpose.21.7 Required test apparatus includes selected test blocks anda precision external attenuator (where specified) in addition tothe instrument or system to be evaluated.1.8 Precautions relating to the applicability of the proce
11、-dures and interpretation of the results are included.1.9 Alternate procedures, such as examples described in thisdocument, or others, may only be used with customer approval.1.10 This standard does not purport to address all of thesafety concerns, if any, associated with its use. It is theresponsib
12、ility of the user of this standard to establish appro-priate safety and health practices and determine the applica-bility of regulatory limitations prior to use.2. Referenced Documents2.1 ASTM Standards:3E114 Practice for Ultrasonic Pulse-Echo Straight-BeamExamination by the Contact MethodE 127 Prac
13、tice for Fabricating and Checking AluminumAlloy Ultrasonic Standard Reference BlocksE 214 Practice for Immersed Ultrasonic Testing by theReflection Method Using Pulsed Longitudinal WavesE 428 Practice for Fabrication and Control of Metal, Otherthan Aluminum Reference, Blocks Used in UltrasonicExamin
14、ationE 1316 Terminology for Nondestructive ExaminationsE 1324 Guide for Measuring Some Electronic Characteris-tics of Ultrasonic Examination Instruments1This practice is under the jurisdiction of ASTM Committee E07 on Nonde-structive Testing and is the direct responsibility of Subcommittee E07.06 on
15、Ultrasonic Method.Current edition approved Dec. 1, 2006. Published January 2007. Originallyapproved in 1967. Last previous edition approved in 2006 as E 317 - 06.2Beck, K. H., “Limitations to the Use of Reference Blocks for Periodic andPreinspection Calibration of Ultrasonic Inspection Instruments a
16、nd Systems,”Materials Evaluation, Vol 57, No. 3, March 1999.3For referenced ASTM standards, visit the ASTM website, www.astm.org, orcontact ASTM Customer Service at serviceastm.org. For Annual Book of ASTMStandards volume information, refer to the standards Document Summary page onthe ASTM website.1
17、Copyright ASTM International, 100 Barr Harbor Drive, PO Box C700, West Conshohocken, PA 19428-2959, United States.2.2 Other Standard:IEEE Std 100, IEEE Standard Dictionary of Electrical andElectronic Terms43. Terminology3.1 DefinitionsFor definitions of terms used in this prac-tice see Terminology E
18、 1316. Other relevant definitions may befound in IEEE Standard 100.4. Summary of Practice4.1 An examination system to be evaluated comprises anultrasonic pulse-echo instrument, search unit, interconnectingcables, and couplant; for immersion examination systemssuitable fixturing is required.4.2 When
19、checking an entire system to be used for a givenexamination, test conditions are selected that are consistentwith the intended end-use as determined by the user.4.3 The ultrasonic response from appropriate test blocks isobtained, and presented in numerical or graphical form.5. Significance and Use5.
20、1 This practice describes procedures applicable to bothshop and field conditions. More comprehensive or precisemeasurements of the characteristics of complete systems andtheir components will generally require laboratory techniquesand electronic equipment such as oscilloscopes and signalgenerators.
21、Substitution of these methods is not precludedwhere appropriate; however, their usage is not within the scopeof this practice.5.2 This document does not establish system acceptancelimits, nor is it intended as a comprehensive equipmentspecification.5.3 While several important characteristics are inc
22、luded,others of possible significance in some applications are notcovered.5.4 Since the parameters to be evaluated and the applicabletest conditions must be specified, this practice should beprescribed only by those familiar with ultrasonic NDT tech-nology and the required tests should be performed
23、either bysuch a qualified person or under his supervision.5.5 Implementation may require more detailed proceduralinstructions in the format of the using facility.5.6 In the case of evaluation of a complete system selectionof the specific tests to be made should be done cautiously; if therelated para
24、meters are not critical in the intended application,then their inclusion may be unjustified. For example, verticallinearity may be irrelevant for a go/no-go test with a flaw gatealarm, while horizontal linearity might be required only foraccurate flaw-depth or thickness measurement from the displays
- 1.请仔细阅读文档,确保文档完整性,对于不预览、不比对内容而直接下载带来的问题本站不予受理。
- 2.下载的文档,不会出现我们的网址水印。
- 3、该文档所得收入(下载+内容+预览)归上传者、原创作者;如果您是本文档原作者,请点此认领!既往收益都归您。
下载文档到电脑,查找使用更方便
5000 积分 0人已下载
下载 | 加入VIP,交流精品资源 |
- 配套讲稿:
如PPT文件的首页显示word图标,表示该PPT已包含配套word讲稿。双击word图标可打开word文档。
- 特殊限制:
部分文档作品中含有的国旗、国徽等图片,仅作为作品整体效果示例展示,禁止商用。设计者仅对作品中独创性部分享有著作权。
- 关 键 词:
- ASTME3172006ASTANDARDPRACTICEFOREVALUATINGPERFORMANCECHARACTERISTICSOFULTRASONICPULSEECHOTESTINGINSTRUMENTSANDSYSTEMSWITHOUTTHEUSEOFELECTRONICMEASUREMENTINSTPDF

链接地址:http://www.mydoc123.com/p-527388.html